EndoVive Safety PEG Kit M00509010Recalls & reported problems
Boston Scientific Corporation · Percutaneous Endoscopic Gastrostomy Kit ENFit · Class 2
All models of EndoVive Safety PEG Kit (8)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 08714729880400), received between Nov 3, 2020 and Feb 5, 2025 — classified as 1 injury, 15 malfunctions.
The most frequently reported specific device problem is “Obstruction of Flow” (11 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Feb 20, 2024Class IIOpen, Classified
Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration
Z-1513-2024 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Obstruction of Flow11
- Detachment of Device or Device Component3
- Break1
- Defective Device1
- Difficult to Advance1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions15
- No Code Available1
- Perforation1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE PUSH PEG KIT BROKE AT THE CONNECTION AS THEY WERE PULLING IT THROUGH THE SKIN. THEY WERE ABLE TO GET A HOLD OF THE PLASTIC PIECE AND GRADUALLY WORKED IT THROUGH THE INCISION. IT WAS REPORTED THAT THE INCISION WAS ENLARGED. THE PLASTIC PIECE HOWEVER BROKE AGAIN IN TWO DIFFERENT PLACES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21318331 ↗MalfunctionJul 8, 2024Obstruction of Flow; Obstruction of Flow
NOTE: THIS REPORT PERTAINS TO THREE ENDOVIVE SAFETY PEG KIT USED IN THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT WAS ATTEMPTED TO BE USED DURING PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) TUBE PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE WAS NOT ABLE TO ADVANCE THRU THE PEG TUBE DUE TO AN OCCLUSION. THE CUSTOMER REPORTED THAT TWO ADDITIONAL PEG KITS WERE OPENED AND ATTEMPTED TO BE USED BUT THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19690832 ↗MalfunctionJul 8, 2024Obstruction of Flow; Obstruction of Flow
NOTE: THIS REPORT PERTAINS TO THREE ENDOVIVE SAFETY PEG KIT USED IN THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT WAS ATTEMPTED TO BE USED DURING PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) TUBE PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE WAS NOT ABLE TO ADVANCE THRU THE PEG TUBE DUE TO AN OCCLUSION. THE CUSTOMER REPORTED THAT TWO ADDITIONAL PEG KITS WERE OPENED AND ATTEMPTED TO BE USED BUT THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19690786 ↗MalfunctionJul 5, 2024Obstruction of Flow; Obstruction of Flow
NOTE: THIS REPORT PERTAINS TO THREE ENDOVIVE SAFETY PEG KIT USED IN THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT WAS ATTEMPTED TO BE USED DURING PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) TUBE PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE WAS NOT ABLE TO ADVANCE THRU THE PEG TUBE DUE TO AN OCCLUSION. THE CUSTOMER REPORTED THAT TWO ADDITIONAL PEG KITS WERE OPENED AND ATTEMPTED TO BE USED BUT THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19684475 ↗MalfunctionMay 20, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS ATTEMPTED TO BE USED DURING PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) TUBE PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING INSERTION, THE TUBE BROKE APART AT THE MIDDLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19354875 ↗MalfunctionMay 2, 2024Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS ATTEMPTED TO BE USED DURING PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) TUBE PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE PHYSICIAN WAS UNABLE TO PASS THE WIRE THROUGH THE INTRODUCER. IT WAS REPORTED THAT THE TIP OF THE INTRODUCER WAS CLOGGED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19230451 ↗MalfunctionFeb 28, 2024Obstruction of Flow; Obstruction of Flow
