- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729978831), received between Apr 18, 2019 and May 14, 2021 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 14, 2021Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT GUIDEWIRE TIP DETACHMENT OCCURRED. THE CHRONIC TOTALLY OCCLUDED TARGET LESION WITH MODERATE TO SEVERE CALCIFICATION WAS LOCATED IN THE MODERATELY TORTUOUS ANTERIOR TIBIAL ARTERY (ATA). THE VESSEL WAS VERY DISEASED AND HAD SEVERAL OCCLUSIONS DOWN THE VESSEL BEFORE REFILLING CLOSE TO THE FOOT. A SAVION FLX GUIDEWIRE WAS PREPPED AND ADVANCED THROUGH A 2.6F NON-BSC SUPPORT CATHETER. WITHIN A MINUTE, THE TIP OF THE GUIDEWIRE BROKE OFF APPROXIMATELY HALFWAY DOWN THE ATA AND BECAME LODGED IN SITU. THE REMAINDER OF THE SAVION FLX GUIDEWIRE WAS REMOVED, AND THEN A NON-BSC GUIDEWIRE WAS USED WITHOUT ISSUE. THE CASE WAS ABANDONED. THE PATIENT WAS STABLE. AN ATTEMPT WILL BE MADE AGAIN SHOULD THE BLOOD FLOW TO THE TOE AMPUTATION WOUND SITE NOT BE SUFFICIENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 11828352 ↗InjurySep 3, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE GUIDEWIRE TIP DETACHED INSIDE THE PATIENT. A SAVION FLX GUIDEWIRE WAS SELECTED FOR USE IN A PROCEDURE WITH RETROGRADE ACCESS AND A 5F SHEATH. DURING THE PROCEDURE, THE TIP OF THE GUIDEWIRE DETACHED IN THE DISTAL ANTERIOR TIBIAL ARTERY. THERE WAS NO WAY TO RETRIEVE THE GUIDEWIRE FRAGMENT. FLOW TO THE PATIENT'S ARTERY WAS NOT AFFECTED. IT WAS FURTHER REPORTED THAT THE LESION WAS EXTREMELY CALCIFIED AND STENOSIS WAS 90%. ATTEMPTS WERE MADE TO RETRIEVE THE DETACHED WIRE BY ASPIRATION AND USING A SNARE CATHETER, BUT IT WAS IRRETRIEVABLE. THE PROCEDURE WAS RESCHEDULED AND A STENT WAS USED TO AFFIX AND APPOSE THE DETACHED TIP AGAINST THE VESSEL WALL.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10488554 ↗MalfunctionApr 18, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A TIP SEPARATION OCCURRED. A SAVION FLX GUIDEWIRE WAS SELECTED FOR USE. DURING PREPARATION OUTSIDE THE PATIENT, THE PHYSICIAN ATTEMPTED TO GENTLY SHAPE THE DISTAL TIP. NO STRENGTH WAS USED AND NO RESISTANCE WAS FELT WHILE TRYING TO SHAPE THE TIP; HOWEVER THE DISTAL TIP BROKE AND TORE APART. THE PRODUCT WAS NOT USED INSIDE THE PATIENT AND THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 8528076 ↗
Other Wire, guide, catheter devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page