Skip to content

WallFlex™ Duodenal M00555860Recalls & reported problems

Boston Scientific Corporation · Stent System with Anchor Lock Delivery System · Class 2 · Implant

All models of WallFlex Duodenal (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729902492), received between Jun 23, 2022 and Jul 27, 2023 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Migration1
  • Premature Activation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 27, 2023Migration; Migration

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6)2023 THAT A WALLFEX DUODENAL UNCOVERED STENT WAS IMPLANTED TO TREAT A GASTRODUODENAL STRICTURE DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6)2023. ON (B)(6) 2023, POST STENT PLACEMENT, IT WAS NOTED THAT THE STENT HAD MIGRATED. THERE ARE NO KNOWN PATIENT COMPLICATIONS DUE TO THIS EVENT. NOTE: BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17410800 ↗
  • InjuryJun 23, 2022Premature Activation; Premature Activation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2022 THAT A WALLFLEX DUODENAL STENT WAS TO BE IMPLANTED IN THE FIRST PART OF THE DUODENUM TO TREAT A 10 MM OVER 2 CM MALIGNANT DUODENAL STRICTURE DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2022. THE PATIENT'S ANATOMY WAS NOT TORTUOUS. DURING THE PROCEDURE, WHEN THE WALLFLEX DUODENAL STENT DELIVERY SYSTEM WAS DEPLOYED HALFWAY, THERE WAS NO MOVEMENT OF THE OUTER SHEATH THAT RELEASES THE STENT. THEN THE WALLFLEX DUODENAL STENT SUDDENLY DEPLOYED PREMATURELY. THE STENT WAS REMOVED FROM THE PATIENT WITH THE USE OF RESCUE COMBO FORCEPS. ANOTHER WALLFLEX DUODENAL STENT WAS IMPLANTED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14786571 ↗

Other Stent, metallic, expandable, duodenal devices

All Stent, metallic, expandable, duodenal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page