Skip to content

Ultraflex™ Tracheobronchial M00576490Recalls & reported problems

Boston Scientific Corporation · Proximal Release Stent System · Class 2 · Implant

All models of Ultraflex Tracheobronchial (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729837855), received between Nov 28, 2016 and May 5, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Activation, Positioning or Separation Problem1
  • Break1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 5, 2021Break; Positioning Failure; Break; Positioning Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON APRIL 13, 2021 THAT AN ULTRAFLEX TRACHEOBRONCHIAL UNCOVERED PROXIMAL RELEASE STENT WAS TO BE IMPLANTED TO TREAT A MALIGNANT CENTRAL OBSTRUCTION IN THE LEFT LUNG DURING A BRONCHOSCOPY WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2021. REPORTEDLY, THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE BLACK STENT DEPLOYMENT SUTURE BROKE AND THE STENT WAS NOT DEPLOYED. THE STENT WAS FULLY COVERED BY THE DEPLOYMENT SUTURE WHEN IT WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER ULTRAFLEX TRACHEOBRONCHIAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11776770 ↗
  • MalfunctionNov 28, 2016Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT 3005099803-2016-03620 AND MANUFACTURER REPORT 3005099803-2016-03621 FOR THE ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON NOVEMBER 03, 2016 THAT TWO ULTRAFLEX TRACHEOBRONCHIAL STENTS WERE TO BE USED IN THE LUNGS TO TREAT A MALIGNANT STRICTURE DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2016. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, BOTH STENTS WOULD NOT FULLY DEPLOY. THE PARTIALLY DEPLOYED STENTS WERE REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER ULTRAFLEX TRACHEOBRONCHIAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6130300 ↗

Other Prosthesis, tracheal, expandable devices

All Prosthesis, tracheal, expandable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page