Skip to content

Expo™ H74908526891Recalls & reported problems

Boston Scientific Corporation · Angiographic Catheter · Class 2

All models of Expo (47)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729366768), received between Oct 6, 2021 and Oct 6, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 21, 2024Class IOpen, Classified

    An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.

    Z-1682-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Difficult to Remove1
  • Entrapment of Device1

Patient problems

  • Fistula1
  • Perforation of Vessels1
  • Pseudoaneurysm1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 6, 2021Entrapment of Device; Difficult to Remove; Detachment of Device or Device Component; Entrapment of Device; Difficult to Remove; Detachment of Device or Device Component

    IT WAS REPORTED VIA MEDWATCH MW5103876 THAT CATHETER ENTRAPMENT AND DETACHMENT OCCURRED. THE PATIENT PRESENTED FOR AN OUTPATIENT CARDIAC CATHETERIZATION PROCEDURE. A NON-BOSTON SCIENTIFIC SHEATH AND A 5F EXPO GUIDE CATHETER WERE INSERTED WITH NO ISSUES OR CONCERNS. WHEN THE PHYSICIAN TRIED TO REMOVE THE CATHETER, IT BECAME STUCK IN THE SHEATH AND CAME APART. THE PIECES WERE SUCCESSFULLY REMOVED AND IN THE PROCESS A VESSEL WAS PERFORATED. THE PATIENT VESSELS WERE TORTUOUS WHICH CONTRIBUTED TO THE COMPLICATION. MANUAL PRESSURE WAS APPLIED AND BLEEDING STOPPED. THE PATIENT EXPERIENCED FOREARM PAIN AND SWELLING. THE PATIENT WAS ADMITTED AND PROVIDED PAIN CONTROL, A REPEAT ULTRASOUND AND CARDIOLOGY CARE. THE PATIENT EXPERIENCED A RADIAL ARTERY PSEUDOANEURYSM AND ARTERIOVENOUS (AV) FISTULA CONSISTENT WITH A PERFORATED VESSEL.

    Patient problems: Fistula; Perforation of Vessels; Pseudoaneurysm

    MAUDE report 12586588 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page