NA 170003-28Recalls & reported problems

Nucletron B.V. · Oncentra External Beam/Oncentra Brachy v4.5.3 · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717213052314), received between Oct 12, 2021 and Dec 20, 2022 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

4 other recalls cover devices cleared under the same FDA submission (K121448, K132816); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 5, 2022Class IIOpen, Classified

    A reconstruction error may occur when using the Catheter Bending functionality in Applicator Modeling or Implant Modeling

    Z-0719-2023 · Root cause: Software design · Recall details · FDA record ↗

  2. Oct 13, 2021Class IIOpen, Classified

    Under certain circumstances, there can be a difference between the planned and delivered dose of the two ovoid channels.

    Z-0182-2022 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Use of Device Problem2
  • Product Quality Problem1

Patient problems

  • Insufficient Information2
  • Radiation Burn1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 20, 2022Use of Device Problem; Use of Device Problem

    THE CUSTOMER REPORTED THAT THEY HAD A MISADMINISTRATION INVOLVING ONCENTRA AND FLEXITRON UNIT.

    Patient problems: Insufficient Information

    MAUDE report 16014803 ↗
  • InjuryOct 12, 2021Use of Device Problem; Product Quality Problem; Use of Device Problem

    THE CUSTOMER REPORTED POSSIBLE CATHETER RECONSTRUCTION ISSUE WITH SAVI PLANS.

    Patient problems: Insufficient Information; Radiation Burn

    MAUDE report 12613703 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121448, K132816. One submission often covers many models, so check the product and lot details in each record.

  • Z-0050-2018 · Aug 10, 2017 · Terminated
    Oncentra Brachy 4.5 radiation therapy software The firm name on the label is Nucletron B.V. Oncentra is a radiation therapy planning software designed to analyze and plan radiati
  • Z-0987-2017 · Dec 23, 2016 · Terminated
    Oncentra External Beam Oncentra Brachy Product Usage: Oncentra is radiation therapy planning software designed to analyze and plan radiation treatment is three dimensions for the
  • Z-0604-2016 · Dec 18, 2015 · Terminated
    Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software designed to analyze and plan radia
  • Z-1712-2015 · May 21, 2015 · Terminated
    Oncentra Brachy radiation therapy planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.

Other System,planning,radiation therapy treatment devices

All System,planning,radiation therapy treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page