EndoVive™ One Step Button™ M00568510Recalls & reported problems
Boston Scientific Corporation · Low Profile Percutaneous Endoscopic Gastrostomy Kit · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729151081), received between Jan 7, 2019 and Jul 6, 2021 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Difficult to Advance1
- Obstruction of Flow1
Patient problems
- Hemorrhage/Blood Loss/Bleeding1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 6, 2021Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE ONESTEP BUTTON WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE ON (B)(6) 2021. POST PROCEDURE, THE RIGHT ANGLE FEEDING PORT WAS OBSTRUCTED. THE DEVICE WAS REPLACED WITH A NEW RIGHT ANGLE FEEDING TUBE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12121777 ↗MalfunctionSep 10, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE ONE STEP BUTTON WAS USED IN A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE DEVICE WAS DEFECTIVE. THE PROCEDURE WAS COMPLETED WITH A NEW ENDOVIVE ONE STEP BUTTON. THERE WERE NO PATIENT COMPLICATIONS REPORTED DUE TO THIS EVENT. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS: ONE STEP BUTTON DELIVERY SYSTEM WAS BROKEN INTO TWO PIECES AND STRETCHED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10513625 ↗InjuryJun 20, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE ONE STEP BUTTON WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE DEVICE WAS BEING PULLED INSIDE THE BODY, SEVERE RESISTANCE WAS FELT. FORCE WAS APPLIED SLOWLY; HOWEVER, THE SHEATH WITH THE BUTTON INSIDE DETACHED FROM THE TUBE. THE PROCEDURE WAS COMPLETED WITH A NEW ENDOVIVE ONE STEP BUTTON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8715975 ↗MalfunctionJan 7, 2019Difficult to Advance; Difficult to Advance
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE ONE STEP BUTTON PULL METHOD USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN PULLING THE DEVICE THROUGH THE INCISION SITE, NORMALLY IT WOULD GET CAUGHT DUE TO THE BUTTON; HOWEVER, WHEN THEY PULLED THIS DEVICE, THERE WAS NO RESISTANCE MET AND IT WAS COMPLETELY TAKEN OUT FROM THE PATIENT'S BODY AS ONE UNIT. IT WAS ALSO NOTICED THAT THE SITE HAD BLEEDING BUT WAS RESOLVED THROUGH CLIPPING. THE PROCEDURE WAS NOT COMPLETED DUE TO ANOTHER REASON. THERE WERE NO REPORTED PATIENT COMPLICATIONS DUE TO THIS EVENT AND THERE WAS NO ISSUE OBSERVED WITH THE PATIENT'S CONDITION.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 8221693 ↗
Other TUBE, GASTRO-ENTEROSTOMY devices
- AVANOS 98431Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY93 reports
- AVANOS 98432Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY72 reports
- Gastrointestinal Anchor Set, SAF-T-PEXY* T-Fasteners 98701Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY1 death61 reports
- Halyard 98701Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY1 death40 reports
- Halyard 98431Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY35 reports
- AVANOS 98433Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY34 reports
- AVANOS 98430Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY1 death29 reports
- Halyard 98433Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY27 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.