Boston Scientific Corporation · High Performance Guidewire · Class 2
All models of Jagwire (10)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Jagwire high performance guidewire is used to guide scopes and stents through digestive and urinary tracts. Most of the 48 reports describe the wire's coating peeling or flaking off, the tip breaking or detaching inside the patient, or the wire being hard to advance; a few mention bleeding or a perforation. Reports include 1 death and 5 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 48
- Deaths reported
- 1
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 48 adverse event reports that list this exact device (DI 08714729747215), received between Jan 13, 2017 and Sep 11, 2026 — classified as 1 death, 5 injuries, 42 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (21 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem22
- Peeled/Delaminated21
- Break18
- Appropriate Device Problem Term/Code Not Available8
- Detachment of Device or Device Component8
- Difficult to Advance3
- Tear, Rip or Hole in Device Packaging2
- Defective Device1
Patient problems
- No Consequences Or Impact To Patient20
- No Clinical Signs, Symptoms or Conditions18
- No Known Impact Or Consequence To Patient6
- Hemorrhage/Blood Loss/Bleeding2
- Perforation2
- Death1
- No Patient Involvement1
- Pancreatitis1
Recent adverse event reports
The 25 most recent of 48 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 11, 2026Difficult to Advance; Material Deformation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50744400 ↗MalfunctionJul 9, 2026Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS USED IN THE COLON DURING A COLON STENT PLACEMENT PROCEDURE TO TREAT COLORECTAL CANCER PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, IT WAS REPORTED THAT THE DISTAL TIP WAS SEPARATED AND REMAINED IN THE BODY. IT WAS RETRIEVED USING FORCEPS. THE PROCEDURE WAS DISCONTINUED AND EVENTUALLY CANCELLED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25771548 ↗MalfunctionApr 1, 2026Tear, Rip or Hole in Device Packaging
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS RECEIVED BY THE CUSTOMER ON (B)(6) 2026. IT WAS REPORTED THAT A PERFORATION WAS REPORTED ON THE PRODUCT PACKAGING, CAUSING THE CONTENTS TO BECOME VISIBLE. THE DEVICE WAS NOT USED IN A PROCEDURE. THE REPORTED EVENT MAY SUGGEST THAT THE STERILE BARRIER WAS COMPROMISED DUE TO THE DAMAGED PACKAGING. THE PATIENT WAS NOT PRESENT AT THE TIME OF THE EVENT AND THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24760559 ↗MalfunctionFeb 3, 2026Nonstandard Device
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE JAGWIRE WAS REPORTED AS DEFECTIVE. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. IT WAS REPORTED THAT THE PATIENT HOSPITALIZATION WAS PROLONGED. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24248822 ↗MalfunctionJan 12, 2026Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem
NOTE: THIS REPORT PERTAINS TO ONE OF FOUR JAGWIRES USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, WHILE ADVANCING THE GUIDEWIRE INTO THE COMMON BILE DUCT, PEELING OCCURRED ON THE BLACK PART OF THE HYDROPHILIC TIP. THE SAME PROBLEM OCCURRED WITH THE OTHER THREE JAGWIRES. DUE TO THIS PROBLEM, THE PROCEDURE HAD TO BE TERMINATED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24049823 ↗MalfunctionJan 12, 2026Peeled/Delaminated
NOTE: THIS REPORT PERTAINS TO ONE OF FOUR JAGWIRES USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, WHILE ADVANCING THE GUIDEWIRE INTO THE COMMON BILE DUCT, PEELING OCCURRED ON THE BLACK PART OF THE HYDROPHILIC TIP. THE SAME PROBLEM OCCURRED WITH THE OTHER THREE JAGWIRES. DUE TO THIS PROBLEM, THE PROCEDURE HAD TO BE TERMINATED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24049647 ↗MalfunctionJan 12, 2026Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem
NOTE: THIS REPORT PERTAINS TO ONE OF FOUR JAGWIRES USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, WHILE ADVANCING THE GUIDEWIRE INTO THE COMMON BILE DUCT, PEELING OCCURRED ON THE BLACK PART OF THE HYDROPHILIC TIP. THE SAME PROBLEM OCCURRED WITH THE OTHER THREE JAGWIRES. DUE TO THIS PROBLEM, THE PROCEDURE HAD TO BE TERMINATED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24049627 ↗MalfunctionJan 12, 2026Peeled/Delaminated
NOTE: THIS REPORT PERTAINS TO ONE OF FOUR JAGWIRES USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, WHILE ADVANCING THE GUIDEWIRE INTO THE COMMON BILE DUCT, PEELING OCCURRED ON THE BLACK PART OF THE HYDROPHILIC TIP. THE SAME PROBLEM OCCURRED WITH THE OTHER THREE JAGWIRES. DUE TO THIS PROBLEM, THE PROCEDURE HAD TO BE TERMINATED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24049515 ↗InjuryJun 27, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE GUIDEWIRE WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE GUIDEWIRE EXPERIENCED AN UNANTICIPATED DISTAL FRACTURE INTO SEVERAL SEGMENTS, AS IT WAS EMERGING FROM THE TIP OF THE SCOPE AND PRIOR TO ENTERING THE COMMON BILE DUCT. ALSO, IT WAS NOTED THAT THE DISTAL TIP DETACHED. IT WAS REPORTED THAT THE GUIDEWIRE WAS RETRIEVED IN TWO STAGES. FIRST AS AN INTACT UNIT FOLLOWED BY REMOVAL OF THE FRACTURED SEGMENTS BY BIOPSY GRASPING FORCEPS. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22346865 ↗InjuryJun 27, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE GUIDEWIRE WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE GUIDEWIRE EXPERIENCED AN UNANTICIPATED DISTAL FRACTURE INTO SEVERAL SEGMENTS, AS IT WAS EMERGING FROM THE TIP OF THE SCOPE AND PRIOR TO ENTERING THE COMMON BILE DUCT. ALSO, IT WAS NOTED THAT THE DISTAL TIP DETACHED. IT WAS REPORTED THAT THE GUIDEWIRE WAS RETRIEVED IN TWO STAGES. FIRST AS AN INTACT UNIT FOLLOWED BY REMOVAL OF THE FRACTURED SEGMENTS BY BIOPSY GRASPING FORCEPS. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22346834 ↗MalfunctionApr 30, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS BEING PREPARED FOR USE IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO BE PERFORMED ON (B)(6) 2025. DURING PREPARATION, THE HYDROPHILIC COATING OF THE GUIDEWIRE HAD COME OFF. THE PROCEDURE WAS NOT COMPLETED SUCCESSFULLY BECAUSE THEY DID NOT HAVE MORE MATERIAL. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21936457 ↗InjuryDec 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NOTE: THIS REPORT PERTAINS TO JAGWIRES THAT WERE USED IN THE RETROSPECTIVE STUDY. BOSTON SCIENTIFIC BECAME AWARE OF EVENTS INVOLVING CAPTIVATOR II SNARES, INTERJECT NEEDLES, JAGWIRES AND RESOLUTION CLIPS THROUGH THE ARTICLE "THE USEFULNESS OF TRACTION-ASSISTED ENDOSCOPIC PAPILLECTOMY FOR AMPULLARY EARLY TUMORS" BY XIE JIAO ET AL. PER ARTICLE, BETWEEN JANUARY 2010 AND AUGUST 2022, STUDY OF 173 PATIENTS SUFFERING AMPULLARY TUMORS UNDERWENT ENDOSCOPIC PAPILLECTOMY PROCEDURES. TWO TECHNIQUES WERE USED: TRACTION ASSISTED, AND NON-TRACTION ASSISTED METHODS. THE FOLLOWING OCCURRED WITH THE STUDY PATIENTS: PANCREATITIS, CHOLANGITIS, PERFORATIONS, INTRAOPERATIVE BLEEDING THAT REQUIRED ENDOSCOPIC HEMOSTASIS AND BLOOD TRANSFUSION. PLEASE SEE THE REFERENCE ARTICLE FOR FULL DETAILS.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Pancreatitis; Perforation
MAUDE report 21035624 ↗MalfunctionMay 1, 2024Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS USED IN THE PANCREATIC DUCT (PD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE CORE/INNER PART OF THE JAGWIRE WAS STRIPPED OFF AND REMAINED IN THE PD. THE PROCEDURE WAS COMPLETED WITH THE SAME JAGWIRE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19226847 ↗MalfunctionDec 14, 2022Break; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Break; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE ON (B)(6) 2022. DURING PREPARATION, THE GUIDEWIRE PEELED OFF. THE PHYSICIAN REPLACED IT WITH A SECOND JAGWIRE GUIDEWIRE, BUT THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS COMPLETED WITH THE THIRD JAGWIRE GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF CORE WIRE BREAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15979998 ↗MalfunctionDec 8, 2022Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS USED DURING AN ENDOSCOPIC ULTRASOUND PROCEDURE PERFORMED ON (B)(6) 2022. DURING PROCEDURE, THE PTFE COATING PEELED OFF WHILE REMOVING THE GUIDEWIRE FROM THE SCOPE. THE COATING WAS NOT REMOVED FROM THE PATIENT. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15940994 ↗MalfunctionSep 14, 2022Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS USED DURING INTERVENTIONAL PROCEDURE ON (B)(6) 2022. DURING THE PROCEDURE, THE GUIDEWIRE PEELED OFF. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15422044 ↗MalfunctionMar 28, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE AS BEING USED IN COMMON BILE DUCT DURING ERCP AND REMOVAL OF STONES PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, THE PHYSICIAN USED A STONE RETRIEVAL EXACTOR BALLOON IN THE COMMON BILE DUCT OVER THE GUIDEWIRE. AT THE TIME OF RETRACTION OF THE GUIDEWIRE, THE PHYSICIAN NOTICED A SMALL FRAGMENT ON THE X-RAY. THEY PULLED OUT THE GUIDEWIRE AND SAW THAT A PART OF IT WAS PEELED OFF. IT WAS REPORTED THAT THE COATING ON THE TIP WAS NOT ACTIVATED USING SALINE BEFORE INSERTING. THE PROCEDURE WAS CANCELLED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13929156 ↗MalfunctionFeb 4, 2022Break; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Break; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS USED IN THE COMMON BILE DUCT DURING AN ERCP (ENDOSCOPIC RETROGRADE CHOLANGIO PANCREATOGRAPHY) PROCEDURE ON (B)(6) 2021. DURING PREPARATION, THE NURSE NOTICED THAT THE GUIDEWIRE TIP WAS PEELED. THE PROCEDURE WAS COMPLETED WITH A NEW JAGWIRE GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED DUE TO THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS: COREWIRE WAS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13453718 ↗MalfunctionMar 29, 2021Break; Difficult to Advance; Adverse Event Without Identified Device or Use Problem; Break; Difficult to Advance; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS USED IN THE BILIARY TRACT DURING A PROCEDURE ON (B)(6) 2021. DURING THE PROCEDURE, THE TIP OF THE GUIDEWIRE COULD NOT CROSS THE SCOPE. THE PROCEDURE WAS COMPLETED WITH A NEW JAGWIRE GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS: COREWIRE BREAK OR FRACTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11585537 ↗DeathJan 13, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE AND EXTRACTOR PRO WERE USED IN THE COLON DURING A WALLFLEX COLONIC METAL STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, THE PHYSICIAN INITIALLY USED AN EGD SCOPE TO CHECK THE TUMOR/STRICTURE, THEN USED A JAGWIRE AND EXTRACTOR PRO THROUGH AN ADULT COLONOSCOPE TO GAIN ACCESS THROUGH THE STRICTURE. THE BALLOON WAS ADVANCED DOWN THE GUIDEWIRE AND ENTERED THE BIOPSY CHANNEL OF THE SCOPE HOWEVER IT DID NOT EXIT THE DISTAL TIP. ACCESS WAS LOST AND THE GUIDEWIRE SLIPPED OUT OF THE STRICTURE. AT THIS POINT, ONLY THE GUIDEWIRE WAS IN THE COLONOSCOPE. A PERFORATION OCCURRED WHEN THE PHYSICIAN REPOSITIONED THE SCOPE NEAR THE TUMOR. IT IS UNKNOWN WHAT CAUSED THE PERFORATION. AT THAT POINT THE PATIENT WAS SENT TO SURGERY. THE PATIENT DIED, HOWEVER, IT IS UNKNOWN IF IT HAPPENED DURING OR AFTER THE SURGERY. THERE IS NO INFORMATION REGARDING THE CAUSE OF THE PATIENT'S DEATH, AND AUTOPSY RESULTS WERE NOT MADE AVAILABLE. REPORTEDLY, THE PATIENT HAD A MALIGNANT TUMOR IN THE COLON, BUT NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Death; Perforation
