Starter™ Guidewire M001491181Recalls & reported problems
Lake Region Medical Limited · "J" Curved Fixed Core Guidewire · Class 2
All models of Starter Guidewire (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729201304), received between Nov 6, 2020 and Nov 17, 2024 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Integrity Problem4
- Fracture3
- Material Deformation3
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 17, 2024Fracture; Fracture
EVENT DESCRIPTION: ECN EVENT DESCRIPTION: THE HEAD OF THE HEMODYNAMICS SERVICE REPORTS THAT WHEN THE PRODUCT WAS ROTATED TO THE TABLE FOR PREPARATION, THE TIP OF THE GUIDE WAS FRACTURED. WHAT TROUBLESHOOTING STEPS, IF ANY, RESOLVED THE ISSUE?: THE PROBLEM WAS SOLVED BY USING ANOTHER GUIDE OF THE SAME SIZE. WHAT IS THE NEXT COURSE OF ACTION?: REPLENISH THE PRODUCT PATIENT PRESENT AT TIME OF EVENT?: YES, PATIENT COMPLICATIONS: NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20702653 ↗MalfunctionFeb 2, 2023Fracture; Fracture
EVENT DESCRIPTION: IT WAS REPORTED THAT: DIFFICULT REMOVAL OF THE GUIDEWIRE, WHICH RESULTED IN THE SEVERING OF THE GUIDEWIRE DURING REMOVAL. DURING CHECK-UP POST-PROCEDURE, THE REST OF THE GUIDEWIRE WAS POSITIONED DISTALLY TO THE IN-PLACE DRAIN. ACTIONS TAKEN IN THE HEALTHCARE FACILITY FOR PATIENT MANAGEMENT: THE DRAIN AND THE REST OF THE GUIDEWIRE HAVE BEEN REMOVED FROM THE PATIENT. WHEN REMOVING THE DRAIN: WE HAD TO CONTROL THE REMOVAL OF THE REST OF THE GUIDEWIRE. THE DEVICE WAS NOT CONSERVED.EVENT DATE: (B)(6).
Patient problems: Insufficient Information
MAUDE report 16282942 ↗MalfunctionSep 29, 2021Fracture; Fracture
PER EMAIL, IT WAS REPORTED THAT: STARTER WIRE LOOKS DEFORMED BEFORE USE. HERE IS SOME ADDITIONAL INFORMATION ON THE GUIDEWIRE. 05AUG2021 JLC: RESPONSE RECEIVED FROM GAYATHRY N VADIVAL ON 02AUG2021 WHAT WAS THE ANATOMY LOCATION THAT THE PHYSICIAN WAS WORKING ON? LEFT MAIN TO LAD . PLEASE PROVIDE LESION DETAILS: % STENOSIS? LEFT MAIN 50%, LAD 70%. CALCIFICATION? (NOT CALCIFIED, MILD, MODERATE, SEVERE). CONCENTRIC CALCIUM IN LEFT MAIN, INTIMAL CALCIFICATION AT LAD. TORTUOSITY? (NOT TORTUOUS, MILD, MODERATE, SEVERE). NOT STATED . WHEN IS THE EXPECTED DATE THIS DEVICE WILL BE RETURNED? BY THIS MONTH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12545507 ↗MalfunctionNov 6, 2020Material Deformation; Material Integrity Problem; Material Deformation; Material Integrity Problem
UNUSABLE, KINKED GUIDEWIRE. THE WARD DOCTOR NOTICED THE PROBLEM WHEN INSERTING THE GUIDE. THERE WAS A HITCH, IT WAS NOT GOING THROUGH. THE GUIDEWIRE WOULD HAVE BROKEN AT THE "J" SHAPE EXTREMITY LEVEL. THIS IS A RECURRING PROBLEM, THIS IS THE 4TH OCCURRENCE, ALL GUIDEWIRES FROM THE SAME BATCH. NOTED CONSEQUENCES: THE PROCEDURE TIME WAS LENGTHENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10799798 ↗MalfunctionNov 6, 2020Material Deformation; Material Integrity Problem; Material Deformation; Material Integrity Problem
UNUSABLE, KINKED GUIDEWIRE. THE WARD DOCTOR NOTICED THE PROBLEM WHEN INSERTING THE GUIDE. THERE WAS A HITCH, IT WAS NOT GOING THROUGH. THE GUIDEWIRE WOULD HAVE BROKEN AT THE "J" SHAPE EXTREMITY LEVEL. THIS IS A RECURRING PROBLEM, THIS IS THE 4TH OCCURRENCE, ALL GUIDEWIRES FROM THE SAME BATCH. NOTED CONSEQUENCES: THE PROCEDURE TIME WAS LENGTHENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10799797 ↗MalfunctionNov 6, 2020Material Integrity Problem; Material Integrity Problem
UNUSABLE, KINKED GUIDEWIRE. THE WARD DOCTOR NOTICED THE PROBLEM WHEN INSERTING THE GUIDE. THERE WAS A HITCH, IT WAS NOT GOING THROUGH. THE GUIDEWIRE WOULD HAVE BROKEN AT THE "J" SHAPE EXTREMITY LEVEL. THIS IS A RECURRING PROBLEM, THIS IS THE 4TH OCCURRENCE, ALL GUIDEWIRES FROM THE SAME BATCH. NOTED CONSEQUENCES: THE PROCEDURE TIME WAS LENGTHENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10799796 ↗MalfunctionNov 6, 2020Material Deformation; Material Integrity Problem; Material Deformation; Material Integrity Problem
UNUSABLE, KINKED GUIDEWIRE. THE WARD DOCTOR NOTICED THE PROBLEM WHEN INSERTING THE GUIDE. THERE WAS A HITCH, IT WAS NOT GOING THROUGH. THE GUIDEWIRE WOULD HAVE BROKEN AT THE "J" SHAPE EXTREMITY LEVEL. THIS IS A RECURRING PROBLEM, THIS IS THE 4TH OCCURRENCE, ALL GUIDEWIRES FROM THE SAME BATCH. NOTED CONSEQUENCES: THE PROCEDURE TIME WAS LENGTHENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10799795 ↗
Other Wire, Guide, Catheter devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.