Greenfield™ M001505010Recalls & reported problems
Boston Scientific Corporation · Vena Cava Filter · Class 2 · Implant
All models of Greenfield (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Greenfield vena cava filter is a small implant placed in the large vein of the abdomen to catch blood clots. Many of the 113 reports describe the filter tilting or shifting out of position, with 85 listing perforation of the vessel wall. The reports include 97 injuries and no deaths, and 5 list a pulmonary embolism. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 113
- Deaths reported
- 0
- Injuries reported
- 97
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 113 adverse event reports that list this exact device (DI 08714729124993), received between Jun 4, 2018 and Nov 2, 2022 — classified as 97 injuries, 16 malfunctions.
The most frequently reported specific device problem is “Positioning Problem” (71 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Positioning Problem71
- Adverse Event Without Identified Device or Use Problem36
- Migration12
- Break4
- Detachment Of Device Component1
Patient problems
- Perforation85
- No Consequences Or Impact To Patient9
- No Known Impact Or Consequence To Patient7
- Pulmonary Embolism5
- Perforation of Vessels3
- Thrombosis3
- No Code Available2
- Bowel Perforation1
Recent adverse event reports
The 25 most recent of 113 reports, in the reporter's own words as published by the FDA.
InjuryNov 2, 2022Positioning Problem; Positioning Problem
IT WAS REPORTED THAT THE FILTER WAS TILTED AND HAD PERFORATED THE CONFINES OF THE INFERIOR VENA CAVA (IVC) AND INTO SURROUNDING STRUCTURES. ON (B)(6) 2009, THE PATIENT WAS IMPLANTED WITH THE GREENFIELD FILTER. ON (B)(6) 2020, THE PATIENT UNDERWENT AN ABDOMINAL AND PELVIC CT SCAN. THE FILTER WAS OBSERVED TO BE TILTED, WITH THE APEX ABUTTING THE IVC WALL. MULTIPLE STRUTS HAD PERFORATED THE IVC WALL, WITH ONE STRUT PERFORATING THE SMALL INTESTINE. THERE WERE NO FURTHER REPORTED PATIENT COMPLICATIONS.
Patient problems: Perforation
MAUDE report 15718733 ↗InjuryOct 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE FILTER HAD PERFORATED THE SURROUNDING STRUCTURES. ON (B)(6) 2010, THE PATIENT WAS IMPLANTED WITH GREENFIELD FILTER FOR CONTROL OF PULMONARY EMBOLISM (PE). ON (B)(6) 2020, THE PATIENT RECEIVED AN ABDOMINAL CT SCAN. THE INFERIOR VENA CAVA (IVC) FILTER WAS NOTED TO HAVE PERFORATED THE CONFINES OF THE IVC. THERE WAS A 3 MM AORTIC PERFORATION, A 4 MM RENAL PERFORATION, AND 3 MM MESENTERIC PERFORATION. THERE WAS NO EVIDENCE OF TILTING, FRACTURE, STENOSIS OR MIGRATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Perforation
MAUDE report 15572473 ↗InjurySep 8, 2022Positioning Problem; Positioning Problem
IT WAS REPORTED THAT THE FILTER WAS TILTED AND HAD PERFORATED THE INFERIOR VENA CAVA (IVC). ON (B)(6) 2007, THE PATIENT WAS IMPLANTED WITH A GREENFIELD FILTER. ON (B)(6) 2020, THE PATIENT RECEIVED AN ABDOMINAL CT SCAN. THE SUPERIOR TIP OF THE FILTER WAS TILTED LATERALLY TOWARD THE RIGHT. ADDITIONALLY, THE FILTER STRUTS EXTENDED BEYOND THE IVC LUMEN ANTERIORLY AND LATERALLY TOWARD THE RIGHT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Perforation
MAUDE report 15381268 ↗InjuryJul 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON OR ABOUT (B)(6) 2002, THE PATIENT WAS IMPLANTED WITH A GREENFIELD FILTER AFTER DIAGNOSIS AND TREATMENT OF A DEEP VEIN THROMBOSIS. ON OR ABOUT (B)(6) 2019, THE PATIENT UNDERWENT A COMPUTERIZED TOMOGRAPHY SCAN (CT SCAN) OF THEIR ABDOMEN AND PELVIS. THE PATIENT WAS INFORMED THAT THE FILTER'S STRUTS PERFORATED BEYOND THE INFERIOR VENA CAVA (IVC) WALL.
