Skip to content

HemosIL AcuStar HIT-IgG (PF4-H) 0009800127Recalls & reported problems

Instrumentation Laboratory Company · 1. HIT- IgG(PF4-H)Cartridges for 25 determinations (Cat No 0009800128) 2. HIT-IgG (PF4-H) Calibaratior 1 (Cat No 0009800129) 3. HIT-IgG (PF4-H) Calibaratior 2 (Cat No 0009800130) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08426951126545), received between Apr 2, 2025 and Sep 22, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K170854); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 3, 2026Open, Classified

    Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.

    Z-0001-2027 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • False Negative Result2
  • No Apparent Adverse Event2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2026False Negative Result

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50859735 ↗
  • MalfunctionAug 20, 2026False Negative Result

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50347530 ↗
  • MalfunctionJun 6, 2025No Apparent Adverse Event; No Apparent Adverse Event

    ON (B)(6) 2025, THE CUSTOMER REPORTED 3 PATIENTS WITH DISCREPANT RESULTS BETWEEN HEMOSIL ACUSTAR HIT IGG (PF4-H) (NEGATIVE RESULTS) AND ELISA METHOD (POSITIVE RESULTS). NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22161249 ↗
  • MalfunctionApr 2, 2025No Apparent Adverse Event; No Apparent Adverse Event

    CUSTOMER REPORTS 2 DISCREPANT HIT PATIENT RESULTS WITH ACL TOP. NO FALSE RESULTS WERE REPORTED. NO PATIENT IMPACT AS RESULTS FROM THE ACL ACUSTAR WERE NOT RELEASED DUE TO ON GOING VALIDATIONS STUDIES. 2 PATIENT SAMPLES RESULTED AS NEGATIVE ON THE ACUSTAR. SAME SAMPLES WERE RUN ON THE ACL TOP AND RESULTED AS POSITIVE. SAMPLES WERE SENT OUT FOR SRA-BOTH SAMPLES RESULTED AS POSITIVE. PATIENTS HAVE BEEN CONFIRMED TO CLINICALLY HAVE HIT. BOTH PATIENTS WERE ON HEPARIN AT TIME OF TESTING. PATIENT FROM 1/20 HAD A PLATELET COUNT OF 200 DROP TO 65. PATIENT FROM 2/27 HAD A PLATELET COUNT OF 200 DROP TO 87. PLATELET COUNTS CAME BACK UP ONCE THE HEPARIN WAS STOPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21752118 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K170854. One submission often covers many models, so check the product and lot details in each record.

  • Z-0002-2027 · Sep 3, 2026 · Open, Classified
    HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)

Other PLATELET FACTOR 4 RADIOIMMUNOASSAY devices

All PLATELET FACTOR 4 RADIOIMMUNOASSAY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page