NOTE: THIS REPORT PERTAINS TO THE THIRD OF THREE ENDOVIVE SAFETY PEG KITS PUSH METHOD THAT WERE USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS ATTEMPTED TO BE USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE COULD NOT PASS THROUGH THE TRANSITION ZONE BETWEEN THE CONNECTOR, BETWEEN THE HARD AND SOFT PLASTIC. THE CUSTOMER REPORTED THAT TWO ADDITIONAL PEG KITS WERE OPENED AND ATTEMPTED TO BE USED BUT THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS COMPLETED WITH A FOURTH ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18800174 ↗MalfunctionFeb 28, 2024Obstruction of Flow; Obstruction of Flow
NOTE: THIS REPORT PERTAINS TO THE SECOND OF THREE ENDOVIVE SAFETY PEG KITS PUSH METHOD THAT WERE USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS ATTEMPTED TO BE USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE COULD NOT PASS THROUGH THE TRANSITION ZONE BETWEEN THE CONNECTOR, BETWEEN THE HARD AND SOFT PLASTIC. THE CUSTOMER REPORTED THAT TWO ADDITIONAL PEG KITS WERE OPENED AND ATTEMPTED TO BE USED BUT THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS COMPLETED WITH A FOURTH ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18800154 ↗MalfunctionFeb 28, 2024Obstruction of Flow; Obstruction of Flow
NOTE: THIS REPORT PERTAINS TO THE FIRST OF THREE ENDOVIVE SAFETY PEG KITS PUSH METHOD THAT WERE USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS ATTEMPTED TO BE USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE COULD NOT PASS THROUGH THE TRANSITION ZONE BETWEEN THE CONNECTOR, BETWEEN THE HARD AND SOFT PLASTIC. THE CUSTOMER REPORTED THAT TWO ADDITIONAL PEG KITS WERE OPENED AND ATTEMPTED TO BE USED BUT THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS COMPLETED WITH A FOURTH ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18799508 ↗MalfunctionFeb 15, 2024Obstruction of Flow; Obstruction of Flow
NOTE: THIS REPORT PERTAINS TO THE SECOND OF TWO ENDOVIVE SAFETY PEG KITS PUSH METHOD THAT WERE USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE COULD NOT PASS THROUGH THE TRANSITION ZONE BETWEEN THE CONNECTOR, BETWEEN THE HARD AND SOFT PLASTIC. ANOTHER PEG KIT WAS USED; HOWEVER, THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS COMPLETED WITH A THIRD ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18716080 ↗MalfunctionFeb 15, 2024Obstruction of Flow; Obstruction of Flow
NOTE: THIS REPORT PERTAINS TO THE FIRST OF TWO ENDOVIVE SAFETY PEG KITS PUSH METHOD THAT WERE USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE COULD NOT PASS THROUGH THE TRANSITION ZONE BETWEEN THE CONNECTOR, BETWEEN THE HARD AND SOFT PLASTIC. ANOTHER PEG KIT WAS USED; HOWEVER, THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS COMPLETED WITH A THIRD ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18716071 ↗MalfunctionFeb 15, 2024Obstruction of Flow; Obstruction of Flow
NOTE: THIS REPORT PERTAINS TO THE SECOND OF TWO ENDOVIVE SAFETY PEG KITS PUSH METHOD THAT WERE USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE COULD NOT PASS THROUGH THE TRANSITION ZONE BETWEEN THE CONNECTOR, BETWEEN THE HARD AND SOFT PLASTIC. ANOTHER PEG KIT WAS USED; HOWEVER, THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS COMPLETED WITH A THIRD ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18715536 ↗MalfunctionFeb 15, 2024Obstruction of Flow; Obstruction of Flow
NOTE: THIS REPORT PERTAINS TO THE FIRST OF TWO ENDOVIVE SAFETY PEG KITS PUSH METHOD THAT WERE USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE COULD NOT PASS THROUGH THE TRANSITION ZONE BETWEEN THE CONNECTOR, BETWEEN THE HARD AND SOFT PLASTIC. ANOTHER PEG KIT WAS USED; HOWEVER, THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS COMPLETED WITH A THIRD ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18715209 ↗MalfunctionSep 21, 2023Break; Defective Device; Break; Defective Device