Patient problems: Perforation
MAUDE report 15129299 ↗InjuryJul 14, 2022Positioning Problem; Positioning Problem
IT WAS REPORTED THAT THE FILTER HAD TITLED AND PERFORATED THE IVC. A GREENFIELD IVC FILTER WAS IMPLANTED INTO THE INFERIOR VENA CAVA (IVC) ON (B)(6) 2008 AFTER TREATMENT OF DEEP VEIN THROMBOSIS (DVT). ON (B)(6) 2020, A COMPUTATED TOMOGRAPHY (CT) SCAN WAS PERFORMED OF THE ABDOMEN AND PELVIS. THE SCAN REVEALED THAT THE FILTER HAD TITLED AND CONTACTS THE ANTERIOR WALL OF THE IVC. THE LEFT LATERAL STRUTS PERFORATED THE IVC WALL AND THERE IS A 3MM VERTEBRAL PERFORATION.
Patient problems: Perforation
MAUDE report 15021554 ↗InjuryJul 14, 2022Break; Break
ON OR ABOUT (B)(6) 2005, THE PATIENT WAS SURGICALLY IMPLANTED WITH A GREENFIELD STAINLESS STEEL VENA CAVA FILTER AFTER DIAGNOSIS AND TREATMENT OF A DEEP VEIN THROMBOSIS. ON OR ABOUT (B)(6) 2020, THE PATIENT UNDERWENT A COMPUTERIZED TOMOGRAPHY SCAN (CT SCAN) OF THEIR ABDOMEN AND PELVIS. THE CT SCAN REVEALED THEIR INFERIOR VENA CAVA (IVC) WAS SMALL AT THE LEVEL OF THEIR IVC FILTER WHICH MAY BE SECONDARY TO IVC THROMBOSIS. ADDITIONALLY, A POSTERIOR STRUT PERFORATED THE IVC WALL BY 3.5MM, AND A STRUT FRACTURE WAS SEEN.
Patient problems: Perforation; Thrombosis/Thrombus
MAUDE report 15021146 ↗InjuryMar 29, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THERE WAS A PULMONARY EMBOLISM. A GREENFIELD FILTER WAS IMPLANTED ON (B)(6) 2007. ON (B)(6) 2008 THE PATIENT WAS DIAGNOSED WITH HAVING A PULMONARY EMBOLISM. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pulmonary Embolism
MAUDE report 11581831 ↗InjuryMar 10, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A PULMONARY EMBOLISM. A GREENFIELD FILTER WAS IMPLANTED ON (B)(6) 2010. ON (B)(6) 2014 A PULMONARY EMBOLISM WAS OBSERVED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Embolism/Embolus
MAUDE report 11453696 ↗InjuryApr 13, 2020Positioning Problem; Positioning Problem
ON (B)(6)2007 THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2018 A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED THE APEX OF THE FILTER WAS 16-17 DEGREES TO THE LEFT AND WAS TOUCHING THE LEFT SIDE OF THE INFERIOR VENA CAVA (IVC) WALL. ALL 6 STRUTS PERFORATE THE IVC WALL AND EXTEND 1-3 MM INTO SURROUNDING FAT. A MESENTERIC PERFORATION WAS NOTED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: Perforation
MAUDE report 9955868 ↗InjuryApr 10, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED A PERFORATION OCCURRED. ON (B)(6) 2007, THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2019, A COMPUTED TOMOGRAPHY (CT) SCAN CONFIRMED NO SIGNIFICANT TILT. THE FILTER TINES EXTEND BEYOND THE WALL OF INFERIOR VENA CAVA (IVC). IT WAS FURTHER REPORTED THAT THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER ON (B)(6) 2007.
Patient problems: Perforation
MAUDE report 9949160 ↗InjuryApr 10, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2009 THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2019, A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED EXTRALUMINAL EXTENSION OF THE STRUT ALONG THE POSTERIOR ASPECT MEASURING 4MM. NO FRACTURE OF THE FILTER OR STRUT WAS NOTED.
Patient problems: Perforation
MAUDE report 9949028 ↗InjuryApr 10, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2001, THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2019, A COMPUTED TOMOGRAPHY (CT) SCAN SHOWED POSITIVE MESENTERIC PERFORATION. PERIRENAL INFERIOR VENA CAVA FILTER WAS NOTED TO HAVE ONE OF THE PRONGS EXTENDING 1MM OUTSIDE THE INFERIOR VENA CAVA (IVC) WALL.
Patient problems: Perforation
MAUDE report 9949010 ↗MalfunctionApr 10, 2020Positioning Problem; Positioning Problem
ON (B)(6) 2015 THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2020, A COMPUTED TOMOGRAPHY (CT) SCAN OF THE ABDOMEN REVEALED THAT THE FILTER WAS SLIGHTLY TILED BACKWARDS WITH ITS SUPERIOR TIP APPEARING TO BE IMPLANTED WITHIN THE POSTERIOR INFERIOR VENA CAVA (IVC) WALL.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9949009 ↗InjuryApr 10, 2020Positioning Problem; Positioning Problem
ON (B)(6) 2000, THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER (IVC). ON (B)(6) 2019 A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED THAT THE GREENFIELD VENA CAVA FILTER WAS IN PLACE WITH A MODERATE LEFTWARD TILT. A COUPLE OF STRUTS WERE ALSO NOTED TO BE PERFORATING THE INFERIOR VENA CAVA (IVC) UP TO 6MM WITHOUT ENCROACHMENT INTO BOWEL OR AORTA.
Patient problems: Perforation
MAUDE report 9948888 ↗InjuryApr 10, 2020Positioning Problem; Positioning Problem
ON (B)(6) 2014 THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2019, A COMPUTATED TOMOGRAPHY (CT) SCAN REVEALED THE FILTER WAS POSITION BELOW THE LEVEL OF THE RENAL VEINS. THERE IS TILTING PROXIMALLY TO RIGHT BY 12 DEGREES. DISTAL ANCHORING STRUTS DO NOT EXTEND BEYOND VESSEL WALL. SECOND IVC FILTER LIES OUTSIDE VASCULAR SYSTEM IN RIGHT RETROPERITONEUM ALONG POSTERO-MEDIAL ASPECT OF ROOT PSOAS MUSCLE. ANCHORING STRUTS MAY BE WITHIN RIGHT COMMON ILIAC VEIN.
Patient problems: Perforation
MAUDE report 9948494 ↗InjuryApr 9, 2020Positioning Problem; Positioning Problem
ON (B)(6) 2012, THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD INFERIOR VENA CAVA (IVC) FILTER. ON (B)(6) 2019, A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED TILT AND PERFORATION. THE IVC FILTER WAS LOCATED BELOW THE LEVEL OF THE RENAL VEINS APPROXIMATELY ABOUT 1.8CM. THE APEX OF THE IVC FILTER WAS TOUCHING THE ANTERIOR WALL OF THE IVC. THERE APPEARED TO BE MILD ANTERIOR TILT AT THE APEX WHICH MEASURED UP TO 10 DEGREES. THE TIPS OF ALL LIMBS OF THE IVC FILTER PROTRUDED BEYOND THE WALL OF THE IVC UP TO 2-3MM.
Patient problems: Perforation
MAUDE report 9947462 ↗InjuryApr 9, 2020Positioning Problem; Positioning Problem
ON (B)(6) 2005, THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD INFERIOR VENA CAVA (IVC) FILTER. ON (B)(6) 2019, A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED POSITIVE TILT AND MESENTERIC PERFORATION. THE FILTER WAS TILTED TO THE LEFT WITH ITS PROXIMAL CONE LYING ON THE WALL OF THE IVC, POSSIBLY EMBEDDED IN IT. SOME OF THE FILTER STRUTS HAVE PENETRATED THROUGH THE WALL OF THE IVC INTO THE PRECAVAL/MESENTERIC FAT. THERE WAS NO SIGN OF MIGRATION, FRACTURE OR STENOSIS.
Patient problems: Perforation
MAUDE report 9946855 ↗InjuryApr 9, 2020Positioning Problem; Migration; Positioning Problem; Migration
ON (B)(6) 2001, THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2019, A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED THAT THE IVC FILTER TIP WAS LOCATED ABNORMALLY LOW BELOW THE LOWEST RENAL VEIN, BEGINNING 3CM INFERIOR TO LEFT RENAL VEIN. IT WAS ABNORMALLY TILTED BY 32 DEGREES TO THE LEFT AND 10 DEGREES POSTERIORLY. THERE WAS A PERFORATION OF ALL STRUTS WITHOUT FRACTURE. NUMEROUS STRUTS PERFORATED UP TO 14MM BEYOND THE LUMEN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED IN RELATION TO THIS EVENT.
Patient problems: Perforation
MAUDE report 9946850 ↗InjuryApr 9, 2020Positioning Problem; Positioning Problem
ON (B)(6) 2016, THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD INFERIOR VENA CAVA (IVC) FILTER. ON (B)(6) 2018, A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED ABNORMAL TILTING OF THE FILTER OF OVER 20 DEGREES. IT WAS ALSO NOTED THAT MULTIPLE STRUTS WERE PERFORATING THE IVC WALL UP TO 4MM AND THE IVC STRUTS EXTENDED AND ABUTTED THE THIRD PORTION OF THE DUODENUM AND AORTIC WALL. THERE WAS NO EVIDENCE OF MIGRATION, FRACTURE OR STENOSIS.
Patient problems: Perforation
MAUDE report 9946792 ↗InjuryApr 9, 2020Positioning Problem; Positioning Problem
IT WAS REPORTED A PERFORATION AND TILT OCCURRED. ON (B)(6) 2003, THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6)2018, A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED THE FILTER IS TILTED POSTERIORLY WITH THE TIP ABUTTING THE POSTERIOR WALL OF THE IVC. ALSO, FOUR OF THE SIX STRUTS PERFORATE THE IVC WALL.
Patient problems: Perforation
MAUDE report 9944333 ↗InjuryApr 9, 2020Positioning Problem; Positioning Problem
ON (B)(6) 2008 THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2019, A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED TILT OF THE FILTER AND MESENTERIC PERFORATION. THERE WAS MINIMAL STRUT WALL PERFORATION ALONG THE TIP OF FIVE (5) AND SIX (6) OF THE CAVAL LEGS, WITH NO ORGAN PENETRATION. THERE WAS A 30 DEGREE TILT TO THE RIGHT. NO STENOSIS, FRACTURE, OR MIGRATION OF THE FILTER WAS NOTED.
Patient problems: Perforation
MAUDE report 9943899 ↗InjuryApr 9, 2020Positioning Problem; Positioning Problem
ON (B)(6) 2007, THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2018, MSR2 COMPUTED TOMOGRAPHY (CT) SCAN REVEALED TWO OF THE FILTER PRONGS LAYING ANTERIOR AND APPEAR TO EXTEND BEYOND THE MARGIN OF THE WALL OF THE INFERIOR VENA CAVA (IVC) AND TWO OTHER PRONGS MAY ALSO EXTEND BEYOND THE LUMEN. THE FILTER IS SOMEWHAT TILTED POINTING POSTERIOR AND LATERALLY TO THE RIGHT SIDE.
