FATHOM™ -16 M001509210Recalls & reported problems

Boston Scientific Corporation · Steerable Guidewire · Class 2

All models of FATHOM -16 (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729961123), received between May 13, 2019 and May 2, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Material Integrity Problem1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 2, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT TIP DETACHMENT OCCURRED VIA ATTACHED VOLUNTARY MEDWATCH REPORT # (B)(4). THE STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS VESSEL. A FATHOM -16 GUIDEWIRE WAS SELECTED FOR USE IN A VISCERAL ANGIOGRAM. DURING THE PROCEDURE, IT WAS NOTED THAT THE TIP OF THE WIRE INADVERTENTLY SHEARED OFF INTO THE BRANCH OF SUPERIOR MESENTERIC ARTERY. THE TIP OF THE WIRE WAS COMPLETELY REMOVED WITH A GOOSENECK SNARE. A REPEAT ANGIOGRAM WAS PERFORMED AND CONFIRMED THE RESOLUTION OF HEMORRHAGE. PATIENT WAS STABLE AND SCHEDULED FOR A COLONOSCOPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21954315 ↗
  • InjuryApr 27, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE GUIDEWIRE WAS FRACTURED AND REMAINED INSIDE THE PATIENT. THE NON-STENOSED TARGET LESION WAS LOCATED IN THE TORTUOUS AND NON-CALCIFIED VESSEL IN THE ABDOMEN. A 200X25CM ANGLED FATHOM -16 GUIDEWIRE WAS SELECTED FOR USE IN AN ENDOLEAK PROCEDURE. DURING THE PROCEDURE, THE DEVICE WAS USED TO NAVIGATE THE VESSEL. AFTER MULTIPLE PASSES, IT WAS NOTED THAT THE GUIDEWIRE FRACTURED AT THE ILEOLUMBUS BRANCH. AS THE FRACTURE OCCURRED IN A VESSEL THAT WAS PLANNED TO BE EMBOLIZED, THE FRACTURED PORTION WAS LEFT INSIDE THE VESSEL. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 14229294 ↗
  • InjuryMay 13, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE TIP OF THE GUIDEWIRE BROKE AND REMAINED INSIDE THE PATIENT. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE VERY TORTUOUS AND SEVERELY CALCIFIED LIVER. A 200X25CM ANGLED FATHOM -16 GUIDEWIRE WAS SELECTED FOR USE IN AN EMBOLISATION PROCEDURE TO STOP BLEEDING. DURING PROCEDURE, IT WAS NOTED THAT THE TIP OF THE WIRE CAME OFF. STRONG ATTEMPT WERE MADE TO REMOVE THE FRACTURED TIP WITHOUT SUCCESS. HOWEVER, THE REST OF THE WIRE WAS RETRIEVED FROM THE PATIENT'S BODY. THE BLEEDING THEN SEEMED TO HAVE STOPPED, THUS THE PROCEDURE WAS ABANDONED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS FINE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8604884 ↗