EMBOLD™ Fibered M001394240200500Recalls & reported problems

Boston Scientific Corporation · Detachable Coil System · Class 2 · Implant

All models of EMBOLD Fibered (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729983736), received between Oct 15, 2025 and Sep 24, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Remove1
  • Failure to Advance1
  • Material Protrusion/Extrusion1
  • Stretched1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 24, 2026Device-Device Incompatibility; Material Protrusion/Extrusion

    It was reported that the previously deployed coil was pulled out of position, and intervention was required to correct its position.Two 20mm x 50cm EMBOLD? Fibered coils were selected for use in the embolization of a splenic artery aneurysm. Other non-Boston Scientific coils were also used in the procedure. The microcatheter was flushed immediately prior to coil insertion, and continuous flushing was performed throughout the procedure. During the procedure, resistance was felt during coil placement. An attempt was made to retrieve the coil; however, the coil became unraveled, making retrieval impossible. An attempt was made to withdraw the coil together with the microcatheter; however, the previously deployed coil, another 20mm x 50cm EMBOLD. Fibered coil, also moved. It might have been the coil that caught on the previously deployed coil. As a result, retrieval was unsuccessful. The unraveled coil was secured with a balloon and cut outside the body. The cut end was pushed into the aneurysm using another coil, resulting in a portion of the coil remaining in the body. The previously deployed coil moved out of position, partially in the target vessel and partially protruding out of t

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889517 ↗
  • InjuryAug 27, 2026Difficult to Remove; Stretched; Failure to Advance; Device-Device Incompatibility

    It was reported that the stretched coil required intervention to retrieve, but there were unretrieved device fragments.A 20mm x 50cm EMBOLD Fibered coil was selected for use in the embolization of a splenic artery aneurysm. During the procedure, resistance was felt during coil placement. An attempt was made to retrieve the coil; however, the coil became unraveled, making retrieval impossible. An attempt was made to withdraw the coil together with the microcatheter; however, a previously deployed coil also moved. As a result, retrieval was unsuccessful. The unraveled coil was secured with a balloon and cut outside the body. The cut end was pushed into the aneurysm using another coil, resulting in a portion of the coil remaining in the body. There were no patient complications as a result of this event.

    Patient problems: Foreign Body In Patient

    MAUDE report 50478241 ↗
  • MalfunctionOct 15, 2025Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Use of Device Problem; Adverse Event Without Identified Device or Use Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 16SEP2025. IT WAS REPORTED THAT THE PROXIMAL RELEASE WAS DIFFICULT TO BREAK. AN EMBOLD FIBERED 20X50 WAS SELECTED FOR A BALLOON-OCCLUDED RETROGRADE TRANSVENOUS OBLITERATION (BRTO) PROCEDURE IN THE GASTRORENAL SHUNT. WHEN ATTEMPTING TO DEPLOY THE COIL, HOWEVER, THE PROXIMAL RELEASE COULD NOT BE DETACHED WHETHER IT WAS BENT OR PULLED. A NEW COIL OF A DIFFERENT MODEL WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE COIL WAS SEPARATED AT THE COUPLERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23299999 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →