Skip to content

FATHOM™ -16 M001509200Recalls & reported problems

Boston Scientific Corporation · Steerable Guidewire · Class 2

All models of FATHOM -16 (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729961116), received between Dec 8, 2022 and Nov 14, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 14, 2024Fracture; Fracture

    IT WAS REPORTED THAT GUIDEWIRE FRACTURE OCCURRED REQUIRING ADDITIONAL INTERVENTION. THE TARGET LESION WAS LOCATED IN A SEVERELY TORTUOUS AND MILDLY CALCIFIED POSTERIOR SEGMENTAL BRANCH OFF THE RENAL ARTERY. A 200X25CM FATHOM -16 GUIDEWIRE WAS SELECTED FOR USE. DURING REMOVAL, IT WAS NOTED THAT THE PART OF THE WIRE BROKE OFF INSIDE THE CATHETER AND AN ESTIMATED 10CM PORTION OF THE WIRE BROKE INSIDE THE PATIENT. A HEADLINER NON-BOSTON SCIENTIFIC WIRE WAS USED TO WEDGE IT IN BETWEEN THE FRAGMENTED FATHOM AND WAS ABLE TO LATCH ONTO THE FATHOM WIRE AND DRAG WHAT WAS INSIDE THE PATIENT OUT WITH THE HEADLINER THROUGH THE SAME NON-BOSTON SCIENTIFIC MICROCATHETER. CASE WAS COMPLETED SUCCESSFULLY, AND NOTHING WAS LEFT IN PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20689347 ↗
  • InjuryDec 8, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE FATHOM WIRE SNAPPED, FELL OFF IN THE ARTERY AND WAS REMOVED BY SNARING. THE NON-STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND NON-CALCIFIED FEMORAL ARTERY. A 200X25CM FATHOM -16 WAS SELECTED FOR USE. DURING THE PROCEDURE, THE FATHOM WIRE SNAPPED WHILE NAVIGATING INSIDE THE ARTERY. AS THE PHYSICIAN PULLED BACK ON WIRE, IT SNAPPED AND FELL OFF IN THE FEMORAL ARTERY. THE PIECE WAS SNARED OUT OF THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15941165 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →