Expo™ H74908526573A1Recalls & reported problems
Boston Scientific Corporation · Angiographic Catheter · Class 2
All models of Expo (47)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Mar 21, 2024Class IOpen, Classified
An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.
Z-1677-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H749085263001Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page