Magic Torque™ M001465920Recalls & reported problems
Boston Scientific Corporation · Glidex™ Coated Hydrophilic Guidewire · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 08714729741428), received between Mar 21, 2019 and Jan 10, 2023 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Fracture2
- Unraveled Material2
- Entrapment of Device1
- Material Integrity Problem1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 10, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT GUIDEWIRE TIP DETACHMENT OCCURRED. A 035/260 (BX/3) MAGIC TORQUE GUIDEWIRE WAS SELECTED FOR USE; HOWEVER, IT WAS NOTED THAT THE TIP WAS BROKEN. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16134002 ↗MalfunctionNov 9, 2022Stretched; Detachment of Device or Device Component; Stretched; Detachment of Device or Device Component
IT WAS REPORTED THAT GUIDEWIRE DETACHMENT OCCURRED. THE PATIENT PRESENTED WITH PERIPHERAL ARTERY DISEASE (PAD) AND UNDERWENT PERIPHERAL ANGIOGRAM. A GW/035/260 MAGIC TORQUE GUIDEWIRE WAS SELECTED FOR USED. HOWEVER, DURING THE PREPARATION, THE FLOPPY TIP WAS NOTICED TO BE STRETCHED UPON REMOVAL FROM THE COIL, AND THE INNER COIL OF THE WIRE SEEMED TO BE DETACHED FROM THE TIP. THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15765074 ↗MalfunctionNov 2, 2022Unraveled Material; Unraveled Material
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2022. IT WAS REPORTED THAT THE TIP OF THE GUIDEWIRE UNRAVELED. A 035/260 (BX/3) MAGIC TORQUE GUIDEWIRE WAS ADVANCED INTO THE LESION. HOWEVER, THE TIP OF THE GUIDEWIRE UNRAVELED WHILE INSIDE THE PATIENT. IT WAS NOTED THAT THIS ALSO HAPPENED IN A SEPARATE CASE USING ANOTHER OF SAME DEVICE. IN BOTH CASES, THE GUIDEWIRE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH GOOD BLOOD FLOW. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A DETACHED CORE WIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15715018 ↗MalfunctionNov 2, 2022Unraveled Material; Unraveled Material
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 13-OCT-2022. IT WAS REPORTED THAT THE TIP OF THE GUIDEWIRE UNRAVELED. A 035/260 (BX/3) MAGIC TORQUE GUIDEWIRE WAS ADVANCED INTO THE LESION. HOWEVER, THE TIP OF THE GUIDEWIRE UNRAVELED WHILE INSIDE THE PATIENT. IT WAS NOTED THAT THIS ALSO HAPPENED IN A SEPARATE CASE USING ANOTHER OF SAME DEVICE. IN BOTH CASES, THE GUIDEWIRE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH GOOD BLOOD FLOW. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A DETACHED CORE WIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15715017 ↗MalfunctionJan 28, 2021Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE GUIDEWIRE'S COATING HAS PEELED. THE 60% STENOSED TARGET LESION WAS LOCATED IN THE COMMON/EXTERNAL ILIAC VEIN. A 035/260 (BX/3) MAGIC TORQUE GUIDEWIRE WAS USED. HOWEVER, AFTER THE GUIDEWIRE WAS REMOVED FROM THE PATIENT'S BODY, IT WAS NOTICED THAT THE COATING WAS PEELING OFF. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11244438 ↗MalfunctionAug 9, 2019Fracture; Fracture
IT WAS REPORTED THAT GUIDEWIRE TIP SHEARED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. A 035/260 MAGIC TORQUE GUIDEWIRE WAS ADVANCED TO CROSS THE LESION. HOWEVER, IT WAS NOTED THAT THE TIP WAS SHEARED. THE TIP WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER MAGIC TORQUE GUIDE WIRE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8879175 ↗MalfunctionAug 9, 2019Fracture; Fracture
IT WAS REPORTED THAT THE WIRE FRACTURED. A GW/035/260 MAGIC TORQUE GUIDE WIRE WAS SELECTED FOR USE. HOWEVER, DURING PROCEDURE, IT WAS NOTED THAT THE WIRE SHEARED. THE WIRE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED. IT WAS FURTHER REPORTED THAT ABOUT 85-90% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. THERE WAS NO COATING LEFT INSIDE THE PATIENT'S BODY. THE PHYSICIAN NOTICED THE SHEARING OF THE DEVICE RIGHT AFTER IT WAS NORMALLY REMOVED FROM THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8879140 ↗MalfunctionMar 21, 2019Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT THE INNOVA DELIVERY SYSTEM WAS STUCK ON THE WIRE. A MAGIC TORQUE GUIDEWIRE AND A 7X150X130 INNOVA VASCULAR STENT WERE SELECTED FOR A STENT PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY (SFA). DURING THE PROCEDURE, THERE WAS DIFFICULTY ADVANCING THE INNOVA DEVICE. THE INNOVA STENT WAS DEPLOYED SAFELY INTO THE PATIENT. THE DELIVERY SYSTEM WAS DIFFICULT TO REMOVE AND BECAME STUCK ON THE MAGIC TORQUE GUIDEWIRE. EVERYTHING WAS REMOVED TOGETHER. THE PHYSICIAN LOST WIRE POSITION. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8441644 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page