ChoICE™ H7491211901J1Recalls & reported problems
Boston Scientific Corporation · Guidewire with ICE™ Hydrophilic Coating · Class 2
All models of ChoICE (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729252290), received between Jul 26, 2019 and Nov 7, 2019 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Difficult to Remove2
- Break1
Patient problems
- Device Embedded In Tissue or Plaque2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 7, 2019Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT A CATHETER AND GUIDEWIRE ENTRAPMENT OCCURRED. AFTER A 300 ES J CHOICE GUIDEWIRE CROSSED THE LESION, A 10.0-31 CAROTID WALLSTENT WAS ADVANCED FOR TREATMENT. HOWEVER, WHEN THE STENT WAS ATTEMPTED TO BE RELEASED, IT WAS NOTED THAT THE GUIDEWIRE STUCK WITHIN THE STENT SYSTEM AND COULD NOT BE RELEASED. THE WALLSTENT DELIVERY SYSTEM AND THE CHOICE GUIDEWIRE WERE REMOVED TOGETHER AND THE PROCEDURE WAS COMPLETED WITH ANOTHER WALLSTENT AND ANOTHER GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT'S CONDITION WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9291578 ↗MalfunctionJul 26, 2019Break; Detachment of Device or Device Component
DURING PCI OF THE CIRCUMFLEX ARTERY, WHILE PHYSICIAN WAS ATTEMPTING TO PLACE PCI WIRE ACROSS THE STENOTIC LESION THE PROXIMAL FLEXIBLE PORTION OF THE WIRE TIP BROKE OFF AND REMAINED IN THE CIRCUMFLEX ARTERY. MULTIPLE ATTEMPTS WERE MADE TO RETRIEVE THE BROKEN WIRE FROM THE VESSEL BUT ALL FAILED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8836435 ↗InjuryJul 26, 2019Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component
IT WAS REPORTED THAT GUIDE WIRE TIP DETACHMENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT CIRCUMFLEX CORONARY ARTERY. A 300CM J CHOICE EXTRA SUPPORT GUIDEWIRE WAS ADVANCED TO CROSS THE LESION. HOWEVER, IT WAS NOTED THAT BALLOON COULD NOT PASS ACROSS THAT WIRE. PHYSICIAN WANTED TO USE A DIFFERENT WIRE AND PULLED THE WIRE OUT. WHEN DEVICE WAS REMOVED, IT WAS NOTED THAT THE WIRE CAUGHT IN THE VESSEL AND THE TIP PORTION WAS FRACTURED AND BROKE OFF INSIDE THE PATIENT. THE PHYSICIAN PULLED THE REST OF THE REMAINING OF THE WIRE OUT, BUT PART OF THE WIRE STILL STAYED INSIDE THE PATIENT. AN ATTEMPT TO SNARE THE DETACHED PART OF THE TIP WAS DONE, BUT IT WAS UNSUCCESSFUL. THE DETACHED PART OF THE TIP WAS LEFT INSIDE THE PATIENT'S BODY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS FINE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8831502 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.