CyberWand USL Generator CW-USLGRecalls & reported problems
Cybersonics Inc · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI H891CWUSLG2), received between May 6, 2022 and May 6, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Output Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 6, 2022Output Problem; Output Problem
IT WAS REPORTED THAT WHEN THE CW-USLTA WAS ACTIVATED WITH THE CYBERWAND GENERATOR, THE CABLE CAUGHT FIRE AND FLAMES SHOT OUT OF THE END OF THE HANDPIECE. CUSTOMER SITE FACILITY BIOMED CONFIRMED THAT THE HANDPIECE WAS NOT IN THE PATIENT AT THE TIME OF THE INCIDENT, THE USER HAD THE HANDPIECE IN HIS HAND TESTING THE HANDPIECE. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THE INTENDED PROCEDURE ACCORDING TO THE CUSTOMER COMPLETED WITH ANOTHER DEVICE. CUSTOMER SITE BIOMED CONFIRMED THAT THE FIRE REPORTED TO HAVE COME FROM THE TRANSDUCER TIP. IT WAS NOT FROM THE GENERATOR AND THERE WAS NO PATIENT INJURY. THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED. THIS REPORT IS RELATED TO A REPORT WITH PATIENT IDENTIFIER (B)(6) .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14312431 ↗
Other Lithotriptor, Electro-Hydraulic devices
- Lithotripsy Probe 1.9FR 375cm DS M00546620Northgate Technologies Inc. · Lithotriptor, Electro-Hydraulic1 recall1 death19 reports
- Autolith Touch M005466800Northgate Technologies Inc. · Lithotriptor, Electro-Hydraulic1 recall12 reports
- Ultrasound Probe M0068407170Boston Scientific Corporation · Lithotriptor, electro-hydraulic15 reports
- CyberWand USL Ultrasonic Lithotripsy System CW-USLSACybersonics Inc · Lithotriptor, Electro-Hydraulic4 reports
- Ultrasound Probe M0068407160Boston Scientific Corporation · Lithotriptor, electro-hydraulic3 reports
- CyberWand USL Advanced Transducer CW-USLTACybersonics Inc · Lithotriptor, Electro-Hydraulic2 reports
- CyberWand USL NoseCone Kit CW-USLNCCybersonics Inc · Lithotriptor, Electro-Hydraulic1 report
- CyberWand USL Advanced Transducer CW-USLTACybersonics Inc · Lithotriptor, Electro-Hydraulic1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page