IMAGER™ II M001314580Recalls & reported problems
Boston Scientific Corporation · Angiographic Catheter · Class 2
All models of IMAGER II (47)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729736608), received between Nov 25, 2019 and Dec 19, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Aug 24, 2026Open, Classified
Tip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include prolonged procedure or life-threatening embolism.
Z-3214-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Detachment of Device or Device Component2
- Material Integrity Problem1
Patient problems
- No Patient Involvement2
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryDec 19, 2025Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ADDITIONAL INTERVENTION WAS REQUIRED TO REMOVE A DETACHED TIP. AN IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FOR USE IN EMBOLIZATION OF THE RENAL ARTERY. THE CONDITION OF THE VESSEL WAS A STRAIGHTFORWARD RENAL ARTERY. WHILE INSERTING THE CATHETER INTO THE PATIENT, THE DISTAL TIP DETACHED. THERE WAS NO RESISTANCE NOTED DURING USE OF THE DEVICE. THE TIP MIGRATED TOWARD THE AORTA. AN INITIAL ATTEMPT AT RETRIEVAL WAS UNSUCCESSFUL BECAUSE THE TIP HAD BECOME ADHERED TO THE AORTIC WALL. A NEW VASCULAR ACCESS SITE WAS CREATED, AND A NEW SHEATH WAS USED FROM THE NEW SITE. WHEN THE DETACHED TIP EVENTUALLY MOVED, IT WAS REMOVED FROM THE LOWER PART OF THE LEG VIA SNARE. WHEN THE TIP WAS EXTRACTED FROM THE SHEATH, IT HAD BROKEN INTO SMALL FRAGMENTS. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THE EVENT. THE PATIENT WAS STABLE FOLLOWING THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23853499 ↗MalfunctionSep 1, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AN UNUSED 5F CONTRA IMAGER II ANGIOGRAPHIC CATHETER IN IT'S UNOPENED ORIGINAL PACKAGING WAS RETURNED TO BOSTON SCIENTIFIC FROM A CUSTOMER. THERE WAS NO REPORTED COMPLAINT AGAINST THIS RETURNED DEVICE. THE DEVICE WAS INTACT UPON RECEIPT. BOSTON SCIENTIFIC CARRIED OUT FUNCTIONAL TESTING ON THE RETURNED DEVICE WHICH RESULTED IN A TIP DETACHMENT. THIS EVENT IS ASSOCIATED WITH PREVIOUSLY REPORTED FIELD ACTION 92484513-FA WITH PRODUCT BEING REMOVED FROM THE FIELD.
Patient problems: No Patient Involvement
MAUDE report 10475944 ↗MalfunctionAug 31, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AN UNUSED 5F CONTRA IMAGER II ANGIOGRAPHIC CATHETER IN IT'S UNOPENED ORIGINAL PACKAGING WAS RETURNED TO BOSTON SCIENTIFIC FROM A CUSTOMER. THERE WAS NO REPORTED COMPLAINT AGAINST THIS RETURNED DEVICE. THE DEVICE WAS INTACT UPON RECEIPT. BOSTON SCIENTIFIC CARRIED OUT FUNCTIONAL TESTING ON THE RETURNED DEVICE WHICH RESULTED IN A TIP DETACHMENT. THIS EVENT IS ASSOCIATED WITH PREVIOUSLY REPORTED FIELD ACTION (B)(4) WITH PRODUCT BEING REMOVED FROM THE FIELD.
Patient problems: No Patient Involvement
MAUDE report 10474505 ↗MalfunctionNov 25, 2019Detachment of Device or Device Component; Material Integrity Problem; Detachment of Device or Device Component; Material Integrity Problem
IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. AN IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FOR USE. THE CATHETER WAS PLACED OVER A NON-BSC GUIDEWIRE. AS THE CATHETER TIP WAS STRAIGHTENED IT SNAPPED AT THE APEX OF THE TIP ANGLE. ANOTHER IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FROM THE SAME LOT NUMBER. PRIOR TO USE, THE CATHETER TIP WAS STRAIGHTENED MANUALLY TO TEST IF TIP WOULD BREAK. THE TIP REMAINED INTACT, AND ANGIOGRAPHIC CATHETER WAS PLACED OVER NON-BSC GUIDEWIRE. AS THE CATHETER TIP WAS STRAIGHTENED IT SNAPPED AGAIN AT THE APEX OF THE TIP ANGLE. ANOTHER IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FROM A DIFFERENT LOT NUMBER. THE CATHETER WAS SUCCESSFULLY LOADED ONTO THE NON-BSC GUIDEWIRE AND THE PROCEDURE WAS COMPLETED. UPON REMOVAL OF THE CATHETER, WHICH HAD A NON-BSC MICROCATHETER WITHIN THE DEVICE, IT WAS NOTICED ON FLUOROSCOPY THAT THE TIP OF THE IMAGER CATHETER HAD AGAIN SNAPPED. THE TIP WAS FREE FLOATING AROUND THE TIP OF THE NON-BSC MICROCATHETER. THE IMAGER II CATHETER AND MICROCATHETER WERE REMOVED SUCCESSFULLY FROM THE SHEATH AND PATIENT. AFTER THE PROCEDURE THE SCRUB NURSE STATED THAT THE CATHETER TIPS FELT BRITTLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9370615 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page