ChoICE™ H74912133011Recalls & reported problems

Boston Scientific Corporation · Guidewire with ICE™ Hydrophilic Coating · Class 2

All models of ChoICE (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729252368), received between Jul 25, 2022 and Aug 6, 2026 — classified as 1 death, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Difficult to Remove1
  • Failure to Advance1
  • Material Integrity Problem1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2026Nonstandard Device; Difficult to Remove

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50162963 ↗
  • DeathJul 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE DEVICE DETACHED AND THE PATIENT DIED. THE PATIENT PRESENTED WITH CORONARY ARTERY DISEASE. DURING PERCUTANEOUS INTERVENTION, A 182CM CHOICE GUIDEWIRE WAS ADVANCED AND THE TIP OF THE GUIDEWIRE BROKE OFF IN THE CIRCUMFLEX ARTERY. INSTEAD OF REMOVING THE 6MM DEVICE FRAGMENT, THE PHYSICIAN DECIDED TO PRESS A PORTION OF THE GUIDEWIRE TIP TO THE VESSEL WALL USING A STENT. THE ENTIRE LENGTH OF THE DEVICE FRAGMENT COULD NOT BE STENTED TO THE VESSEL WALL. THE PATIENT REMAINED STABLE DURING THE PROCEDURE. THE FOLLOWING DAY, DIAGNOSTIC CORONARY ANGIOGRAPHY REVEALED THAT THE LUMEN OF THE ARTERY WAS NORMAL, BLOOD FLOW WAS NOT LIMITED. THE PATIENT WAS TRANSFERRED TO ANOTHER HOSPITAL FOR ADDITIONAL INTERVENTION. TWO ADDITIONAL STENTS WERE DEPLOYED IN THE CIRCUMFLEX PRESSING THE REMAINDER OF THE GUIDEWIRE TIP AGAINST THE VESSEL WALL. THROMBOSIS THEN OCCURRED RESULTING IN THE DEATH OF THE PATIENT.

    Patient problems: Foreign Body In Patient; Thrombosis/Thrombus

    MAUDE report 19745390 ↗
  • MalfunctionJul 25, 2022Failure to Advance; Material Integrity Problem; Failure to Advance; Material Integrity Problem

    IT WAS REPORTED THAT COATING ISSUE OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND NON-CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 182CM (5PK) CHOICE GUIDEWIRE WAS SELECTED FOR USE. HOWEVER, RESISTANCE WAS ENCOUNTERED WHEN THE DEVICE WAS TRIED TO CROSS THE LESION AND THE TIP OF THE WIRE WAS PEELED OFF. THE PROCEDURE WAS COMPLETED WITH ANOTHER WIRE OF SAME BRAND. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15094893 ↗