EMBOLD™ Fibered M001394240100200Recalls & reported problems

Boston Scientific Corporation · Detachable Coil System · Class 2 · Implant

All models of EMBOLD Fibered (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729983637), received between Mar 5, 2025 and Jul 7, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Difficult to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance2
  • Material Protrusion/Extrusion2
  • Stretched2
  • Difficult or Delayed Separation1
  • Premature Activation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 7, 2026Premature Activation; Difficult to Advance

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 08JUN2026. IT WAS REPORTED THAT PREMATURE DEPLOYMENT OCCURRED. A 10X20 EMBOLD FIBERED WAS SELECTED FOR AN EMBOLIZATION PROCEDURE. THE COIL WAS DIFFICULT TO ADVANCE AND DEPLOYED IN THE CATHETER BEFORE 25 PERCENT OF THE COIL WAS DEPLOYED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. AN ALTERNATE DEVICE WAS USED TO COMPLETE THE PROCEDURE. HOWEVER, DEVICE ANALYSIS REVEALED COIL SEPARATION FROM THE DISTAL COUPLER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25738424 ↗
  • MalfunctionFeb 26, 2026Stretched; Material Protrusion/Extrusion; Difficult or Delayed Separation

    IT WAS REPORTED THAT THE DEVICE WAS PROTRUDING FROM THE TARGET LOCATION. THIS 10X20 EMBOLD? FIBERED DEVICE WAS SELECTED FOR USE DURING A SPLENIC ARTERY EMBOLIZATION. DURING THE PROCEDURE, THE COIL DID NOT DETACH FROM THE DELIVERY WIRE WHEN THE PROXIMAL RELEASE MECHANISM WAS ACTIVATED. THE COIL WAS AT THE TARGET LOCATION WITH A PIECE OF THE COIL THAT DID NOT FULLY DETACH. WHEN THE SYSTEM WAS BEING PULLED BACK, THE COIL PULLED BACK WITH IT. WITH TIME AND MANEUVERING THEY WERE ABLE TO DETACH THE COIL ENOUGH WHERE ONLY A SHORT PIECE WAS LEFT DANGLING. IT WAS ADDITIONALLY NOTED THAT THE COIL BECAME STRETCHED AS A RESULT OF THIS. A ROAD RUNNER GUIDEWIRE AND ENVOY GUIDEWIRE WERE USED TO BREAK THE FILAMENT/TETHER STRING FROM THE DISTAL END OF THE COIL. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24462584 ↗
  • MalfunctionMar 5, 2025Stretched; Difficult to Advance; Material Protrusion/Extrusion; Stretched; Difficult to Advance; Material Protrusion/Extrusion

    IT WAS REPORTED THAT THE COIL STRETCHED AND WAS PROTRUDING FROM THE TARGET LOCATION. TWO 10X20 EMBOLD FIBERED DEVICES WERE SELECTED FOR USE FOR AN EMBOLIZATION PROCEDURE. DURING THE PROCEDURE, UPON ADVANCING ONE OF THE COILS FORWARD THROUGH THE MICROCATHETER, THE COILS NORMAL RESISTANCE WHEN PUSHING FORWARD DISAPPEARED. IT APPEARED VIA IMAGING THAT THE COIL STRETCHED, AND THE COUPLER COULD BE SEEN AS WELL AS A LONG AREA OF STRETCHED COIL EXTENDING INTO AN AREA THEY DID NOT INTEND TO COIL. BOTH COILS REMAIN IMPLANTED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21527508 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →