Luge™ H7491213001J0Recalls & reported problems
Boston Scientific Corporation · Guidewire with ICE™ Hydrophilic Coating · Class 2
All models of Luge (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 1
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 08714729169963), received between Aug 21, 2023 and Apr 3, 2024 — classified as 1 death, 13 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component9
- Adverse Event Without Identified Device or Use Problem1
- Device-Device Incompatibility1
- Fracture1
Patient problems
- Foreign Body In Patient11
- No Clinical Signs, Symptoms or Conditions5
- Device Embedded In Tissue or Plaque3
- Hematoma2
- Needle Stick/Puncture2
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryApr 3, 2024Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component
IT WAS REPORTED THAT A GUIDEWIRE FRACTURE OCCURRED. PATIENT UNDERWENT PERCUTANEOUS CORONARY INTERVENTION. THE TARGET LESION WAS LOCATED IN THE DISTAL LEFT ANTERIOR DESCENDING ARTERY. A 182 CM LUGE GUIDEWIRE WAS SELECTED FOR USE. HOWEVER, DURING REMOVAL, THE TIP OF THE WIRE WAS BROKEN AND COULD NOT BE RETRIEVED. THE PROCEDURE WAS COMPLETED.
Patient problems: Foreign Body In Patient
MAUDE report 19032661 ↗InjuryMar 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED VIA FACILITY MEDWATCH REPORT MW5150888 THAT TIP DETACHMENT OCCURRED. TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT (TIPS) WAS PERFORMED. A GADANI WIRE WAS PLACED AS A MARKER IN BOTH THE RIGHT AND MIDDLE HEPATIC PORTAL VEIN. DUE TO OBLITERATION OF PERIPORTAL FAT AND HETEROGENOUS LIVER, TRANSHEPATIC PORTAL WAS DIFFICULT TO ACCESS. SOME INJECTIONS SEEMED TO BE WITHIN INTRAHEPATIC HEMATOMA. A 182CM LUGE GUIDEWIRES WERE SELECTED FOR USED. THREE OCCURRENCES WHERE THE TIP OF THE GUIDEWIRE SHEARED OFF INSIDE THE PATIENT. ON TWO OF THOSE OCCURRENCES, THE PHYSICIAN WAS ABLE TO SNARE AND RETRIEVED THE DEVICE, WHILE ON ONE OCCURRENCE; THE TIP OF THE WIRE HAD TO BE LEFT EMBEDDED IN THE LIVER PARENCHYMA. NO PATIENT COMPLICATIONS WERE REPORTED. THE REPORTED EVENT INFORMATION REFERENCES THREE OCCURRENCES CAPTURED IN EMDR 2124215-2024-07135 (17722511), EMDR 2124215-2024-07133 (17722598), AND EMDR 2124215-2024-07134 (17722711). FOLLOW-UP WAS SENT TO CLARIFY NUMBER OF EVENTS WAS SENT TO THE USER FACILITY WITH NO RESPONSE. THERE IS NO INDICATION OF A 4TH OCCURRENCE; MDR SUBMITTED IN ERROR.
Patient problems: Foreign Body In Patient; Hematoma; Needle Stick/Puncture
MAUDE report 18869206 ↗InjuryMar 8, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED VIA FACILITY MEDWATCH REPORT MW5150887 THAT TIP DETACHMENT OCCURRED. TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT (TIPS) WAS PERFORMED. A GADANI WIRE WAS PLACED AS A MARKER IN BOTH THE RIGHT AND MIDDLE HEPATIC PORTAL VEIN. DUE TO OBLITERATION OF PERIPORTAL FAT AND HETEROGENOUS LIVER, TRANSHEPATIC PORTAL WAS DIFFICULT TO ACCESS. SOME INJECTIONS SEEMED TO BE WITHIN INTRAHEPATIC HEMATOMA. A 182CM LUGE GUIDEWIRES WERE SELECTED FOR USED. THREE OCCURRENCES WHERE THE TIP OF THE GUIDEWIRE SHEARED OFF INSIDE THE PATIENT. ON TWO OF THOSE OCCURRENCES, THE PHYSICIAN WAS ABLE TO SNARE AND RETRIEVED THE DEVICE, WHILE ON ONE OCCURRENCE; THE TIP OF THE WIRE HAD TO BE LEFT EMBEDDED IN THE LIVER PARENCHYMA. NO PATIENT COMPLICATIONS WERE REPORTED. THE RECORD IS A DUPLICATE OF EMDR 2124215-2024-07133 (17722598).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18869204 ↗InjuryMar 8, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED VIA FACILITY MEDWATCH REPORT MW5150890 THAT TIP DETACHMENT OCCURRED. TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT (TIPS) WAS PERFORMED. A GADANI WIRE WAS PLACED AS A MARKER IN BOTH THE RIGHT AND MIDDLE HEPATIC PORTAL VEIN. DUE TO OBLITERATION OF PERIPORTAL FAT AND HETEROGENOUS LIVER, TRANSHEPATIC PORTAL WAS DIFFICULT TO ACCESS. SOME INJECTIONS SEEMED TO BE WITHIN INTRAHEPATIC HEMATOMA. A 182CM LUGE GUIDEWIRES WERE SELECTED FOR USED. THREE OCCURRENCES WHERE THE TIP OF THE GUIDEWIRE SHEARED OFF INSIDE THE PATIENT. ON TWO OF THOSE OCCURRENCES, THE PHYSICIAN WAS ABLE TO SNARE AND RETRIEVED THE DEVICE, WHILE ON ONE OCCURRENCE; THE TIP OF THE WIRE HAD TO BE LEFT EMBEDDED IN THE LIVER PARENCHYMA. NO PATIENT COMPLICATIONS WERE REPORTED. THE REPORT IS A DUPLICATE OF EMDR 2124215-2024-07135 (17722511).
