ChoICE™ H7491213201J1Recalls & reported problems

Boston Scientific Corporation · Guidewire with ICE™ Hydrophilic Coating · Class 2

All models of ChoICE (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 08714729252351), received between Aug 21, 2019 and Feb 2, 2024 — classified as 4 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 3 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component7
  • Fracture3
  • Device-Device Incompatibility1
  • Difficult to Remove1
  • Entrapment of Device1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Device Embedded In Tissue or Plaque2
  • Foreign Body In Patient1
  • No Patient Involvement1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 2, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE DEVICE FRACTURED. UPON OPENING THE PACKAGE, A CHOICE GUIDEWIRE APPEARED "FRAYED." THE DEVICE WAS NOT USED FOR THE PROCEDURE. NO PATIENT CONTACT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18634330 ↗
  • MalfunctionFeb 2, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE DEVICE FRACTURED. UPON OPENING THE PACKAGE, A CHOICE GUIDEWIRE APPEARED "FRAYED." THE DEVICE WAS NOT USED FOR THE PROCEDURE. NO PATIENT CONTACT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18634261 ↗
  • InjuryFeb 2, 2024Detachment of Device or Device Component; Device-Device Incompatibility; Detachment of Device or Device Component; Device-Device Incompatibility

    IT WAS REPORTED THAT THE DEVICE DETACHED REQUIRING ADDITIONAL INTERVENTION. A CHOICE GUIDEWIRE WAS ADVANCED AND BECAME STUCK UNDER A STENT. DURING ATTEMPTED REMOVAL, THE GUIDEWIRE "FRAYED" AND BROKE. THE GUIDEWIRE IS BROKEN IN THE PATIENT. THREE STENTS WERE PLACED CRUSHING THE DISTAL STENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 18632603 ↗
  • MalfunctionNov 16, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DISTAL TIP FRACTURE OCCURRED. DURING PREPARATION OF A 182CM J CHOICE FLOPPY GUIDEWIRE, THE TIP WAS FOUND TO BE FRACTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18152169 ↗
  • InjuryJun 16, 2022Use of Device Problem; Detachment of Device or Device Component; Use of Device Problem; Detachment of Device or Device Component

    IT WAS REPORTED THAT DISTAL TIP DETACHMENT OCCURRED. A 182CM J (5PK) CHOICE GUIDEWIRE WAS USED IN A RIGHT HEART CATHERIZATION PROCEDURE. HOWEVER, IT WAS NOTED THAT THE RADIOPAQUE PORTION OF THE WIRE BROKE OFF. THE DETACHED PORTION WAS SUCCESSFULLY SNARED AND REMOVED FROM THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14716812 ↗
  • MalfunctionFeb 23, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT GUIDEWIRE DETACHMENT OCCURRED. THE PATIENT PRESENTED WITH CORONARY ARTERIAL OCCLUSION. A 182CM J CHOICE GUIDEWIRE WAS SELECTED FOR USE. HOWEVER, DURING INTRODUCTION, THE DEVICE WAS FRACTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13588882 ↗
  • MalfunctionJul 27, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT GUIDEWIRE TIP DETACHMENT OCCURRED. A 182CM J CHOICE GUIDEWIRE WAS SELECTED FOR USE; HOWEVER, UPON REMOVAL FROM THE PROTECTIVE CASE, IT WAS NOTICED THAT THE GUIDEWIRE TIP WAS DAMAGED AND BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12233197 ↗
  • InjuryDec 11, 2019Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    IT WAS REPORTED THAT DETACHMENT OF DEVICE OCCURRED. THE TARGET LESION WAS LOCATED IN THE CORONARY ARTERY. A FLOPPY 182CM J CHOICE GUIDEWIRE WAS USED FOR TREATMENT. HOWEVER, DURING PROCEDURE, ENTRAPMENT OCCURRED WHEN REMOVING THE DEVICE. THE GUIDEWIRE TIP FRACTURE REMAINED INSIDE THE PATIENT'S BODY.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9450499 ↗
  • InjuryDec 9, 2019Entrapment of Device; Detachment of Device or Device Component; Entrapment of Device; Detachment of Device or Device Component

    IT WAS REPORTED THAT DETACHMENT OF DEVICE OCCURRED. THE TARGET LESION WAS LOCATED IN THE CORONARY ARTERY. A FLOPPY 182CM J CHOICE GUIDEWIRE WAS USED FOR TREATMENT. HOWEVER, DURING PROCEDURE, ENTRAPMENT OCCURRED WHEN REMOVING THE DEVICE. THE GUIDEWIRE TIP FRACTURE REMAINED INSIDE THE PATIENT'S BODY.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9438400 ↗
  • MalfunctionAug 21, 2019Fracture; Fracture

    IT WAS REPORTED THAT GUIDE WIRE TIP FRACTURE OCCURRED. DURING UNPACKING OF A 182CM CHOICE GUIDE WIRE, A FRACTURE OF THE TIP WAS NOTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Patient Involvement

    MAUDE report 8914640 ↗