Skip to content

CyberWand USL Generator CW-USLGRecalls & reported problems

Cybersonics Inc · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI H891CWUSLG2), received between May 6, 2022 and May 6, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Output Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 6, 2022Output Problem; Output Problem

    IT WAS REPORTED THAT WHEN THE CW-USLTA WAS ACTIVATED WITH THE CYBERWAND GENERATOR, THE CABLE CAUGHT FIRE AND FLAMES SHOT OUT OF THE END OF THE HANDPIECE. CUSTOMER SITE FACILITY BIOMED CONFIRMED THAT THE HANDPIECE WAS NOT IN THE PATIENT AT THE TIME OF THE INCIDENT, THE USER HAD THE HANDPIECE IN HIS HAND TESTING THE HANDPIECE. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THE INTENDED PROCEDURE ACCORDING TO THE CUSTOMER COMPLETED WITH ANOTHER DEVICE. CUSTOMER SITE BIOMED CONFIRMED THAT THE FIRE REPORTED TO HAVE COME FROM THE TRANSDUCER TIP. IT WAS NOT FROM THE GENERATOR AND THERE WAS NO PATIENT INJURY. THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED. THIS REPORT IS RELATED TO A REPORT WITH PATIENT IDENTIFIER (B)(6) .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14312431 ↗

Other Lithotriptor, Electro-Hydraulic devices

All Lithotriptor, Electro-Hydraulic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page