Patient problems: Foreign Body In Patient
MAUDE report 18869177 ↗InjuryMar 8, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED VIA FACILITY MEDWATCH REPORT MW5150889 THAT TIP DETACHMENT OCCURRED. TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT (TIPS) WAS PERFORMED. A GADANI WIRE WAS PLACED AS A MARKER IN BOTH THE RIGHT AND MIDDLE HEPATIC PORTAL VEIN. DUE TO OBLITERATION OF PERIPORTAL FAT AND HETEROGENOUS LIVER, TRANSHEPATIC PORTAL WAS DIFFICULT TO ACCESS. SOME INJECTIONS SEEMED TO BE WITHIN INTRAHEPATIC HEMATOMA. A 182CM LUGE GUIDEWIRES WERE SELECTED FOR USED. THREE OCCURRENCES WHERE THE TIP OF THE GUIDEWIRE SHEARED OFF INSIDE THE PATIENT. ON TWO OF THOSE OCCURRENCES, THE PHYSICIAN WAS ABLE TO SNARE AND RETRIEVED THE DEVICE, WHILE ON ONE OCCURRENCE; THE TIP OF THE WIRE HAD TO BE LEFT EMBEDDED IN THE LIVER PARENCHYMA. NO PATIENT COMPLICATIONS WERE REPORTED. THE RECORD IS A DUPLICATE OF EMDR 2124215-2024-07134 (17722711).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18869170 ↗DeathFeb 28, 2024Detachment of Device or Device Component; Device-Device Incompatibility; Detachment of Device or Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT THE PATIENT DIED. A 2.75 X 8MM SYNERGY XD DRUG ELUTING-STENT AND LUGE GUIDEWIRE WERE USED DURING PROCEDURE. THE GUIDEWIRE BECAME STUCK ON THE STENT AND THE GUIDEWIRE SHEARED OFF IN THE STENT. ANOTHER STENT WAS PLACED TO TRAP THE GUIDEWIRE IN THE VESSEL WALL. THE PATIENT DIED.
Patient problems: Foreign Body In Patient
MAUDE report 18796498 ↗InjuryFeb 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE GUIDEWIRE DETACHED WITHIN THE PATIENT. A LUGE GUIDEWIRE WAS SELECTED FOR USE DURING THE PROCEDURE. HOWEVER, THE TIP OF THE WIRE SHEARED OFF AND HAD TO BE LEFT EMBEDDED IN THE LIVER PARENCHYMA. NO FURTHER COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED VIA FACILITY MEDWATCH REPORT MW5150887 THAT TIP DETACHMENT OCCURRED. TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT (TIPS) WAS PERFORMED. A NON-BOSTON SCIENTIFIC GUIDEWIRE WAS PLACED AS A MARKER IN BOTH THE RIGHT AND MIDDLE HEPATIC PORTAL VEIN. DUE TO OBLITERATION OF PERIPORTAL FAT AND HETEROGENOUS LIVER, TRANSHEPATIC PORTAL WAS DIFFICULT TO ACCESS. SOME INJECTIONS SEEMED TO BE WITHIN INTRAHEPATIC HEMATOMA. A 182CM LUGE GUIDEWIRES WERE SELECTED FOR USED. THREE OCCURRENCES WHERE THE TIP OF THE GUIDEWIRE SHEARED OFF INSIDE THE PATIENT. ON TWO OF THOSE OCCURRENCES, THE PHYSICIAN WAS ABLE TO SNARE AND RETRIEVED THE DEVICE, WHILE ON ONE OCCURRENCE; THE TIP OF THE WIRE HAD TO BE LEFT EMBEDDED IN THE LIVER PARENCHYMA. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 18682042 ↗InjuryFeb 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DISTAL TIP DETACHMENT OCCURRED REQUIRING ADDITIONAL INTERVENTION. A 182CM J LUGE GUIDEWIRE WAS USED DURING THE PROCEDURE. HOWEVER, THE TIP OF THE WIRE SHEARED OFF INSIDE THE PATIENT. THE PHYSICIAN SNARED THE TIP AND RETRIEVED IT. NO FURTHER COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED VIA FACILITY MEDWATCH REPORT MW5150890 THAT TIP DETACHMENT OCCURRED. TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT (TIPS) WAS PERFORMED. A NON-BOSTON SCIENTIFIC GUIDEWIRE WAS PLACED AS A MARKER IN BOTH THE RIGHT AND MIDDLE HEPATIC PORTAL VEIN. DUE TO OBLITERATION OF PERIPORTAL FAT AND HETEROGENOUS LIVER, TRANSHEPATIC PORTAL WAS DIFFICULT TO ACCESS. SOME INJECTIONS SEEMED TO BE WITHIN INTRAHEPATIC HEMATOMA. A 182CM LUGE GUIDEWIRES WERE SELECTED FOR USED. THREE OCCURRENCES WHERE THE TIP OF THE GUIDEWIRE SHEARED OFF INSIDE THE PATIENT. ON TWO OF THOSE OCCURRENCES, THE PHYSICIAN WAS ABLE TO SNARE AND RETRIEVED THE DEVICE, WHILE ON ONE OCCURRENCE; THE TIP OF THE WIRE HAD TO BE LEFT EMBEDDED IN THE LIVER PARENCHYMA. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 18678027 ↗InjuryFeb 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DISTAL TIP DETACHMENT OCCURRED REQUIRING ADDITIONAL INTERVENTION. A LUGE GUIDEWIRE WAS USED DURING THE PROCEDURE. HOWEVER, THE TIP OF THE WIRE SHEARED OFF INSIDE THE PATIENT. THE PHYSICIAN SNARED THE TIP AND RETRIEVED IT. NO FURTHER COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED VIA FACILITY MEDWATCH REPORT MW5150887 THAT TIP DETACHMENT OCCURRED. TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT (TIPS) WAS PERFORMED. A NON-BOSTON SCIENTIFIC GUIDEWIRE WAS PLACED AS A MARKER IN BOTH THE RIGHT AND MIDDLE HEPATIC PORTAL VEIN. DUE TO OBLITERATION OF PERIPORTAL FAT AND HETEROGENOUS LIVER, TRANSHEPATIC PORTAL WAS DIFFICULT TO ACCESS. SOME INJECTIONS SEEMED TO BE WITHIN INTRAHEPATIC HEMATOMA. A 182CM LUGE GUIDEWIRES WERE SELECTED FOR USED. THREE OCCURRENCES WHERE THE TIP OF THE GUIDEWIRE SHEARED OFF INSIDE THE PATIENT. ON TWO OF THOSE OCCURRENCES, THE PHYSICIAN WAS ABLE TO SNARE AND RETRIEVED THE DEVICE, WHILE ON ONE OCCURRENCE; THE TIP OF THE WIRE HAD TO BE LEFT EMBEDDED IN THE LIVER PARENCHYMA. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18677974 ↗InjuryJan 30, 2024
THREE OCCURRENCES WHERE THE TIP OF CATHETER/GUIDEWIRE SHEARED OFF INSIDE OF THE PATIENT. ON TWO OF THOSE OCCURRENCES THE INTERVENTIONAL RADIOLOGIST WAS ABLE TO SNARE THE TIP AND RETRIEVE. ON ONE OCCURRENCE THE TIP THAT SHEARED OFF HAD TO BE LEFT EMBEDDED IN THE LIVER PARENCHYMA. DIFFICULT TRANSHEPATIC PORTAL ACCESS DUE TO OBLITERATION OF PERIPORTAL FAT AND HETEROGENOUS LIVER. MULTIPLE PUNCTURES TO GET INTO PORTAL VEIN. "GADANI WIRE" PLACED AS A MARKER IN BOTH THE RIGHT AND MIDDLE HEPATIC PORTAL VEIN. JUGULAR ACCESS WITH LIVERTY TIPS SET DURING ACCESS, SOME INJECTIONS SEEMED TO BE WITHIN INTRAHEPATIC HEMATOMA. REFERENCE REPORTS: MW5150887, MW5150888, MW5150889.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 18613190 ↗InjuryJan 30, 2024
