V-18™ Control Wire™ M001468480Recalls & reported problems

Boston Scientific Corporation · Guidewire with ICE™ Hydrophilic coating · Class 2

All models of V-18 Control Wire (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729715566), received between Jan 6, 2021 and Jan 10, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Entrapment of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Entrapment of Device2
  • Material Integrity Problem1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 10, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT TIP DETACHMENT OCCURRED AND IT WAS REMAINED INSIDE THE PATIENT'S BODY. THE PATIENT UNDERWENT COMPLEX RECANALIZATION/THROMBOLYSIS CATHETER PLACEMENT FOR A THROMBOSED LEFT LOWER EXTREMITY FEMORAL POPLITEAL. A 110CM, 8CM POLY TIP V-18 CONTROL WIRE WAS SELECTED FOR USE. DURING THE PROCESS OF ACCESSING A TIGHT CHANNEL FOR PLACING A CATHETER, IT WAS NOTED THAT A PORTION OF THE WIRE'S TIP SHEARED OFF THE NEEDLE AND WAS RETAINED WITHIN A BRANCH OF THE PATIENT'S PROFUNDAL ARTERY. MINIMAL EFFECT WAS NOTED ON PATIENT'S CARE AND THE PATIENT WAS ABLE TO GET A BYPASS THE FOLLOWING DAYS UPON A PRESUMED DIFFERENT/LARGER PROFUNDAL BRANCH. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 18487313 ↗
  • MalfunctionNov 23, 2022Entrapment of Device; Entrapment of Device

    IT WAS REPORTED THAT GUIDEWIRE ENTRAPMENT OCCURRED. THE PATIENT PRESENTED WITH VASCULAR ACCESS FAILURE AND UNDERWENT VASCULAR ACCESS INTERVENTION THERAPY. THE 99% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN A MILDLY TORTUOUS VESSEL. A 110CM, 8CM V-18 CONTROL WIRE GUIDEWIRE WAS ADVANCED. HOWEVER, DURING REPEATED EXPANSION, THE BALLOON AND GUIDEWIRE GOT STUCK AROUND THE HUB. THE DEVICE WAS REMOVED TOGETHER WITH THE BALLOON AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15853298 ↗
  • MalfunctionJan 6, 2021Entrapment of Device; Entrapment of Device

    IT WAS REPORTED THAT GUIDEWIRE ENTRAPMENT OCCURRED. A 110CM, 8CM V-18 CONTROL WIRE GUIDEWIRE WAS SELECTED FOR A PROCEDURE IN THE RADIAL CUTANEOUS VEIN. DURING THE PROCEDURE, A BALLOON CATHETER BECAME STUCK ON THE WIRE AND COULD NOT MOVE BACK AND FORTH. THE ENTIRE GUIDEWIRE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER V-18 WIRE AND BALLOON CATHETER. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11127220 ↗