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PLACEMENT PROCEDURE PERFORMED ON (B)(6), 2023. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF A LOOP WIRE DISTAL END SECTION OF THE FEEDING TUBE WAS DETACHED. PLEASE SEE BLOCK H10 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17787638 ↗MalfunctionJan 26, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED DURING AN ESOPHAGOGASTRODUODENOSCOPY WITH PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PLACEMENT PROCEDURE ON (B)(6) 2022. DURING THE PROCEDURE, THE PEG TUBE DETACHED AT THE TRANSITION ZONE BETWEEN THE HARD AND SOFT PLASTIC. THE PROCEDURE WAS COMPLETED WITH A NEW ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13360276 ↗InjuryNov 3, 2020Use of Device Problem; Difficult to Advance; Use of Device Problem; Difficult to Advance
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ENDOVIVE SAFETY PEG KIT PULL METHOD USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE PEG TUBE WAS DIFFICULT TO PULL THROUGH THE STOMA. THE PHYSICIAN INTRODUCED THE TROCAR HOWEVER AN INITIAL INCISION WAS NOT DONE FIRST. REPORTEDLY, PROBLEMS BEGAN WHEN INCISIONS WERE MADE ALONGSIDE THE TUBING WHILE TRYING TO PULL THE TUBE THROUGH THE SITE. THIS CREATED SEVERAL MICRO PERFORATIONS. THE PATIENT WENT TO SURGERY. PER PHYSICIAN ASSESSMENT, THE PEG TUBE CONTRIBUTE TO THE MICRO PERFORATIONS. THE IFU (INSTRUCTIONS FOR USE) ENDOVIVE SAFETY PEG KIT PULL METHOD FOURTH STEP STATES THE FOLLOWING: USING THE EXPOSED BLADE OF THE SAFETY SCALPEL PROVIDED, MAKE A 1 TO 1.5 CM INCISION AT THE SELECTED INCISION SITE.
Patient problems: No Code Available; Perforation
MAUDE report 10777112 ↗
FDA clearance and classification
The FDA regulates product code PIF (gastrointestinal tubes with enteral specific connectors) under 21 CFR 876.5980 as a Class II device. 21 CFR 876.5980 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the EndoVive Safety PEG Kit M00509010
Has the EndoVive Safety PEG Kit M00509010 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1513-2024 (Class II), began on Feb 20, 2024: Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the EndoVive Safety PEG Kit M00509010?
The FDA's MAUDE database lists 16 adverse event reports for this device, received between Nov 3, 2020 and Feb 5, 2025. In those reports, the most common specific device problems are “Obstruction of Flow” (11), “Detachment of Device or Device Component” (3) and “Break” (1); the most common patient problem is “Perforation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EndoVive Safety PEG Kit M00509010?
None of the reports are classified as deaths, 1 as an injury and 15 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the EndoVive Safety PEG Kit M00509010, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand EndoVive Safety PEG Kit, model M00509010). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Gastrointestinal Tubes with Enteral Specific Connectors devices
- EndoVive Safety PEG Kit M00509000Boston Scientific Corporation · Gastrointestinal Tubes with Enteral Specific Connectors1 recall30 reports
- EndoVive Standard PEG Kit M00509050Boston Scientific Corporation · Gastrointestinal Tubes with Enteral Specific Connectors1 recall6 reports
- Salem Sump 8888264986ECardinal Health 200, LLC · Gastrointestinal Tubes with Enteral Specific Connectors1 recall1 death2 reports
- Salem Sump 8888268060ECardinal Health 200, LLC · Gastrointestinal Tubes with Enteral Specific Connectors1 recall0 reports
- Salem Sump 7771810ECardinal Health, Inc. · Gastrointestinal Tubes with Enteral Specific Connectors28 reports
- Salem Sump 7771610ECardinal Health, Inc. · Gastrointestinal Tubes with Enteral Specific Connectors1 death18 reports
- Edi Catheter ENFit 6Fr/49 cm, 5pcs 6883898Maquet Critical Care AB · Gastrointestinal Tubes With Enteral Specific Connectors14 reports
- MEDLINE ENFIT9000SMedline Industries, Inc. · Gastrointestinal Tubes with Enteral Specific Connectors14 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice