Skip to content

Amplatz Super Stiff™ M001465000Recalls & reported problems

Boston Scientific Corporation · Guidewire · Class 2

All models of Amplatz Super Stiff (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729295631), received between Mar 8, 2023 and Jan 31, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Difficult to Remove1
  • Material Deformation1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 31, 2025Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE GUIDEWIRE SEPARATED. AN AMPLATZ SUPER STIFF GUIDEWIRE WAS SELECTED FOR USE IN A BILIARY DRAINAGE PROCEDURE. AN 8.5 FRENCH NON-BOSTON SCIENTIFIC DRAINAGE CATHETER FOR INTERNAL AND EXTERNAL BILIARY DRAINAGE WAS USED, EQUIPPED WITH A SOFT MANDRIN. AFTER CORRECTLY POSITIONING THE DRAIN, AN ISSUE AROSE WHILE ATTEMPTING TO SIMULTANEOUSLY PULL OUT THE SOFT MANDRIN AND THE AMPLATZ SUPER STIFF GUIDEWIRE. RESISTANCE WAS ENCOUNTERED DURING REMOVAL, BUT THE USER CONTINUED TO PULL THE MANDRIN AND THE GUIDEWIRE OUT TOGETHER. UPON EXTRACTING THE MANDRIN AND GUIDEWIRE, IT WAS OBSERVED THAT THE SOFT MANDRIN HAD STRETCHED DURING THE REMOVAL PROCESS. ULTIMATELY, A PORTION OF THE MANDRIN WAS LEFT INSIDE THE DRAINAGE CATHETER, AND THE AMPLATZ SUPER STIFF GUIDEWIRE HAD DISINTEGRATED, SEPARATING INTO TWO PIECES BUT REMAINING CONNECTED AT THE MIDPOINT. BOTH DEVICES WERE SUCCESSFULLY RETRIEVED FROM THE DRAIN. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT'S CONDITION WAS STABLE FOLLOWING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21286860 ↗
  • MalfunctionMar 8, 2023Unraveled Material; Detachment of Device or Device Component; Material Deformation; Unraveled Material; Detachment of Device or Device Component; Material Deformation

    IT WAS REPORTED THAT CORE WIRE DETACHMENT OCCURRED. THE 50% STENOSED TARGET LESION WAS LOCATED IN THE NON- TORTUOUS AND NON-CALCIFIED FEMORAL ARTERY. A 035/145/3MM AMPLATZ SUPER STIFF GUIDEWIRE WAS ADVANCED, HOWEVER, IT WAS FOUND UNDER FLUOROSCOPY THAT THE TIP WAS SERIOUSLY KINKED. THE PHYSICIAN PULLED THE WIRE BACK OUT OF THE SHEATH AND INSPECTED IT. THE OUTER SUPPORT COIL WAS LOOSENED, AND THE INNER CORE WIRE OF AMPLATZ WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16507247 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources and updates