HemosIL HIT-Ab(PF4-H) 00020014600Recalls & reported problems

Instrumentation Laboratory Company · HemosIL HIT-Ab(PF4-H) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
1
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08426950975380), received between Apr 22, 2024 and Aug 7, 2025 — classified as 1 death, 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result4
  • Adverse Event Without Identified Device or Use Problem1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2
  • Stroke/CVA1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    POTENTIALLY ERRONEOUS RESULT GENERATED DUE TO HIT ALTERNATIVE PREDILUTION (D) TEST BEING ORDERED INSTEAD OF THE MAIN TEST USING HEMOSIL HIT-AB (PF4-H) LOT B37012 ON ACL TOP 550 CTS SN (B)(6). A COMPLAINT WAS SUBMITTED ON HEMOSIL HIT-AB (PF4-H), LOT B37012, PN 00020014600, INDICATING THAT A LABORATORY USER MISTAKENLY ORDERED AND PERFORMED A HIT ALTERNATIVE PREDILUTION (D) TEST INSTEAD OF THE STANDARD HIT_AB TEST ON (B)(6) 2025. AT THIS TIME, THE IMPACT ON THE PATIENT REMAINS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 22737450 ↗
  • MalfunctionJun 18, 2025False Negative Result; Adverse Event Without Identified Device or Use Problem; False Negative Result; Adverse Event Without Identified Device or Use Problem

    A COMPLAINT WAS FILED ON MAY 20, 2025 REGARDING A DISCREPANT HIT TEST RESULT. THE PATIENT'S HIT RESULTS ON THE ACL TOP INSTRUMENT RETURNED A NEGATIVE RESULT, BUT A FOLLOW-UP SRA CONDUCTED AT AN EXTERNAL FACILITY ON YIELDED A POSITIVE RESULT. THE INSTRUMENT WAS CALIBRATED ON (B)(6) 2025. DUE TO CLINICAL SUSPICION, THE PATIENT WAS INITIALLY TREATED WITH AN ALTERNATIVE ANTICOAGULANT, BRIEFLY SWITCHED BACK TO HEPARIN, AND THEN RESUMED APPROPRIATE THERAPY (BIVALIRUDIN AND LATER APIXABAN) UPON RECEIVING THE POSITIVE SRA. THE ORIGINAL SERUM SAMPLE HAS BEEN DISCARDED, AND NO ELISA TESTING WAS PERFORMED. NO KNOWN PATIENT IMPACT WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 22284252 ↗
  • DeathMay 28, 2025False Negative Result; False Negative Result

    A COMPLAINT HAS BEEN FILED FOR A DISCREPANT HIT TEST RESULT. THE PATIENT SAMPLE ID (B)(6), TESTED ON THE ACL TOP USING REAGENT/CALIBRATOR PN 00020014600, YIELDED A NEGATIVE RESULT. THE SAME SAMPLE, WHEN SENT TO MAYO CLINIC, RETURNED A POSITIVE SRA RESULT (CUT-OFF: >0.400 POSITIVE). THE SAMPLE WAS INITIALLY SENT TO MAYO AFTER PHARMACY QUESTIONED THE NEGATIVE RESULT, GIVEN THE PATIENT'S CLINICAL HISTORY OF A LEFT APICAL THROMBUS, STROKE, AND SUBSEQUENT DEATH UNDER PALLIATIVE CARE. THE SITE HAS NOW DISCONTINUED HIT TESTING ON THE ACL TOP AND IS SENDING SAMPLES TO QUEST.

    Patient problems: Stroke/CVA; Thrombosis/Thrombus

    MAUDE report 22100252 ↗
  • InjuryApr 24, 2024False Negative Result; False Negative Result

    CUSTOMER REPORTED TWO COMPLAINTS REGARDING A NEGATIVE RESULT FOR HIT-AB RUN ON TWO SEPARATE ACL TOP 700 INSTRUMENTS. A SEPARATE REPORT (1217183-2024-00002) HAS BEEN FILED FOR THE SAMPLE ANALYZED ON ACL TOP 700 SERIAL NUMBER (B)(6). IT WAS REPORTED THAT A PATIENT SUSPECTED OF HIT RECEIVED A NEGATIVE RESULT ON THE ACL TOP 700 SERIAL NUMBER (B)(6). OTHER METHODS OF TESTING (ELISA AND SRA) WERE POSITIVE. THE CUSTOMER CONSIDERS THE NEGATIVE RESULT FROM THE ACL TOP 700 ERRONEOUS. THE NEGATIVE RESULT FROM THE ACL TOP 700 WAS NOT REPORTED, AS THE POSITIVE ELISA TEST FIT THE PATIENT'S PROFILE. NO CORRECTED REPORT WAS ISSUED AND THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19175720 ↗
  • InjuryApr 22, 2024False Negative Result; False Negative Result

    CUSTOMER REPORTED TWO COMPLAINTS REGARDING A NEGATIVE RESULT FOR HIT-AB RUN ON TWO SEPARATE ACL TOP 700 INSTRUMENTS. A SEPARATE REPORT (1217183-2024-00003) HAS BEEN FILED FOR THE SAMPLE ANALYZED ON ACL TOP 700 SERIAL NUMBER (B)(6). IT WAS REPORTED THAT A PATIENT SUSPECTED OF HIT RECEIVED A NEGATIVE RESULT ON THE ACL TOP 700 SERIAL NUMBER (B)(6). OTHER METHODS OF TESTING (ELISA AND SRA) WERE POSITIVE. THE CUSTOMER CONSIDERS THE NEGATIVE RESULT FROM THE ACL TOP 700 ERRONEOUS. THE NEGATIVE RESULT FROM THE ACL TOP 700 WAS NOT REPORTED, AS THE POSITIVE ELISA TEST FIT THE PATIENT'S PROFILE. NO CORRECTED REPORT WAS ISSUED AND THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19160534 ↗

Other PLATELET FACTOR 4 RADIOIMMUNOASSAY devices

All PLATELET FACTOR 4 RADIOIMMUNOASSAY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page