WALLSTENT® Endoprosthesis H965404310Recalls & reported problems
Boston Scientific Corporation · Self-Expanding Stent · Class 2 · Implant
All models of WALLSTENT Endoprosthesis (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729204107), received between Jan 30, 2020 and Jan 25, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Deformation1
- Mechanical Problem1
- Positioning Failure1
Patient problems
- No Patient Involvement2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 25, 2023Positioning Failure; Mechanical Problem; Positioning Failure; Mechanical Problem
IT WAS REPORTED THAT THE DEVICE WAS REMOVED NOT FULLY CONSTRAINED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED BRONCHIAL ARTERY. A 20X55/10FR UNI PLUS 75CM WALLSTENT ENDOPROSTHESIS STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING THE PROCEDURE, THE STENT FAILED TO DEPLOY. THE STENT WAS NOT FULLY CONSTRAINED WHEN REMOVED AND THE TIP WAS OUTSIDE THE SHEATH. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT CONDITION WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16240330 ↗MalfunctionSep 7, 2020Material Deformation; Material Deformation
IT WAS REPORTED THAT STENT DAMAGE OCCURRED. A 20X55/10FR UNI PLUS 75CM WALLSTENT ENDOPROSTHESIS STENT WAS SELECTED FOR USE. HOWEVER, IT WAS NOTICED THAT THE STENT WAS KINKED WHEN IT CAME OUT FROM THE PACKAGE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE.
Patient problems: No Patient Involvement
MAUDE report 10499525 ↗MalfunctionJan 30, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. DURING UNPACKING OF A 20X55/10FR UNI PLUS 75CM WALLSTENT, IT WAS NOTED THAT THE STENT DEPLOYMENT TIP WAS MISSING WHILE THE DISTAL RADIOPAQUE MARKER BAND WAS STILL INTACT. UPON FURTHER INSPECTION, THE MISSING TIP WAS NOT IN THE PACKING EITHER. THE PROCEDURE WAS COMPLETED WITH ANOTHER WALL STENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Patient Involvement
MAUDE report 9646749 ↗
Other Prosthesis, tracheal, expandable devices
- iCAST 85453Atrium Medical Corporation · Prosthesis, tracheal, expandable43 reports
- iCAST 85419Atrium Medical Corporation · Prosthesis, tracheal, expandable13 reports
- WALLSTENT™ BILIARY H965430500Boston Scientific Corporation · Prosthesis, tracheal, expandable13 reports
- iCAST 85445Atrium Medical Corporation · Prosthesis, tracheal, expandable12 reports
- iCAST 85416Atrium Medical Corporation · Prosthesis, tracheal, expandable12 reports
- WALLSTENT® Endoprosthesis H965404130Boston Scientific Corporation · Prosthesis, tracheal, expandable11 reports
- iCAST 85455Atrium Medical Corporation · Prosthesis, tracheal, expandable11 reports
- iCAST 85414Atrium Medical Corporation · Prosthesis, tracheal, expandable10 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page