MAUDE report 11158301 ↗MalfunctionNov 10, 2020Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT JAGWIRE GUIDEWIRE WAS USED IN THE BILIARY SYSTEM (PAPILLA) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN PAPILLA WAS CUT WITH A SPHINCTEROTOME AND GUIDEWIRE WAS INSERTED IN THE NON BOSTON SCIENTIFIC DUODENOSCOPE, THE PTFE PEELED. THIS HAPPENED WITH SEVEN JAGWIRES. THE PROCEDURE WAS CANCELLED BASED UPON NO ADDITIONAL STOCK AVAILABLE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10817095 ↗MalfunctionNov 10, 2020Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT JAGWIRE GUIDEWIRE WAS USED IN THE BILIARY SYSTEM (PAPILLA) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN PAPILLA WAS CUT WITH A SPHINCTEROTOME AND GUIDEWIRE WAS INSERTED IN THE NON BOSTON SCIENTIFIC DUODENOSCOPE, THE PTFE PEELED. THIS HAPPENED WITH SEVEN JAGWIRES. THE PROCEDURE WAS CANCELLED BASED UPON NO ADDITIONAL STOCK AVAILABLE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10816991 ↗MalfunctionNov 10, 2020Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT JAGWIRE GUIDEWIRE WAS USED IN THE BILIARY SYSTEM (PAPILLA) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN PAPILLA WAS CUT WITH A SPHINCTEROTOME AND GUIDEWIRE WAS INSERTED IN THE NON BOSTON SCIENTIFIC DUODENOSCOPE, THE PTFE PEELED. THIS HAPPENED WITH SEVEN JAGWIRES. THE PROCEDURE WAS CANCELLED BASED UPON NO ADDITIONAL STOCK AVAILABLE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10816919 ↗MalfunctionNov 10, 2020Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT JAGWIRE GUIDEWIRE WAS USED IN THE BILIARY SYSTEM (PAPILLA) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN PAPILLA WAS CUT WITH A SPHINCTEROTOME AND GUIDEWIRE WAS INSERTED IN THE NON BOSTON SCIENTIFIC DUODENOSCOPE, THE PTFE PEELED. THIS HAPPENED WITH SEVEN JAGWIRES. THE PROCEDURE WAS CANCELLED BASED UPON NO ADDITIONAL STOCK AVAILABLE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10816918 ↗MalfunctionNov 10, 2020Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT JAGWIRE GUIDEWIRE WAS USED IN THE BILIARY SYSTEM (PAPILLA) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN PAPILLA WAS CUT WITH A SPHINCTEROTOME AND GUIDEWIRE WAS INSERTED IN THE NON BOSTON SCIENTIFIC DUODENOSCOPE, THE PTFE PEELED. THIS HAPPENED WITH SEVEN JAGWIRES. THE PROCEDURE WAS CANCELLED BASED UPON NO ADDITIONAL STOCK AVAILABLE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10815708 ↗
FDA clearance and classification
The FDA regulates product code OCY (endoscopic guidewire, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Jagwire M00556580
Has the Jagwire M00556580 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729747215), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Jagwire M00556580?
The FDA's MAUDE database lists 48 adverse event reports for this device, received between Jan 13, 2017 and Sep 11, 2026. In those reports, the most common specific device problems are “Peeled/Delaminated” (21), “Break” (18) and “Detachment of Device or Device Component” (8); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (2), “Perforation” (2) and “Death” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Jagwire M00556580?
1 report is classified as a death, 5 as injuries and 42 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Jagwire M00556580 FDA approved?
Devices of this type (product code OCY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Boston Scientific Corporation.
Who makes the Jagwire M00556580, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Jagwire, model M00556580). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic guidewire, gastroenterology-urology devices
- HiWire G30474Cook Incorporated · Endoscopic guidewire, gastroenterology-urology81 reports
- Roadrunner G18154Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall18 reports
- Roadrunner G17542Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall9 reports
- Roadrunner G18153Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall6 reports
- Roadrunner G18155Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall3 reports
- Roadrunner G17539Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall2 reports
- Roadrunner G17540Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall0 reports
- Acrobat G34266Cook Incorporated · Endoscopic guidewire, gastroenterology-urology44 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.