Skip to content

Marvel™ H749393013000Recalls & reported problems

Boston Scientific Corporation · Guidewire · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729887805), received between Jan 9, 2020 and Aug 14, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Material Integrity Problem1

Patient problems

  • Foreign Body In Patient1
  • Injury1
  • Low Blood Pressure/ Hypotension1
  • Respiratory Failure1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 14, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT GUIDEWIRE DETACHMENT OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND CALCIFIED PROXIMAL ARTERY. A 300CM MARVEL GUIDEWIRE WAS ADVANCED AND ANGIOPLASTY AND STENTING WAS PERFORMED. UPON REMOVAL OF THE GUIDEWIRE, IT WAS NOTICED THAT THE DISTAL WIRE TIP WAS LEFT IN THE PROXIMAL ARTERY. ANOTHER STENT WAS IMPLANTED TO HOLD THE WIRE FRAGMENT INTO THE ARTERY WALL SO THAT IT WOULD NOT MIGRATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 19988496 ↗
  • InjuryJan 9, 2020Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT PATIENT EXPERIENCED HYPOTENSION AND WIRE COATING PEELED OFF. THE TARGET LESION WAS LOCATED IN A CORONARY ARTERY. AFTER ADVANCING THE 300CM MARVEL GUIDEWIRE, A 3.00 X 20MM SYNERGY II DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, THE PATIENT EXPERIENCED LOW BLOOD PRESSURE. THE STENT WAS REMOVED AND THE WIRE WAS REPOSITIONED. IT WAS THEN NOTICED THAT THE STENT WAS UNABLE TO ADVANCE OVER THE WIRE AND THE TIP WAS DRAGGING THE COATING MATERIAL OFF THE WIRE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE AND NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Injury; Low Blood Pressure/ Hypotension; Respiratory Failure

    MAUDE report 9570730 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page