SENSATION Short Throw M00560310Recalls & reported problems
Boston Scientific Corporation · Single-Use Polypectomy Snare · Class 2
All models of SENSATION Short Throw (7)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Boston Scientific SENSATION short throw snare is a single-use polypectomy snare, a wire loop used to cut polyps during endoscopic procedures in the intestine. Reports mostly describe the loop failing to cut the polyp — the wire slipping off the tissue when tightened — plus difficulty positioning or retracting it. There are 59 reports, with no deaths and no injuries; most describe the procedure being finished with another snare and no patient complications. One open recall from January 2021 concerns an incomplete pouch seal noticed before use.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 58
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 58 adverse event reports that list this exact device (DI 08714729747598), received between Dec 10, 2018 and Jun 22, 2026 — classified as 58 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (50 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jan 19, 2021Open, Classified
Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient.
Z-1170-2021 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Cut50
- Positioning Problem21
- Retraction Problem5
- Defective Device2
- Device-Device Incompatibility2
- Electrical /Electronic Property Problem2
- Failure to Deliver Energy2
- Nonstandard Device2
Patient problems
- No Clinical Signs, Symptoms or Conditions53
- No Consequences Or Impact To Patient5
Recent adverse event reports
The 25 most recent of 58 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 22, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE IN THE INTESTINAL TRACT, ON (B)(6) 2026. DURING THE PROCEDURE, IT WAS NOTED THAT THE SUPPORT WAS INSUFFICIENT. WHEN TIGHTENING THE METAL WIRE TO ENCASE THE POLYP, IT KEPT SLIPPING OFF AND COULD NOT CUT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25573215 ↗MalfunctionJun 16, 2026Failure to Cut
IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT FOR AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE THE SNARE FAILED TO CUT THE TARGET POLYP. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE SNARE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25520194 ↗MalfunctionJun 4, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A SENSATION SARE WAS USED IN THE INTESTINAL TRACT FOR AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE THE SNARE FORCE WAS INSUFFICIENT, IT WOULD SLIP AWAY FROM THE TISSUE SURFACE WHEN IT WAS TIGHTENED, AND THE TARGET POLYP COULD NOT BE TRAPPED OR CUT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT CONDITION FOLLOWING PROCEDURE WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25388538 ↗MalfunctionJun 3, 2026Failure to Cut
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE DEVICE WAS USED TO TREAT THE POLYP FOUND IN THE DIGESTIVE TRACT IN THE INTESTINAL TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, INSUFFICIENT SUPPORT FORCE CAUSED THE METAL WIRE TO SLIP FROM THE TISSUE SURFACE WHEN TIGHTENED, PREVENTING THE LOOP FROM CUTTING THE TARGET POLYP. THE PROCEDURE WAS COMPLETED WITH ANOTHER SENSATION SNARE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25374879 ↗MalfunctionJun 2, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE PHYSICIAN TRIED TO CUT A POLYP, BUT THE SUPPORT FORCE WAS NOT ENOUGH TO SUCCESSFULLY CUT THE TARGET POLYP, AND THE LOOP CONTINUOUSLY SLIDE OFF THE SURFACE OF THE TISSUE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25368538 ↗MalfunctionJun 2, 2026Failure to Cut
IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT FOR AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, INSUFFICIENT SUPPORTING FORCE WAS REPORTED, CAUSING THE SNARE TO FAIL TO CUT THE TARGET POLYP. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT CONDITION FOLLOWING PROCEDURE WAS STABLE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25368149 ↗MalfunctionJun 2, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE PHYSICIAN TRIED TO CUT A POLYP, BUT THE SUPPORT FORCE WAS NOT ENOUGH TO SUCCESSFULLY CUT THE TARGET POLYP, AND THE LOOP CONTINUOUSLY SLIDE OFF THE SURFACE OF THE TISSUE. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25361707 ↗MalfunctionJun 2, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT FOR AN ENDOSCOPIC POLYPECTOMY PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE THE SNARE LOOP PRESENTED ISSUES POSITIONING AROUND THE POLYP AND COULD NOT CUT IT. PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE SNARE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25359374 ↗MalfunctionJun 1, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON 07-MAY-2026. DURING THE PROCEDURE, THE PHYSICIAN TRIED TO CUT A POLYP, BUT THE SUPPORT FORCE WAS NOT ENOUGH TO SUCCESSFULLY CUT THE TARGET POLYP, AND THE LOOP CONTINUOUSLY SLIDE OFF THE SURFACE OF THE TISSUE. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25355290 ↗MalfunctionJun 1, 2026Failure to Cut
IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT FOR AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE THE SNARE FAILED TO CUT THE POLYP. PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25351189 ↗MalfunctionJun 1, 2026Failure to Cut
IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT FOR AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE THE SNARE FAILED TO CUT THE POLYP. PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25351181 ↗MalfunctionJun 1, 2026Failure to Cut
It was reported to Boston Scientific Corporation that a Sensation Snare was used in the intestinal tract during an Endoscopic Polypectomy procedure performed on (b)(6) 2026. During the procedure, the supporting force was insufficient, preventing the completion of the cutting procedure. The procedure was completed with a different device. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25350648 ↗MalfunctionJun 1, 2026Failure to Cut
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARES WAS USED IN THE INTESTINAL TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. IT WAS REPORTED THAT THE SUPPORTING FORCE WAS INSUFFICIENT, AND THE DEVICE WAS UNABLE TO COMPLETE THE CUTTING DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25349246 ↗MalfunctionJun 1, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE PHYSICIAN TRIED TO CUT A POLYP, BUT THE SUPPORT FORCE WAS NOT ENOUGH TO SUCCESSFULLY CUT THE TARGET POLYP, AND THE LOOP CONTINUOUSLY SLIDE OFF THE SURFACE OF THE TISSUE. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25347902 ↗MalfunctionJun 1, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE SUPPORTING FORCE WAS INSUFFICIENT, CAUSING THE WIRE TO SLIP FROM THE TISSUE SURFACE WHEN TIGHTENED, AND PREVENTING THE TARGET POLYP FROM BEING CAPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25347893 ↗MalfunctionMay 29, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE DIGESTIVE TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE PHYSICIAN TRIED TO CUT A POLYP, BUT THE LOOP CONTINUOUSLY SLIDE OFF THE SURFACE OF THE TISSUE. AS A RESULT, THE SUPPORT FORCE WAS INSUFFICIENT, AND THE SNARE WAS UNABLE TO SUCCESSFULLY CUT THE TARGET POLYP. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25338424 ↗MalfunctionMay 28, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE DIGESTIVE TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE PHYSICIAN TRIED TO CUT A POLYP, BUT THE LOOP CONTINUOUSLY SLIDE OFF THE SURFACE OF THE TISSUE. AS A RESULT, THE SUPPORT FORCE WAS INSUFFICIENT, AND THE SNARE WAS UNABLE TO SUCCESSFULLY CUT THE TARGET POLYP. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25317998 ↗MalfunctionMay 28, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE DIGESTIVE TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON(B)(6) 2026. DURING THE PROCEDURE, THE PHYSICIAN TRIED TO CUT A POLYP, BUT THE LOOP CONTINUOUSLY SLIDE OFF THE SURFACE OF THE TISSUE. AS A RESULT, THE SUPPORT FORCE WAS INSUFFICIENT, AND THE SNARE WAS UNABLE TO SUCCESSFULLY CUT THE TARGET POLYP. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25317993 ↗MalfunctionMay 28, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT FOR AN ENDOSCOPIC POLYPECTOMY PROCEDURE ON (B)(6) 2026. DURING THE PROCEDURE, THE SNARE FAILED TO CUT THE POLYP AND PRESENTED DIFFICULTY TO POSITION. PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25316908 ↗MalfunctionMay 19, 2026Nonstandard Device; Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE DIGESTIVE TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE PHYSICIAN WAS UNABLE TO SUCCESSFULLY CUT THE TARGET POLYP. WHEN THE METAL WIRE WAS TIGHTENED, THE POLYP WOULD SLIP OFF. THE SUPPORT WAS INSUFFICIENT DUE TO THE WORKING LENGTH BEING TOO SHORT. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25216888 ↗MalfunctionMay 19, 2026Failure to Cut; Positioning Problem
It was reported to Boston Scientific Corporation that a Sensation Snare was used in the intestinal tract during an Endoscopic Polypectomy procedure performed on (b)(6) 2026. During the procedure, the supporting force was insufficient; during snaring attempts, the target polyp consistently slipped and could not be cut. The procedure was completed with a different device. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25216862 ↗MalfunctionMay 18, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE SUPPORTING FORCE WAS INADEQUATE; DURING ATTEMPTS TO SNARE THE TARGET POLYP, IT REPEATEDLY SLIPPED AND COULD NOT BE SUCCESSFULLY SNARED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25202342 ↗MalfunctionMay 18, 2026Nonstandard Device; Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE DIGESTIVE TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE PHYSICIAN TRIED TO CUT A POLYP, BUT THE LOOP CONTINUOUSLY SLIPPED OFF THE SURFACE OF THE TISSUE, AND THE SUPPORT WAS INSUFFICIENT DUE THE WORKING LENGTH BEING TOO SHORT. AS A RESULT, THE SNARE WAS UNABLE TO SUCCESSFULLY CUT THE TARGET POLYP. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION FOLLOWING PROCEDURE WAS STABLE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25197105 ↗MalfunctionMay 15, 2026Failure to Cut; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE INTESTINAL TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON APRIL 22, 2026. DURING THE PROCEDURE, THE APPLIED SUPPORT FORCE WAS INSUFFICIENT; DURING ATTEMPTS TO CAPTURE THE TARGET POLYP, IT REPEATEDLY SLIPPED, AND THE TARGET POLYP COULD NOT BE CAPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25193773 ↗MalfunctionMay 13, 2026Failure to Cut
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSATION SNARE WAS USED IN THE DIGESTIVE TRACT DURING AN ENDOSCOPIC POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE PHYSICIAN TRIED TO CUT A POLYP, BUT THE SUPPORT WAS INSUFFICIENT, AND THE SNARE WAS UNABLE TO SUCCESSFULLY CUT THE TARGET POLYP. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25164156 ↗
Other SNARE, FLEXIBLE devices
- CAPTIFLEX M00562400Boston Scientific Corporation · SNARE, FLEXIBLE3 recalls85 reports
- SENSATION Short Throw M00562670Boston Scientific Corporation · SNARE, FLEXIBLE1 recall178 reports
- SENSATION Short Throw M00562650Boston Scientific Corporation · SNARE, FLEXIBLE1 recall173 reports
- Ultratome XL M00535900Boston Scientific Corporation · SNARE, FLEXIBLE205 reports
- CAPTIVATOR M00562340Boston Scientific Corporation · SNARE, FLEXIBLE2 recalls71 reports
- CAPTIVATOR II M00561230Boston Scientific Corporation · SNARE, FLEXIBLE1 recall83 reports
- CAPTIVATOR M00562320Boston Scientific Corporation · SNARE, FLEXIBLE2 recalls32 reports
- CAPTIVATOR M00562300Boston Scientific Corporation · SNARE, FLEXIBLE1 recall78 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page