THREE OCCURRENCES WHERE THE TIP OF CATHETER/GUIDEWIRE SHEARED OFF INSIDE OF THE PATIENT. ON TWO OF THOSE OCCURRENCES THE INTERVENTIONAL RADIOLOGIST WAS ABLE TO SNARE THE TIP AND RETRIEVE. ON ONE OCCURRENCE THE TIP THAT SHEARED OFF HAD TO BE LEFT EMBEDDED IN THE LIVER PARENCHYMA. DIFFICULT TRANSHEPATIC PORTAL ACCESS DUE TO OBLITERATION OF PERIPORTAL FAT AND HETEROGENOUS LIVER. MULTIPLE PUNCTURES TO GET INTO PORTAL VEIN. "GADANI WIRE" PLACED AS A MARKER IN BOTH THE RIGHT AND MIDDLE HEPATIC PORTAL VEIN. JUGULAR ACCESS WITH LIVERTY TIPS SET DURING ACCESS, SOME INJECTIONS SEEMED TO BE WITHIN INTRAHEPATIC HEMATOMA. REFERENCE REPORTS: MW5150887, MW5150888, MW5150890.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 18613177 ↗InjuryJan 30, 2024
THREE OCCURRENCES WHERE THE TIP OF CATHETER/GUIDEWIRE SHEARED OFF INSIDE OF THE PATIENT. ON TWO OF THOSE OCCURRENCES THE INTERVENTIONAL RADIOLOGIST WAS ABLE TO SNARE THE TIP AND RETRIEVE. ON ONE OCCURRENCE THE TIP THAT SHEARED OFF HAD TO BE LEFT EMBEDDED IN THE LIVER PARENCHYMA. DIFFICULT TRANSHEPATIC PORTAL ACCESS DUE TO OBLITERATION OF PERIPORTAL FAT AND HETEROGENOUS LIVER. MULTIPLE PUNCTURES TO GET INTO PORTAL VEIN. "GADANI WIRE" PLACED AS A MARKER IN BOTH THE RIGHT AND MIDDLE HEPATIC PORTAL VEIN. JUGULAR ACCESS WITH LIVERTY TIPS SET DURING ACCESS, SOME INJECTIONS SEEMED TO BE WITHIN INTRAHEPATIC HEMATOMA. REFERENCE REPORTS: MW5150887, MW5150889, MW5150890.
Patient problems: Foreign Body In Patient; Hematoma; Needle Stick/Puncture
MAUDE report 18613161 ↗InjuryJan 30, 2024
THREE OCCURRENCES WHERE THE TIP OF CATHETER/GUIDEWIRE SHEARED OFF INSIDE OF THE PATIENT. ON TWO OF THOSE OCCURRENCES THE INTERVENTIONAL RADIOLOGIST WAS ABLE TO SNARE THE TIP AND RETRIEVE. ON ONE OCCURRENCE THE TIP THAT SHEARED OFF HAD TO BE LEFT EMBEDDED IN THE LIVER PARENCHYMA. DIFFICULT TRANSHEPATIC PORTAL ACCESS DUE TO OBLITERATION OF PERIPORTAL FAT AND HETEROGENOUS LIVER. MULTIPLE PUNCTURES TO GET INTO PORTAL VEIN. "GADANI WIRE" PLACED AS A MARKER IN BOTH THE RIGHT AND MIDDLE HEPATIC PORTAL VEIN. JUGULAR ACCESS WITH LIVERTY TIPS SET DURING ACCESS, SOME INJECTIONS SEEMED TO BE WITHIN INTRAHEPATIC HEMATOMA. REFERENCE REPORTS: MW5150888, MW5150889, MW5150890.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 18613149 ↗InjuryAug 21, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DETACHMENT OCCURRED. THE PATIENT PRESENTED WITH ARRHYTHMIA FOR BIVENTRICULAR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) INSERTION. A 182 CM J LUGE GUIDEWIRE WAS SELECTED FOR USE IN PREPARING THE VESSEL FOR INSERTION OF A PACER LEAD. DURING WIRING USING THE 182CM J LUGE GUIDEWIRE, IT WAS NOTED THAT THE PROXIMAL PART OF THE RADIOPAQUE TIP HAD DETACHED. THE PHYSICIAN DECIDED TO LEAVE THE BROKEN PIECE INSIDE THE PATIENT AND NO REMOVAL ATTEMPTS WERE MADE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT WIRE. THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE. IT WAS FURTHER REPORTED THAT THE TARGET VESSEL WAS IN THE POSTERIOR LATERAL BRANCH OF THE CORONARY SINUS, AND THAT THEY WERE UNABLE TO SAFELY REMOVE THE DEVICE DUE TO ITS LOCATION.
Patient problems: Foreign Body In Patient
MAUDE report 17590906 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page