Skip to content

FATHOM™ -16 M001509110Recalls & reported problems

Boston Scientific Corporation · Steerable Guidewire · Class 2

All models of FATHOM -16 (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729762560), received between Oct 18, 2018 and Sep 2, 2026 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Fracture3
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Difficult to Insert1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient2
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 2, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50569610 ↗
  • MalfunctionApr 3, 2025Fracture; Difficult to Insert; Fracture; Difficult to Insert

    IT WAS REPORTED THAT TIP FRACTURE OCCURRED. A 180X35CM FATHOM -16 GUIDEWIRE WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE TIP WAS DELAMINATED WHEN PULLED OUT OF ITS LOOP AND COULD NOT BE INSERTED INTO THE HUB OF THE MICROCATHETER. THE PROCEDURE WAS COMPLETED USING AN ALTERNATE DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21763477 ↗
  • MalfunctionFeb 8, 2024Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. THE PATIENT WAS BEING TREATED FOR LIVER METASTASES AND UNDERWENT BLAND EMBOLIZATION. THE NON-STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND NON-CALCIFIED VESSEL. A180X35CM FATHOM -16 GUIDEWIRE WAS ADVANCED; HOWEVER, DURING THE PROCEDURE, THE GUIDEWIRE TIP BROKE OFF. THE TIP WAS HELD TOGETHER BY AN INTERNAL THREAD AND WAS PULLED BACK INTO THE MICROCATHETER AND WAS SUCCESSFULLY RETRIEVED. THE PROCEDURE WAS COMPLETED WITH ANOTHER GUIDEWIRE. THERE WERE NO COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18672465 ↗
  • MalfunctionJun 11, 2021Detachment of Device or Device Component; Material Integrity Problem; Detachment of Device or Device Component; Material Integrity Problem

    IT WAS REPORTED THAT THE COATING OF THE GUIDEWIRE WAS COMING OFF. A FATHOM 16 .016, 180X35CM, PERIPHERAL GUIDEWIRE WAS SELECTED FOR USE IN A TRANSARTERIAL CHEMOEMBOLIZATION (TACE) PROCEDURE IN THE LIVER. WHILE RETRACTING THE WIRE BACK THROUGH THE INTRODUCER NEEDLE, IT WAS REPORTED THAT THE COATING OF THE FATHOM WAS COMING OFF. IT WAS CONFIRMED THAT NO COATING AND/OR FRAGMENTS WERE LEFT INSIDE THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11985198 ↗
  • MalfunctionFeb 24, 2020Fracture; Fracture

    IT WAS REPORTED THAT A GUIDEWIRE FRACTURE OCCURRED. A 180X35CM FATHOM -16 GUIDEWIRE WAS SELECTED FOR USE. DURING PREPARATION, IT WAS NOTED THAT THE FATHOM WIRE WAS SPLIT. THE PHYSICIAN REVEALED THAT THE SAME ISSUE HAPPENED TWICE. NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9744596 ↗
  • InjuryJan 21, 2020Fracture; Fracture

    IT WAS REPORTED THAT WIRE SHEARED OFF VIA VOLUNTARY MEDWATCH MW5091859. A 180X35CM FATHOM -16 GUIDEWIRE WAS SELECTED FOR USE. DURING THE STENT INSERTION TO THE RIGHT SIDE NEPHROURETERAL, IT WAS NOTED THAT A SMALL FRAGMENT OF THE WIRE SHEARED OFF. THE WIRE COULD NOT BE RETRIEVED FROM THE COLLECTING SYSTEM. THE FRAGMENT WILL BE REMOVED DURING A NEPHROLITHOTOMY PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9610670 ↗
  • MalfunctionOct 18, 2018Break; Break

    IT WAS REPORTED THAT THE COATING OF THE DEVICE BROKE. A 180X35CM FATHOM 16 WAS SELECTED FOR USE. A BREAK OF THE COATING AT THE TIP OF THE DEVICE WAS NOTED UPON OPENING THE DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7979727 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page