EMBOLD™ Fibered M001394240040080Recalls & reported problems

Boston Scientific Corporation · Detachable Coil System · Class 2 · Implant

All models of EMBOLD Fibered (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729983569), received between Jun 8, 2023 and Sep 1, 2026 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Difficult to Remove3
  • Stretched3
  • Difficult to Advance2
  • Adverse Event Without Identified Device or Use Problem1
  • Failure to Advance1
  • Malposition of Device1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient1
  • Vasoconstriction1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 1, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50551531 ↗
  • MalfunctionMar 24, 2025Failure to Advance; Adverse Event Without Identified Device or Use Problem; Failure to Advance; Adverse Event Without Identified Device or Use Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 24FEB2025. IT WAS REPORTED THAT DIFFICULTY ADVANCING OCCURRED. A 4X8 EMBOLD FIBERED WAS SELECTED FOR AN EMBOLIZATION PROCEDURE TO TREAT A PANCREATIC ANEURYSM. THE 4X8 COIL WAS ADVANCED; HOWEVER, THE PHYSICIAN DECIDED ON USING A SMALLER COIL FIRST. THE 4X8 COIL WAS REMOVED FROM THE PATIENT AND RE-HOUSED IN THE LOOP IN SALINE. THE COIL WAS PULLED OUT OF THE HOOP FOR LATER USE, BUT IT WOULD NOT ADVANCE THROUGH THE MICROCATHETER. A TOTAL OF 17 EMBOLD COILS WERE USED IN THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED COIL DETACHMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21677853 ↗
  • InjuryNov 20, 2023Difficult to Remove; Stretched; Difficult to Advance; Material Protrusion/Extrusion; Difficult to Remove; Stretched; Difficult to Advance; Material Protrusion/Extrusion

    IT WAS REPORTED THAT THE COILS WERE PROTRUDING FROM THE TARGET LOCATION, AND THERE WAS AN UNSUCCESSFUL SNARE ATTEMPT. THIS 4X8 EMBOLD FIBERED DETACHABLE COIL SYSTEM WAS SELECTED FOR USE IN AN EMBOLIZATION PROCEDURE. DURING THE PROCEDURE, TWO NON-BOSTON SCIENTIFIC COILS WERE USED, THEN THE PHYSICIAN WANTED A SMALLER COIL. THEREFORE, THIS EMBOLD COIL WAS USED AND AFTER PACKING, IT WAS PROTRUDING FROM THE TARGET LOCATION. IT WAS THEREFORE RETRACTED ONCE AND THEN PUSHED BACK IN; HOWEVER, IT BECAME STRETCHED. ADDITIONALLY, ISSUES OF DIFFICULT TO RETRACT/REMOVE AND DIFFICULT TO ADVANCE THROUGH THE GUIDE CATHETER WERE ALSO REPORTED. AN ATTEMPT WAS MADE TO RESHEATH THE COIL BUT WAS NOT SUCCESSFUL. ADDITIONALLY, AN ATTEMPT WAS MADE TO SNARE THE COIL; HOWEVER, BECAUSE THE COIL WAS SO STRETCHED THE FILAMENT MADE IT DIFFICULT TO SNARE, AND IT WAS LEFT IN PLACE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18172947 ↗
  • MalfunctionOct 26, 2023Break; Difficult to Remove; Stretched; Difficult to Advance; Break; Difficult to Remove; Stretched; Difficult to Advance

    IT WAS REPORTED THAT THE COIL FRACTURED INTERNALLY, AND ADDITIONAL INTERVENTION WAS REQUIRED TO RETRIEVE A PORTION OF THE COIL. AN EMBOLD FIBERED 4X8 WAS SELECTED FOR USE FOR AN EMBOLIZATION PROCEDURE IN THE GASTRODUODENAL ARTERY (GDA). THE COIL WAS DIFFICULT TO RETRACT INTO THE MICROCATHETER AND STRETCHED AS A RESULT. THE COIL FRACTURED, AND ADDITIONAL INTERVENTION WAS REQUIRED TO REMOVE THE COIL. A PORTION OF THE FRACTURED COIL REMAINED IMPLANTED IN THE PATIENT. THE PATIENT WAS EXPECTED TO FULLY RECOVER, AND THE EMBOLIZATION PROCEDURE WAS REPORTED TO HAVE BEEN SUCCESSFULLY COMPLETED. IT WAS FURTHER REPORTED THAT THE TARGET ANEURYSM WAS ACCESSED USING A .016 FATHOM GUIDEWIRE AND AN UNKNOWN .021 STC MICROCATHETER. THE MICROCATHETER WAS FLUSHED WITH SALINE, AND THERE WERE NO ISSUES LOADING THE COIL INTO THE MICROCATHETER. AS THE COIL WAS ADVANCING WITHIN APPROXIMATELY THE LAST 20 CM OF THE MICROCATHETER, THERE WAS INCREASING RESISTANCE AND FRICTION. THE PHYSICIAN ELECTED TO REMOVE THE COIL FROM THE MICROCATHETER AND OBSERVED THAT THE COIL HAD BECOME UNRAVELED. THERE WAS A LONG, THIN, NON-RADIOPAQUE FILAMENT WITHIN THE CATHETER. THE ENTIRE MICROCATHETER WAS REMOVED AND REPLACED. THE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18012245 ↗
  • InjuryJun 8, 2023Break; Difficult to Remove; Stretched; Malposition of Device; Material Integrity Problem; Break; Difficult to Remove; Stretched; Malposition of Device; Material Integrity Problem

    IT WAS REPORTED THAT THE COIL BROKE AND WAS LEFT IN A NON-TARGET LOCATION. A 4X8 EMBOLD FIBERED DETACHABLE COIL SYSTEM WAS SELECTED FOR USE IN A GASTRODUODENAL ARTERY (GDA) EMBOLIZATION PROCEDURE TO TREAT A GI BLEED. THE TARGET VESSEL WAS NOT EXTREMELY TORTUOUS, BUT THERE WERE BRANCHES COMING OFF THE GDA ON THE PROXIMAL END THAT NEEDED TO BE EMBOLIZED AS WELL. THESE BRANCHES WERE SMALL AND AT A HARD ANGLE MAKING IT DIFFICULT TO CANNULATE. THEY WERE EVENTUALLY EMBOLIZED, BUT THERE WAS SOME SPASMING IN THE MAIN GDA AFTER MULTIPLE ATTEMPTS TO GET INTO THESE SMALL VESSELS. THIS COIL WAS BEING PLACED IN THE PROXIMAL ARTERY BUT NOT FORMING CORRECTLY. AFTER MULTIPLE ATTEMPTS, THE DISTAL END OF THE COIL HOOKED INTO A BRANCH AND ANOTHER ATTEMPT WAS MADE TO FORM IN THE MAIN ARTERY. COIL RETRACTION WAS ATTEMPTED, BUT THE COIL STRETCHED AND BROKE. A 6F SNARE WAS USED TO ATTEMPT TO SNARE THE STRETCHED COIL. IT WAS ALSO ATTEMPTED TO PUSH THE COIL WITH A PUSHER WIRE. AFTER THE FAILED ATTEMPTS TO RETRIEVE THE STRETCHED COIL, THE PROCEDURE WAS COMPLETED. THE STRETCHED COIL REMAINED IN A NON-TARGET LOCATION. THE COIL STARTED IN THE SMALL BRANCH OFF THE GDA AND WAS STRETCHED THROUGH THE COMMON HEPATI

    Patient problems: Foreign Body In Patient; Vasoconstriction

    MAUDE report 17086529 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →