Skip to content

EndoVive Standard PEG Kit M00509050Recalls & reported problems

Boston Scientific Corporation · Percutaneous Endoscopic Gastrostomy Kit ENFit · Class 2

All models of EndoVive Standard PEG Kit (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
5
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729880486), received between Apr 30, 2021 and Apr 29, 2024 — classified as 5 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 20, 2024Open, Classified

    Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration

    Z-1514-2024 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Break1
  • Detachment of Device or Device Component1

Patient problems

  • Unspecified Infection4
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 29, 2024Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG KIT PUSH METHOD WAS ATTEMPTED TO BE USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) TUBE PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE SNAPPED INSIDE THE PATIENT. PART OF THE WIRE DETACHED, AND THE DETACHED PIECE WAS SUCCESSFULLY RETRIEVED USING A SNARE DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER STANDARD PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19203292 ↗
  • MalfunctionJul 7, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG KIT PUSH METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PLACEMENT PROCEDURE ON (B)(6) 2022. DURING PREPARATION, THE PEG TUBE DETACHED. THE PROCEDURE WAS COMPLETED WITH A NEW ENDOVIVE STANDARD PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14945872 ↗
  • InjuryApr 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG PUSH METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PROCEDURE. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, THE PATIENT WENT TO EMERGENCY ROOM FOR AN INCISION SITE INFECTION AND WAS TREATED WITH ANTIBIOTICS. NO ADDITIONAL INFORMATION HAS BEEN ABLE TO BE OBTAINED DESPITE GOOD FAITH EFFORTS.

    Patient problems: Unspecified Infection

    MAUDE report 11757233 ↗
  • InjuryApr 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG PUSH METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PROCEDURE. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, THE PATIENT WENT TO EMERGENCY ROOM FOR AN INCISION SITE INFECTION AND WAS TREATED WITH ANTIBIOTICS. NO ADDITIONAL INFORMATION HAS BEEN ABLE TO BE OBTAINED DESPITE GOOD FAITH EFFORTS.

    Patient problems: Unspecified Infection

    MAUDE report 11757204 ↗
  • InjuryApr 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG PUSH METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PROCEDURE. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, THE PATIENT WENT TO EMERGENCY ROOM FOR AN INCISION SITE INFECTION AND WAS TREATED WITH ANTIBIOTICS. NO ADDITIONAL INFORMATION HAS BEEN ABLE TO BE OBTAINED DESPITE GOOD FAITH EFFORTS.

    Patient problems: Unspecified Infection

    MAUDE report 11757198 ↗
  • InjuryApr 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG PUSH METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PROCEDURE. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, THE PATIENT WENT TO EMERGENCY ROOM FOR AN INCISION SITE INFECTION AND WAS TREATED WITH ANTIBIOTICS. NO ADDITIONAL INFORMATION HAS BEEN ABLE TO BE OBTAINED DESPITE GOOD FAITH EFFORTS.

    Patient problems: Unspecified Infection

    MAUDE report 11756649 ↗

Other Gastrointestinal Tubes with Enteral Specific Connectors devices

All Gastrointestinal Tubes with Enteral Specific Connectors devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page