FlowArt® A1010H-SRecalls & reported problems

Asset Medikal Tasarim Sanayi Ve Ticaret Anonim Sirketi · Needle-Free Split Septum Valve, Single, Hemodialysis · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08699443580044), received between Sep 17, 2024 and Sep 17, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 6 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem6
  • Incomplete or Inadequate Connection1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2024Connection Problem; Connection Problem

    SIX (6) SEPARATE EVENTS HAVE OCCURRED SINCE (B)(6) 2024 WHERE THE DIALYSIS BLOOD LINE BECOMES DISCONNECTED FROM THE FLOWART NEEDLE-FREE SPLIT SEPTUM VALVE WHILE THE PATIENT IS RECEIVING A DIALYSIS TREATMENT. THIS HAS RESULTED IN THE LOSS OF BLOOD ON SEVERAL OF THESE OCCASIONS. THE DISTRIBUTER/MANUFACTURER HAS BEEN NOTIFIED OF EACH EVENT. REFERENCE REPORTS: MW5158285, MW5158286, MW5158287, MW5158288, MW5158289. DURING A ROUTINE DIALYSIS TREATMENT, A NURSE WAS SWABBING THE A1010H-S VALVE WITH ALCOHOL, PER PROCEDURE, WHEN THE VALVE BECAME DISCONNECTED FROM THE BLOOD LINE. THE BLOOD LINE WAS NOT INTENDED TO BE DISCONNECTED AND THE VALVE WAS NOT MANIPULATED IN A MANNER THAT WOULD OTHERWISE RESULT IN A DISCONNECTION. THE A1010H-S VALVE SEEMED TO NOT HAVE BEEN THREADED ONTO THE BLOOD LINE CORRECTLY. THE NURSE WAS EXPERIENCED AND HAS PERFORMED THIS ACTION MANY TIMES WITHOUT A SUDDEN DISCONNECTION OF THE VALVE FROM THE BLOOD LINE. STAFF AT THIS DIALYSIS CENTER HAVE REPORTED THAT THE THREADS OF THE VALVE SOMETIMES CAN BE VERY DIFFICULT TO CONNECT TO SYRINGES OR BLOOD LINES. THEY HAVE NOT REPORTED ISSUES CONNECTING THE VALVE TO THE PATIENT DIALYSIS CATHETER. BASED ON THE DIFFICULTY OF THREADI

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20238442 ↗
  • MalfunctionSep 17, 2024Connection Problem; Connection Problem

    SIX (6) SEPARATE EVENTS HAVE OCCURRED SINCE MAY 2ND, 2024 WHERE THE DIALYSIS BLOOD LINE BECOMES DISCONNECTED FROM THE FLOWART NEEDLE-FREE SPLIT SEPTUM VALVE WHILE THE PATIENT IS RECEIVING A DIALYSIS TREATMENT. THIS HAS RESULTED IN THE LOSS OF BLOOD ON SEVERAL OF THESE OCCASIONS. THE DISTRIBUTER/MANUFACTURER HAS BEEN NOTIFIED OF EACH EVENT. REFERENCE REPORTS: MW5158285, MW5158286, MW5158287, MW5158288, MW5158289. DURING A ROUTINE DIALYSIS TREATMENT, A NURSE WAS SWABBING THE A1010H-S VALVE WITH ALCOHOL, PER PROCEDURE, WHEN THE VALVE BECAME DISCONNECTED FROM THE BLOOD LINE. THE BLOOD LINE WAS NOT INTENDED TO BE DISCONNECTED AND THE VALVE WAS NOT MANIPULATED IN A MANNER THAT WOULD OTHERWISE RESULT IN A DISCONNECTION. THE A1010H-S VALVE SEEMED TO NOT HAVE BEEN THREADED ONTO THE BLOOD LINE CORRECTLY. THE NURSE WAS EXPERIENCED AND HAS PERFORMED THIS ACTION MANY TIMES WITHOUT A SUDDEN DISCONNECTION OF THE VALVE FROM THE BLOOD LINE. STAFF AT THIS DIALYSIS CENTER HAVE REPORTED THAT THE THREADS OF THE VALVE SOMETIMES CAN BE VERY DIFFICULT TO CONNECT TO SYRINGES OR BLOOD LINES. THEY HAVE NOT REPORTED ISSUES CONNECTING THE VALVE TO THE PATIENT DIALYSIS CATHETER. BASED ON THE DIFFICULTY OF THREA

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20238440 ↗
  • MalfunctionSep 17, 2024Connection Problem; Connection Problem

    SIX (6) SEPARATE EVENTS HAVE OCCURRED SINCE MAY 2ND, 2024 WHERE THE DIALYSIS BLOOD LINE BECOMES DISCONNECTED FROM THE FLOWART NEEDLE-FREE SPLIT SEPTUM VALVE WHILE THE PATIENT IS RECEIVING A DIALYSIS TREATMENT. THIS HAS RESULTED IN THE LOSS OF BLOOD ON SEVERAL OF THESE OCCASIONS. THE DISTRIBUTER/MANUFACTURER HAS BEEN NOTIFIED OF EACH EVENT. REFERENCE REPORTS: MW5158285, MW5158286, MW5158287, MW5158288, MW5158289. DURING A ROUTINE DIALYSIS TREATMENT, A NURSE WAS SWABBING THE A1010H-S VALVE WITH ALCOHOL, PER PROCEDURE, WHEN THE VALVE BECAME DISCONNECTED FROM THE BLOOD LINE. THE BLOOD LINE WAS NOT INTENDED TO BE DISCONNECTED AND THE VALVE WAS NOT MANIPULATED IN A MANNER THAT WOULD OTHERWISE RESULT IN A DISCONNECTION. THE A1010H-S VALVE SEEMED TO NOT HAVE BEEN THREADED ONTO THE BLOOD LINE CORRECTLY. THE NURSE WAS EXPERIENCED AND HAS PERFORMED THIS ACTION MANY TIMES WITHOUT A SUDDEN DISCONNECTION OF THE VALVE FROM THE BLOOD LINE. STAFF AT THIS DIALYSIS CENTER HAVE REPORTED THAT THE THREADS OF THE VALVE SOMETIMES CAN BE VERY DIFFICULT TO CONNECT TO SYRINGES OR BLOOD LINES. THEY HAVE NOT REPORTED ISSUES CONNECTING THE VALVE TO THE PATIENT DIALYSIS CATHETER. BASED ON THE DIFFICULTY OF THREA

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20238436 ↗
  • MalfunctionSep 17, 2024Connection Problem; Connection Problem

    SIX (6) SEPARATE EVENTS HAVE OCCURRED SINCE MAY 2ND, 2024 WHERE THE DIALYSIS BLOOD LINE BECOMES DISCONNECTED FROM THE FLOWART NEEDLE-FREE SPLIT SEPTUM VALVE WHILE THE PATIENT IS RECEIVING A DIALYSIS TREATMENT. THIS HAS RESULTED IN THE LOSS OF BLOOD ON SEVERAL OF THESE OCCASIONS. THE DISTRIBUTER/MANUFACTURER HAS BEEN NOTIFIED OF EACH EVENT. REFERENCE REPORTS: MW5158285, MW5158286, MW5158287, MW5158288, MW5158289. DURING A ROUTINE DIALYSIS TREATMENT, A NURSE WAS SWABBING THE A1010H-S VALVE WITH ALCOHOL, PER PROCEDURE, WHEN THE VALVE BECAME DISCONNECTED FROM THE BLOOD LINE. THE BLOOD LINE WAS NOT INTENDED TO BE DISCONNECTED AND THE VALVE WAS NOT MANIPULATED IN A MANNER THAT WOULD OTHERWISE RESULT IN A DISCONNECTION. THE A1010H-S VALVE SEEMED TO NOT HAVE BEEN THREADED ONTO THE BLOOD LINE CORRECTLY. THE NURSE WAS EXPERIENCED AND HAS PERFORMED THIS ACTION MANY TIMES WITHOUT A SUDDEN DISCONNECTION OF THE VALVE FROM THE BLOOD LINE. STAFF AT THIS DIALYSIS CENTER HAVE REPORTED THAT THE THREADS OF THE VALVE SOMETIMES CAN BE VERY DIFFICULT TO CONNECT TO SYRINGES OR BLOOD LINES. THEY HAVE NOT REPORTED ISSUES CONNECTING THE VALVE TO THE PATIENT DIALYSIS CATHETER. BASED ON THE DIFFICULTY OF THREA

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20238422 ↗
  • MalfunctionSep 17, 2024Connection Problem; Incomplete or Inadequate Connection; Connection Problem; Incomplete or Inadequate Connection

    SIX (6) SEPARATE EVENTS HAVE OCCURRED SINCE MAY 2ND, 2024 WHERE THE DIALYSIS BLOOD LINE BECOMES DISCONNECTED FROM THE FLOWART NEEDLE-FREE SPLIT SEPTUM VALVE WHILE THE PATIENT IS RECEIVING A DIALYSIS TREATMENT. THIS HAS RESULTED IN THE LOSS OF BLOOD ON SEVERAL OF THESE OCCASIONS. THE DISTRIBUTER/MANUFACTURER HAS BEEN NOTIFIED OF EACH EVENT. REFERENCE REPORTS: MW5158285, MW5158286, MW5158287, MW5158288, MW5158289. DURING A ROUTINE DIALYSIS TREATMENT, A NURSE WAS SWABBING THE A1010H-S VALVE WITH ALCOHOL, PER PROCEDURE, WHEN THE VALVE BECAME DISCONNECTED FROM THE BLOOD LINE. THE BLOOD LINE WAS NOT INTENDED TO BE DISCONNECTED AND THE VALVE WAS NOT MANIPULATED IN A MANNER THAT WOULD OTHERWISE RESULT IN A DISCONNECTION. THE A1010H-S VALVE SEEMED TO NOT HAVE BEEN THREADED ONTO THE BLOOD LINE CORRECTLY. THE NURSE WAS EXPERIENCED AND HAS PERFORMED THIS ACTION MANY TIMES WITHOUT A SUDDEN DISCONNECTION OF THE VALVE FROM THE BLOOD LINE. STAFF AT THIS DIALYSIS CENTER HAVE REPORTED THAT THE THREADS OF THE VALVE SOMETIMES CAN BE VERY DIFFICULT TO CONNECT TO SYRINGES OR BLOOD LINES. THEY HAVE NOT REPORTED ISSUES CONNECTING THE VALVE TO THE PATIENT DIALYSIS CATHETER. BASED ON THE DIFFICULTY OF THREA

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20238419 ↗
  • MalfunctionSep 17, 2024Connection Problem; Connection Problem

    SIX (6) SEPARATE EVENTS HAVE OCCURRED SINCE MAY 2ND, 2024 WHERE THE DIALYSIS BLOOD LINE BECOMES DISCONNECTED FROM THE FLOWART NEEDLE-FREE SPLIT SEPTUM VALVE WHILE THE PATIENT IS RECEIVING A DIALYSIS TREATMENT. THIS HAS RESULTED IN THE LOSS OF BLOOD ON SEVERAL OF THESE OCCASIONS. THE DISTRIBUTER/MANUFACTURER HAS BEEN NOTIFIED OF EACH EVENT. REFERENCE REPORTS: MW5158285, MW5158286, MW5158287, MW5158288, MW5158289. DURING A ROUTINE DIALYSIS TREATMENT, A NURSE WAS SWABBING THE A1010H-S VALVE WITH ALCOHOL, PER PROCEDURE, WHEN THE VALVE BECAME DISCONNECTED FROM THE BLOOD LINE. THE BLOOD LINE WAS NOT INTENDED TO BE DISCONNECTED AND THE VALVE WAS NOT MANIPULATED IN A MANNER THAT WOULD OTHERWISE RESULT IN A DISCONNECTION. THE A1010H-S VALVE SEEMED TO NOT HAVE BEEN THREADED ONTO THE BLOOD LINE CORRECTLY. THE NURSE WAS EXPERIENCED AND HAS PERFORMED THIS ACTION MANY TIMES WITHOUT A SUDDEN DISCONNECTION OF THE VALVE FROM THE BLOOD LINE. STAFF AT THIS DIALYSIS CENTER HAVE REPORTED THAT THE THREADS OF THE VALVE SOMETIMES CAN BE VERY DIFFICULT TO CONNECT TO SYRINGES OR BLOOD LINES. THEY HAVE NOT REPORTED ISSUES CONNECTING THE VALVE TO THE PATIENT DIALYSIS CATHETER. BASED ON THE DIFFICULTY OF THREA

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20238412 ↗

Other Set, Administration, Intravascular devices