JUDO™ 3 H749393771900Recalls & reported problems
Boston Scientific Corporation · Guidewire · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729974949), received between Nov 25, 2019 and Dec 22, 2023 — classified as 4 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Difficult to Remove2
- Device-Device Incompatibility1
- Difficult to Advance1
- Fracture1
- Material Deformation1
- Unraveled Material1
Patient problems
- Device Embedded In Tissue or Plaque2
- Vasoconstriction2
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryDec 22, 2023Difficult to Advance; Difficult to Advance
IT WAS REPORTED THAT A VASOSPASM OCCURRED AND REQUIRED MEDICATION. A 190 CM JUDO 3 GUIDEWIRE WAS SELECTED FOR USE. DURING THE PROCEDURE, THE GUIDEWIRE ADVANCED TO THE GUIDING CATHETER, BUT IT FAILED TO CROSS SMOOTHLY, AND VASOSPASM OCCURRED. VASOSPASM WAS TREATED WITH NITROGLYCERIN. THE PROCEDURE WAS COMPLETED WITH A NON-BOSTON SCIENTIFIC GUIDEWIRE DEVICE. THE PATIENT RECOVERED AND WAS GOOD POST PROCEDURE.
Patient problems: Vasoconstriction
MAUDE report 18383682 ↗InjuryNov 29, 2023Fracture; Fracture
IT WAS REPORTED THAT GUIDEWIRE FRACTURE OCCURRED. THE OVER 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY (LAD). A 190 CM, STRAIGHT TIP, JUDO 3 GUIDEWIRE WAS ADVANCED FOR TREATMENT. HOWEVER, WHEN THE DEVICE WAS WITHDRAWN FROM THE PATIENT'S BODY, THE SPRING COIL SECTION OF THE GUIDEWIRE FRACTURED AROUND 180CM. THE DEVICE WAS REMOVED USING GUIDEZILLA (GZ), LOWER BALLOON WITH LOW PRESSURE DILATION AND RETRACTION REMOVAL. THE PROCEDURE WAS COMPLETED WITH A NON-BOSTON SCIENTIFIC (BSC) DEVICE. THERE WAS NO PATIENT COMPLICATION REPORTED AND THE PATIENT WAS IN GOOD CONDITION. IT WAS FURTHER REPORTED THAT THERE ARE NO OBVIOUS COMPLICATIONS, AND THE IMMEDIATE MANIFESTATION WAS VASOSPASM.
Patient problems: No Clinical Signs, Symptoms or Conditions; Vasoconstriction
MAUDE report 18226917 ↗InjuryJan 27, 2021Difficult to Remove; Unraveled Material; Detachment of Device or Device Component; Difficult to Remove; Unraveled Material; Detachment of Device or Device Component
IT WAS REPORTED THAT REMOVAL DIFFICULTIES WERE ENCOUNTERED AND DISTAL TIP DETACHMENT OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND CALCIFIED RIGHT CORONARY ARTERY. A 3, 190 CM JUDO 3 GUIDE WIRE WAS ADVANCED TO CROSS THE LESION. HOWEVER, THE WIRE BECAME STUCK IN THE SUBINTIMAL PLAQUE AND THE DISTAL PORTION UNRAVELED. THE PHYSICIAN TRIED TO RETRIEVE THE DISTAL TIP BY BALLOONING PROXIMALLY BUT EVERYTHING CAME OUT AND THE RADIOPAQUE PORTION BROKE OFF. THE DETACHED COMPONENT HAS BEEN STENTED ACROSS AND BURIED IN CALCIUM. THE PROCEDURE WAS COMPLETED BY GOING RETROGRADE AND THE SHEARED OFF PORTION OF THE WIRE REMAINS IN THE RIGHT CORONARY ARTERY BEHIND A STENT. NO COMPLICATIONS WERE REPORTED AND PATIENT IS STABLE.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 11238795 ↗InjuryNov 25, 2019Difficult to Remove; Detachment of Device or Device Component; Device-Device Incompatibility; Material Deformation; Difficult to Remove; Detachment of Device or Device Component; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT THE TIP OF THE GUIDEWIRE DETACHED. A PERCUTANEOUS CORONARY INTERVENTION WAS BEING PERFORMED ON A CTO LESION IN THE CIRCUMFLEX CORONARY ARTERY. A JUDO 3 GUIDEWIRE WAS ABLE TO CROSS THE LESION BUT THE WIRE PROLAPSED AND BECAME TANGLED. THE BODY OF THE GUIDEWIRE WAS INSIDE A MICROCATHETER, BUT THE TIP OF THE GUIDEWIRE WAS OUTSIDE OF THE MICROCATHETER. THE GUIDEWIRE WAS UNABLE TO BE PULLED INTO THE MICROCATHETER AND AS THEY WERE REMOVING EVERYTHING FROM THE BODY, THE GUIDEWIRE TIP GOT STUCK ON THE PREVIOUSLY PLACED STENT STRUT AND IT SNAPPED OFF. THE GUIDEWIRE TIP WAS LEFT INSIDE THE BODY. THE PHYSICIAN ATTEMPTED TO SNARE THE TIP OF THE GUIDEWIRE BUT WAS NOT ABLE TO DO SO. THE PHYSICIAN PLACED A SECOND STENT INSIDE THE PREVIOUSLY PLACED STENT SO THROMBUS WOULD NOT FORM ON THE GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS. THE PATIENT WAS FINE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9371581 ↗
Other Wire, guide, catheter devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page