Patient problems: Perforation
MAUDE report 9943859 ↗InjuryApr 9, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED A PERFORATION OCCURRED. ON (B)(6) 2004, THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2019, A COMPUTED TOMOGRAPHY (CT) SCAN OF THE ABDOMEN WITH CONTRAST WAS PERFORMED AND REVEALED THE INFERIOR VENA CAVA (IVC) FILTER WITH TINES MINIMALLY EXTENDING OUTSIDE IVC. NO SURROUNDING STRANDING WAS NOTICED. THERE WAS A POSITIVE MESENTERIC PERFORATION. RETRO-AORTIC LEFT RENAL VEIN NOTED WITH ONE OF THE TINES SLIGHTLY EXTENDING INTO THE LEFT VEIN.
Patient problems: Perforation
MAUDE report 9943787 ↗InjuryApr 9, 2020Positioning Problem; Positioning Problem
ON (B)(6) 2007, THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED THAT INFRARENAL INFERIOR VENA CAVA (IVC) FILTER NOTED WAS ANGLED AT APPROXIMATELY 10 DEGREES RELATIVE TO THE IVC. A FEW OF THE STRUTS APPEARED TO BE BENDING. ONE OF THE STRUTS WAS SEEN 2 MM FROM THE RIGHT IVC WALL. THE APEX OF THE FILTER ABUTS THE ANTERIOR WALL. IT WAS FURTHER REPORTED THAT THE PATIENT HAD THEIR CT SCAN ON (B)(6) 2018.
Patient problems: Perforation
MAUDE report 9943616 ↗InjuryApr 9, 2020Positioning Problem; Positioning Problem
ON (B)(6) 2007 THE PATIENT UNDERWENT PLACEMENT OF A GREENFIELD VENA CAVA FILTER. ON (B)(6) 2019 A POSITIVE COMPUTED TOMOGRAPHY (CT) SCAN REVEALED A SLIGHT TILT OF THE FILTER TO THE LEFT AND A MESENTERIC PERFORATION.
Patient problems: Perforation
MAUDE report 9943565 ↗
FDA clearance and classification
The FDA regulates product code DTK (filter, intravascular, cardiovascular) under 21 CFR 870.3375 as a Class II device. 21 CFR 870.3375 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Greenfield™ M001505010
Has the Greenfield™ M001505010 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729124993), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Greenfield™ M001505010?
The FDA's MAUDE database lists 113 adverse event reports for this device, received between Jun 4, 2018 and Nov 2, 2022. In those reports, the most common specific device problems are “Positioning Problem” (71), “Migration” (12) and “Break” (4); the most common patient problems are “Perforation” (85), “Pulmonary Embolism” (5) and “Perforation of Vessels” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Greenfield™ M001505010?
None of the reports are classified as deaths, 97 as injuries and 16 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Greenfield™ M001505010 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Greenfield™ M001505010, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Greenfield™, model M001505010). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other FILTER, INTRAVASCULAR, CARDIOVASCULAR devices
- Denali® Vena Cava Filter DL900FBard Peripheral Vascular, Inc. · FILTER, INTRAVASCULAR, CARDIOVASCULAR22 deaths530 reports
- DENALI® Vena Cava Filter DL900JBard Peripheral Vascular, Inc. · FILTER, INTRAVASCULAR, CARDIOVASCULAR15 deaths491 reports
- Option Elite Retrievable Vena Cava Filter System 352506070EArgon Medical Devices, Inc. · Filter, Intravascular, Cardiovascular2 recalls3 deaths340 reports
- Gunther Tulip G33016Cook Incorporated · FILTER, INTRAVASCULAR, CARDIOVASCULAR3 deaths348 reports
- Gunther Tulip G33017Cook Incorporated · FILTER, INTRAVASCULAR, CARDIOVASCULAR3 deaths286 reports
- Cook Celect G34309William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall1 death230 reports
- Gunther Tulip G52916William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall2 deaths189 reports
- Cook Celect G34502William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall2 deaths165 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page