Oncentra Brachy radiation therapy planning system software recallZ-0719-2023 · Elekta, Inc. · initiated Dec 5, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A reconstruction error may occur when using the Catheter Bending functionality in Applicator Modeling or Implant Modeling
FDA root cause category: Software design
What the company is doing
Elekta issued an Important Field Safety Notice 806-01-BTP-003 dated November 2022 to all affected customers on 12/5/2022 via email. The notice informs users of the specific product and version numbers affected by the issue and any work arounds that can be used to avoid the issue. The notice is to be posted in a place accessible to all users until this action is closed. Appropriate personnel working with the product are to be advised of the content of the letter. The customers are instructed to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta representative as soon as possible but no longer than 30 days from receipt. Update: The manufacturer has now released the solution to this issue for affected Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00. All affected customers will receive a software update in order to solve the issue. The solution was released for installation in the field on 20-MAR-23. Customers are being notified by email on 03/20/2023. Customers are being informed that the fix is available for download/request depending on their standard processes. Elekta service teams will pro-actively follow up with customers to insure the fix has been installed before the target completion date.
Recall details
- Recall number
- Z-0719-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 5, 2022
- Posted by the FDA
- Dec 15, 2022
- Recalling firm
- Elekta, Inc.
- Quantity
- 336 systems
- Product code
- MUJ
- FDA submissions
- K132816
- Distribution
- Distribution US nationwide, including Puerto Rico. There was no military/government distribution.
Product description
Oncentra Brachy radiation therapy planning system software.
Code information (models, lots, serial numbers)
Version 4.0 and above with Applicator Modeling or Implant Modeling; UDI GTIN numbers: 08717213052758, 08717213038660, 08717213020610, 08717213052321, 08717213052314, 08717213052307, 08717213052291, 08717213052246, 08717213052239, 08717213051881, 08717213051843, 08717213051782, 08717213051775, 08717213051294, 08717213020610, 08717213053717, 08717213053700.
Devices named in this recall
16 devices in the FDA device registry whose identifier appears in the recall's code information.
- NA 170003-28Nucletron B.V. · System,planning,radiation therapy treatment2 recalls2 reports
- NA 170607-00Nucletron B.V. · System, planning, radiation therapy treatment2 recalls1 report
- NA 170730Nucletron B.V. · System,planning,radiation therapy treatment2 recalls0 reports
- NA Oncentra BrachyNucletron B.V. · System,planning,radiation therapy treatment2 recalls0 reports
- NA 170730Nucletron B.V. · System,planning,radiation therapy treatment2 recalls0 reports
- NA 170003-23Nucletron B.V. · System,planning,radiation therapy treatment2 recalls0 reports
- NA 170537-04Nucletron B.V. · System,planning,radiation therapy treatment2 recalls0 reports
- NA 170003-24Nucletron B.V. · System, applicator, radionuclide, remote-controlled2 recalls0 reports
- NA 170537-05Nucletron B.V. · System,planning,radiation therapy treatment2 recalls0 reports
- NA 170003-25Nucletron B.V. · System,planning,radiation therapy treatment2 recalls0 reports
- NA 170537-06Nucletron B.V. · System,planning,radiation therapy treatment2 recalls0 reports
- NA 170003-27Nucletron B.V. · System,planning,radiation therapy treatment2 recalls0 reports
- NA 170537-06Nucletron B.V. · System,planning,radiation therapy treatment2 recalls0 reports
- NA 170537-08Nucletron B.V. · System,planning,radiation therapy treatment2 recalls0 reports
- NA 170018-00Nucletron B.V. · System, planning, radiation therapy treatment2 recalls0 reports
- NA 170538-00Nucletron B.V. · System, planning, radiation therapy treatment2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page