IMAGER™ II M001314670Recalls & reported problems
Boston Scientific Corporation · Angiographic Catheter · Class 2
All models of IMAGER II (47)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729736691), received between Jul 22, 2020 and Oct 13, 2022 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Aug 24, 2026Class IOpen, Classified
Tip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include prolonged procedure or life-threatening embolism.
Z-3214-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component7
Patient problems
- No Consequences Or Impact To Patient3
- No Patient Involvement3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryOct 13, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA THE BRACHIAL ARTERY. THE DE NOVO TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS ARTERIOVENOUS FISTULA IN THE RIGHT ARM. THE IMAGER II/5/C2/65/035 CATHETER WAS SELECTED FOR USE. DURING AN ATTEMPT TO CANNULATE THE VESSEL, THE TIP OF THE CATHETER BROKE. THERE WAS NO RESISTANCE FELT AND NO INTERACTION WITH ANOTHER DEVICE. THE DETACHED TIP WAS REMOVED THROUGH SURGERY. THE PROCEDURE WAS CANCELLED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AND PATIENT'S CONDITION FOLLOWING PROCEDURE WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15600696 ↗MalfunctionSep 23, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
THE HOSPITAL PREVIOUSLY EXPERIENCED TIP DETACHMENTS WITH THREE 5FR IMAGER II ANGIOGRAPHIC CATHETERS CAPTURED WITHIN MANUFACTURER REPORT #S 2134265-2018-03728, 2134265-2018-04054, AND 2134265-2018-04096. THE PHYSICIAN AND HIS TEAM TOOK THIS 5FR IMAGER II ANGIOGRAPHIC CATHETER OF THE SAME BATCH AND PURPOSELY BROKE THE TIP OF THE CATHETER TO PROVE THE FRAGILITY OF THE DEVICE. THE CATHETER WAS NOT USED ON A PATIENT. THIS EVENT IS ASSOCIATED WITH PREVIOUSLY REPORTED FIELD ACTION (B)(4) WITH PRODUCT BEING REMOVED FROM THE FIELD.
Patient problems: No Patient Involvement
MAUDE report 10577659 ↗MalfunctionSep 23, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
THE HOSPITAL PREVIOUSLY EXPERIENCED TIP DETACHMENTS WITH THREE 5FR IMAGER II ANGIOGRAPHIC CATHETERS CAPTURED WITHIN MANUFACTURER REPORT #S (B)(4). THE PHYSICIAN AND HIS TEAM TOOK THIS 5FR IMAGER II ANGIOGRAPHIC CATHETER OF THE SAME BATCH, AND PURPOSELY BROKE THE TIP OF THE CATHETER TO PROVE THE FRAGILITY OF THE DEVICE. THE CATHETER WAS NOT USED ON A PATIENT. THIS EVENT IS ASSOCIATED WITH PREVIOUSLY REPORTED FIELD ACTION 92484513-FA WITH PRODUCT BEING REMOVED FROM THE FIELD.
Patient problems: No Patient Involvement
MAUDE report 10577655 ↗MalfunctionSep 23, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
THE HOSPITAL PREVIOUSLY EXPERIENCED TIP DETACHMENTS WITH THREE 5FR IMAGER II ANGIOGRAPHIC CATHETERS CAPTURED WITHIN MANUFACTURER REPORT #S 2134265-2018-03728, 2134265-2018-04054, AND 2134265-2018-04096. THE PHYSICIAN AND HIS TEAM TOOK THIS 5FR IMAGER II ANGIOGRAPHIC CATHETER OF THE SAME BATCH AND PURPOSELY BROKE THE TIP OF THE CATHETER TO PROVE THE FRAGILITY OF THE DEVICE. THE CATHETER WAS NOT USED ON A PATIENT. THIS EVENT IS ASSOCIATED WITH PREVIOUSLY REPORTED FIELD ACTION 92484513-FA WITH PRODUCT BEING REMOVED FROM THE FIELD.
Patient problems: No Patient Involvement
MAUDE report 10577654 ↗MalfunctionJul 22, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE TIP BROKE OFF. A 5F C2 IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FOR A URETERIC STENT INSERTION PROCEDURE. AN ANTEGRADE APPROACH WAS USED VIA THE LEFT NEPHROSTOMY TUBE. THE IMAGER II CATHETER WAS PLACED INTO THE ENTRY SHEATH AFTER BEING FLUSHED WITH HEPARINIZED SALINE AND LUBRICATED WITH NORMAL SALINE. WHEN THE CATHETER WAS IN THE LEFT URETER, IT WAS DISCOVERED THAT THE DISTAL TIP WAS BROKEN OFF BUT STILL ATTACHED TO THE CATHETER. THE CATHETER WAS REMOVED AND DISCARDED. ANOTHER OF THE SAME PRODUCT AND BATCH NUMBER WAS OPENED. DURING INSPECTION OF THE CATHETER, WHICH INVOLVED FLUSHING AND LUBRICATING, THE DISTAL TIP OF THE CATHETER DETACHED IN THE RADIOLOGIST'S HAND. A THIRD CATHETER WAS THEN OPENED TO CHECK THE TIP. THE TIP WAS VERY FRAGILE AND EASILY REMOVED FROM THE CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS STABLE POST-PROCEDURE. THE PATIENT WAS ADMITTED FOR ROUTINE OBSERVATIONS POST-PROCEDURE AS PRE-PLANNED, AND NO ESCALATION OF CARE WAS REQUIRED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10308661 ↗MalfunctionJul 22, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE TIP BROKE OFF. A 5F C2 IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FOR A URETERIC STENT INSERTION PROCEDURE. AN ANTEGRADE APPROACH WAS USED VIA THE LEFT NEPHROSTOMY TUBE. THE IMAGER II CATHETER WAS PLACED INTO THE ENTRY SHEATH AFTER BEING FLUSHED WITH HEPARINIZED SALINE AND LUBRICATED WITH NORMAL SALINE. WHEN THE CATHETER WAS IN THE LEFT URETER, IT WAS DISCOVERED THAT THE DISTAL TIP WAS BROKEN OFF BUT STILL ATTACHED TO THE CATHETER. THE CATHETER WAS REMOVED AND DISCARDED. ANOTHER OF THE SAME PRODUCT AND BATCH NUMBER WAS OPENED. DURING INSPECTION OF THE CATHETER, WHICH INVOLVED FLUSHING AND LUBRICATING, THE DISTAL TIP OF THE CATHETER DETACHED IN THE RADIOLOGIST'S HAND. A THIRD CATHETER WAS THEN OPENED TO CHECK THE TIP. THE TIP WAS VERY FRAGILE AND EASILY REMOVED FROM THE CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS STABLE POST-PROCEDURE. THE PATIENT WAS ADMITTED FOR ROUTINE OBSERVATIONS POST-PROCEDURE AS PRE-PLANNED, AND NO ESCALATION OF CARE WAS REQUIRED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10308647 ↗MalfunctionJul 22, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE TIP BROKE OFF. A 5F C2 IMAGER II ANGIOGRAPHIC CATHETER WAS SELECTED FOR A URETERIC STENT INSERTION PROCEDURE. AN ANTEGRADE APPROACH WAS USED VIA THE LEFT NEPHROSTOMY TUBE. THE IMAGER II CATHETER WAS PLACED INTO THE ENTRY SHEATH AFTER BEING FLUSHED WITH HEPARINIZED SALINE AND LUBRICATED WITH NORMAL SALINE. WHEN THE CATHETER WAS IN THE LEFT URETER, IT WAS DISCOVERED THAT THE DISTAL TIP WAS BROKEN OFF BUT STILL ATTACHED TO THE CATHETER. THE CATHETER WAS REMOVED AND DISCARDED. ANOTHER OF THE SAME PRODUCT AND BATCH NUMBER WAS OPENED. DURING INSPECTION OF THE CATHETER, WHICH INVOLVED FLUSHING AND LUBRICATING, THE DISTAL TIP OF THE CATHETER DETACHED IN THE RADIOLOGIST'S HAND. A THIRD CATHETER WAS THEN OPENED TO CHECK THE TIP. THE TIP WAS VERY FRAGILE AND EASILY REMOVED FROM THE CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS STABLE POST-PROCEDURE. THE PATIENT WAS ADMITTED FOR ROUTINE OBSERVATIONS POST-PROCEDURE AS PRE-PLANNED, AND NO ESCALATION OF CARE WAS REQUIRED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10308641 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H749085263001Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page