IMAGER™ II M001314590Recalls & reported problems
Boston Scientific Corporation · Angiographic Catheter · Class 2
All models of IMAGER II (47)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729736615), received between Nov 25, 2019 and Jan 8, 2020 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Material Integrity Problem2
Patient problems
- No Consequences Or Impact To Patient2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJan 8, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT TIP DETACHMENT OCCURRED INSIDE THE PATIENT. THE TARGET LESION WAS LOCATED IN THE ILIAC ARTERY. AN IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FOR USE DURING A SUPERFICIAL FEMORAL ARTERY INTERVENTION. HOWEVER, THE TIP OF THE CATHETER BROKE OFF INSIDE THE PATIENT AND IN THE PROCESS OF SNARING, THE TIP WAS CUT INTO HALF. THE LARGEST PORTION OF THE CATHETER TIP WAS REMOVED, WHILE A SMALL PORTION BECAME EMBEDDED IN THE ILIAC ARTERIAL WALL. SUBSEQUENTLY, A STENT WAS USED TO TRAP THE EMBEDDED TIP IN THE ILIAC PORTION. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9565694 ↗MalfunctionNov 25, 2019Detachment of Device or Device Component; Material Integrity Problem; Detachment of Device or Device Component; Material Integrity Problem
IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. AN IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FOR USE. THE CATHETER WAS PLACED OVER A NON-BSC GUIDEWIRE. AS THE CATHETER TIP WAS STRAIGHTENED IT SNAPPED AT THE APEX OF THE TIP ANGLE. ANOTHER IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FROM THE SAME LOT NUMBER. PRIOR TO USE, THE CATHETER TIP WAS STRAIGHTENED MANUALLY TO TEST IF TIP WOULD BREAK. THE TIP REMAINED INTACT, AND ANGIOGRAPHIC CATHETER WAS PLACED OVER NON-BSC GUIDEWIRE. AS THE CATHETER TIP WAS STRAIGHTENED IT SNAPPED AGAIN AT THE APEX OF THE TIP ANGLE. ANOTHER IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FROM A DIFFERENT LOT NUMBER. THE CATHETER WAS SUCCESSFULLY LOADED ONTO THE NON-BSC GUIDEWIRE AND THE PROCEDURE WAS COMPLETED. UPON REMOVAL OF THE CATHETER, WHICH HAD A NON-BSC MICROCATHETER WITHIN THE DEVICE, IT WAS NOTICED ON FLUOROSCOPY THAT THE TIP OF THE IMAGER CATHETER HAD AGAIN SNAPPED. THE TIP WAS FREE FLOATING AROUND THE TIP OF THE NON-BSC MICROCATHETER. THE IMAGER II CATHETER AND MICROCATHETER WERE REMOVED SUCCESSFULLY FROM THE SHEATH AND PATIENT. AFTER THE PROCEDURE THE SCRUB NURSE STATED THAT THE CATHETER TIPS FELT BRITTLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9370638 ↗MalfunctionNov 25, 2019Detachment of Device or Device Component; Material Integrity Problem; Detachment of Device or Device Component; Material Integrity Problem
IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. AN IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FOR USE. THE CATHETER WAS PLACED OVER A NON-BSC GUIDEWIRE. AS THE CATHETER TIP WAS STRAIGHTENED IT SNAPPED AT THE APEX OF THE TIP ANGLE. ANOTHER IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FROM THE SAME LOT NUMBER. PRIOR TO USE, THE CATHETER TIP WAS STRAIGHTENED MANUALLY TO TEST IF TIP WOULD BREAK. THE TIP REMAINED INTACT, AND ANGIOGRAPHIC CATHETER WAS PLACED OVER NON-BSC GUIDEWIRE. AS THE CATHETER TIP WAS STRAIGHTENED IT SNAPPED AGAIN AT THE APEX OF THE TIP ANGLE. ANOTHER IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FROM A DIFFERENT LOT NUMBER. THE CATHETER WAS SUCCESSFULLY LOADED ONTO THE NON-BSC GUIDEWIRE AND THE PROCEDURE WAS COMPLETED. UPON REMOVAL OF THE CATHETER, WHICH HAD A NON-BSC MICROCATHETER WITHIN THE DEVICE, IT WAS NOTICED ON FLUOROSCOPY THAT THE TIP OF THE IMAGER CATHETER HAD AGAIN SNAPPED. THE TIP WAS FREE FLOATING AROUND THE TIP OF THE NON-BSC MICROCATHETER. THE IMAGER II CATHETER AND MICROCATHETER WERE REMOVED SUCCESSFULLY FROM THE SHEATH AND PATIENT. AFTER THE PROCEDURE THE SCRUB NURSE STATED THAT THE CATHETER TIPS FELT BRITTLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9370608 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page