Skip to content

SENSATION Short Throw recallZ-1170-2021 · Boston Scientific Corporation · initiated Jan 19, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient.

FDA root cause category: Process change control

What the company is doing

On 01/21/2021, Boston Scientific issued Urgent Medical Device Product Removal notices to customers via e-mail. The removal notice is due to an incomplete seal of the inner pouch which could result in a sterile barrier breach rendering the specified lots of polypectomy snare devices non-sterile.

Recall details

Recall number
Z-1170-2021
Status
Open, Classified
Initiated by the company
Jan 19, 2021
Recalling firm
Boston Scientific Corporation
Quantity
335,788 units total (Domestic: 222,882, Foreign: 112,906)
Product code
FDI
FDA submissions
K131700
Distribution
US nationwide distribution.

Product description

Captivator/Captiflex/Sensation Short Throw Single-Use Polypectomy Snares Outer box UPN, Inner pouch UPN: M00560311 (Box 5), M00560310; M00560321 (Box 5), M00560320; M00562651 (Box 5), M00562650; M00562652 (Box 20), M00562650; M00562671 (Box 5), M00562670; M00562672 (Box 20), M00562670; M00562673 (Box 40), M00562670; M00562691 (Box 5); M00562690; M00562692 (Box 20), M00562690; M00562693 (Box 40), M00562690.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Outer box GTIN: 08714729283904 Inner pouch GTIN: 08714729747598 Lot/Batch#: 24818278, 24906823, 25042455  Outer box GTIN: 08714729283928 Inner pouch GTIN: 08714729747611 Lot/Batch#: 24685766, 24817145, 24899710, 24739786, 25006735, 24861901  Outer box GTIN: 08714729158110 Inner pouch GTIN: 08714729748069 Lot/Batch#: 24685764, 24890051, 24824132  Outer box GTIN: 08714729268802 Inner pouch GTIN: 08714729748069 Lot/Batch#: 25023345  Outer box GTIN: 08714729158141 Inner pouch GTIN: 08714729748083 Lot/Batch#: 24699718, 24939626, 24861739  Outer box GTIN: 08714729268819 Inner pouch GTIN: 08714729748083 Lot/Batch#: 24824134, 25006729, 24709534  Outer box GTIN: 08714729501664 Inner pouch GTIN: 08714729748083 Lot/Batch#: 25023203, 24697822, 24739587, 24942292, 24909214, 24824136  Outer box GTIN: 08714729158158 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24856636, 24927642, 24699218  Outer box GTIN: 08714729268826 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24835709, 24818605, 25023343  Outer box GTIN: 08714729501671 Inner pouch GTIN: 08714729748106 Lot/Batch#: 24909216, 24758832, 24872119, 24994171, 25048397, 2475126, 24929431, 24995438

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 87221).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page