Ultraflex™ Tracheobronchial M00576670Recalls & reported problems
Boston Scientific Corporation · Distal Release Stent System · Class 2 · Implant
All models of Ultraflex Tracheobronchial (25)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729842477), received between May 10, 2021 and Jul 6, 2021 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Difficult to Advance” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available2
- Difficult to Advance2
- Material Deformation1
- Positioning Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 6, 2021Difficult to Advance; Appropriate Device Problem Term/Code Not Available; Difficult to Advance; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON JUNE 14, 2021 THAT AN ULTRAFLEX TRACHEOBRONCHIAL COVERED DISTAL RELEASE STENT WAS TO BE IMPLANTED TO TREAT A STRICTURE DURING A BRONCHOSCOPY WITH STENT PLACEMENT PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE STENT WOULD NOT ADVANCE OVER THE GUIDEWIRE. THE PROCEDURE WAS CANCELLED BECAUSE ANOTHER STENT OF THE SAME SIZE WAS NOT AVAILABLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12124052 ↗MalfunctionJul 6, 2021Difficult to Advance; Appropriate Device Problem Term/Code Not Available; Difficult to Advance; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2021 THAT AN ULTRAFLEX TRACHEOBRONCHIAL COVERED DISTAL RELEASE STENT WAS TO BE IMPLANTED TO TREAT A STRICTURE DURING A BRONCHOSCOPY WITH STENT PLACEMENT PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE STENT WOULD NOT ADVANCE OVER THE GUIDEWIRE. THE PROCEDURE WAS CANCELLED BECAUSE ANOTHER STENT OF THE SAME SIZE WAS NOT AVAILABLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12123131 ↗InjuryMay 10, 2021Material Deformation; Positioning Problem; Material Deformation; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON APRIL 20, 2021 THAT AN ULTRAFLEX TRACHEOBRONCHIAL COVERED DISTAL STENT WAS TO BE IMPLANTED TO TREAT A STRICTURE IN THE TRACHEA DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE STENT WAS DEPLOYED TOO DISTAL IN THE TRACHEA. THE PHYSICIAN PULLED THE STENT RETRIEVAL SUTURE WITH A PAIR OF RESCUE FORCEPS AND MOVED THE STENT INTO ITS CORRECT POSITION. HOWEVER, AFTER THE STENT WAS REPOSITIONED, IT WAS NOTICED THAT THE STENT KINKED. THE PHYSICIAN WAS UNABLE TO FLATTEN THE STENT WITH THE SCOPE TIP AND FORCEPS. THE STENT WAS REMOVED FROM THE PATIENT WITH RESCUE FORCEPS AND A DIFFERENT STENT WAS IMPLANTED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11797883 ↗
Other Prosthesis, tracheal, expandable devices
- iCAST 85453Atrium Medical Corporation · Prosthesis, tracheal, expandable43 reports
- iCAST 85419Atrium Medical Corporation · Prosthesis, tracheal, expandable13 reports
- WALLSTENT™ BILIARY H965430500Boston Scientific Corporation · Prosthesis, tracheal, expandable13 reports
- iCAST 85445Atrium Medical Corporation · Prosthesis, tracheal, expandable12 reports
- iCAST 85416Atrium Medical Corporation · Prosthesis, tracheal, expandable12 reports
- WALLSTENT® Endoprosthesis H965404130Boston Scientific Corporation · Prosthesis, tracheal, expandable11 reports
- iCAST 85455Atrium Medical Corporation · Prosthesis, tracheal, expandable11 reports
- iCAST 85414Atrium Medical Corporation · Prosthesis, tracheal, expandable10 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