Interlock™ M001361740Recalls & reported problems

Boston Scientific Corporation · Fibered IDC™ Occlusion System · Class 2 · Implant

All models of Interlock (26)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729765080), received between Apr 14, 2020 and Jan 26, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem3
  • Adverse Event Without Identified Device or Use Problem1
  • Component Missing1
  • Premature Activation1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 26, 2026Separation Failure; Adverse Event Without Identified Device or Use Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 30DEC2025. IT WAS REPORTED THAT THE COIL DID NOT DEPLOY. THE NON STENOSED TARGET LESION WAS LOCATED IN A NON-TORTUOUS AND NON-CALCIFIED VESSEL OF PROSTATE. A 2/3MM X 2.3CM INTERLOCK WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE COIL GOT STUCK AND DID NOT DEPLOY. THE COIL WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED DETACHED PUSHER WIRE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24179468 ↗
  • MalfunctionFeb 24, 2022Component Missing; Material Integrity Problem; Component Missing; Material Integrity Problem

    IT WAS REPORTED THAT THERE WAS NO FIBERED COIL. THE TARGET LESION WAS LOCATED IN THE SPLENIC ARTERY. A 2/3MM X 2.3CM F-IDC VORTX DIAMOND INTERLOCK COIL WAS SELECTED FOR USE. DURING THE PROCEDURE, THE COIL FAILED TO FORM INSIDE THE LESION WHEN THE PHYSICIAN TRIED TO DEPLOY IT. THE COIL WAS WITHDRAWN AND ANOTHER ATTEMPT TO DEPLOY WAS UNSUCCESSFUL AS THE COIL STILL FAILED TO FORM. CONSEQUENTLY, IT WAS WITHDRAWN FOR THE SECOND TIME AND DEPLOYED OUTSIDE THE PATIENT, AND FOUND THERE WAS "NO FIBERED COIL". THE PHYSICIAN NOTED, "AFTER END HAS INTERLOCKING ARM THE FRONT HAS NOTHING". CONFIRMATION WAS MADE NO DEVICE FRAGMENTS WERE LEFT INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED BY UNKNOWN INJECTED EMBOLIC MICROSPHERES. THERE WERE NO COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13590975 ↗
  • InjuryMay 4, 2021Premature Activation; Material Integrity Problem; Premature Activation; Material Integrity Problem

    IT WAS REPORTED THAT THE COIL NEEDED TO BE SNARED TO REMOVE. A 2/3MMX2.3CM INTERLOCK WAS SELECTED FOR USE. THE TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS VESSEL. DURING THE PROCEDURE, THE COIL WAS FED INTO THE MICROCATHETER. THE PHYSICIAN WAS NOT HAPPY WITH THE CATHETER POSITION, SO THE COIL WAS PULLED OUT WITHOUT DETACHING. THE GUIDEWIRE WAS PLACED INTO THE CATHETER TO REPOSITION, BUT THE COIL WAS UNEXPECTEDLY IN MICROCATHETER AND WAS THEN DEPLOYED IN SUBOPTIMAL POSITION. THE COIL NEEDED TO BE SNARED TO REMOVE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11770023 ↗
  • MalfunctionApr 14, 2020Material Integrity Problem; Material Integrity Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 26MAR2020. IT WAS REPORTED THAT THERE WAS NO COILING IN THE VESSEL. A 2/3MM X 2.3CM INTERLOCK WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE DEVICE DID NOT COIL IN THE VESSEL. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS REPORTED AND PATIENT'S CONDITION WAS STABLE. HOWEVER, DEVICE ANALYSIS REVEALED DETACHED INTERLOCKING ARM AND MISSING COIL FIBER BUNDLES. IT WAS FURTHER REPORTED THAT THE COIL WAS NOT DEPLOYED SUCCESSFULLY AND WAS THEN RETRIEVED SAFELY. THERE WERE NO DEVICE FRAGMENTS LEFT INSIDE THE PATIENT'S BODY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9957747 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →