Indeflator 1000183Recalls & reported problems
Abbott Vascular Inc. · INDEFLATOR PLUS 30 Inflation Device 30 atm 10 cc · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 08717648013584), received between Aug 1, 2017 and Sep 29, 2020 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Mar 11, 2022Open, Classified
Due to an increase in complaint trend for leaks and intermittent/loose connections.
Z-0924-2022 · Root cause: Component change control · FDA record ↗
Most reported problems
Device problems
- Leak/Splash7
- Break1
Patient problems
- No Consequences Or Impact To Patient5
- No Patient Involvement3
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2020Break; Break
THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE OSTIUM OF THE LEFT ANTERIOR DESCENDING ARTERY AND THE LEFT MAIN ARTERY. THE HANDLE OF THE PLUS 30 INDEFLATOR BROKE INSIDE THE CHAMBER WHILE ATTEMPTING TO DEFLATE AN UNSPECIFIED BALLOON CATHETER. A SYRINGE WAS USED TO DEFLATE THE BALLOON. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NEW INDEFLATOR. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10603163 ↗MalfunctionJul 17, 2020Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING USE OF THE INDEFLATOR PLUS 30, THERE WAS A LEAK OBSERVED. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10292354 ↗MalfunctionOct 30, 2017Leak/Splash
IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI), TWO PLUS 30 INDEFLATORS DID NOT MAINTAIN PRESSURE WHEN CONNECTED TO BALLOON CATHETERS. THE DEVICES WERE REPLACED WITH AN INDEFLATOR FROM ANOTHER LOT AND THE PROCEDURE WAS COMPLETED WITHOUT ISSUE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED. ANALYSIS OF 3 RETURNED UNUSED/STERILE DEVICES NOTED LEAKS WITH THE STOPCOCKS. ROOT CAUSE ANALYSIS DETERMINED THAT THE ISSUE REPORTED WITH THE INDEFLATORS IN THIS CASE WAS LIKELY DUE TO A LEAK IN THE STOPCOCKS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6988904 ↗MalfunctionOct 30, 2017Leak/Splash
IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI), TWO PLUS 30 INDEFLATORS DID NOT MAINTAIN PRESSURE WHEN CONNECTED TO BALLOON CATHETERS. THE DEVICES WERE REPLACED WITH AN INDEFLATOR FROM ANOTHER LOT AND THE PROCEDURE WAS COMPLETED WITHOUT ISSUE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED. ANALYSIS OF 3 RETURNED UNUSED/STERILE DEVICES NOTED LEAKS WITH THE STOPCOCKS. ROOT CAUSE ANALYSIS DETERMINED THAT THE ISSUE REPORTED WITH THE INDEFLATORS IN THIS CASE WAS LIKELY DUE TO A LEAK IN THE STOPCOCKS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6988865 ↗MalfunctionOct 10, 2017Leak/Splash
UNUSED/STERILE DEVICES WERE RETURNED FOR LEAK TESTING. ANALYSIS OF THE RETURNED STERILE DEVICES NOTED 3 OF THE STOPCOCKS LEAKED DURING FUNCTIONAL TESTING.
Patient problems: No Patient Involvement
MAUDE report 6939092 ↗MalfunctionOct 10, 2017Leak/Splash
UNUSED/STERILE DEVICES WERE RETURNED FOR LEAK TESTING. ANALYSIS OF THE RETURNED STERILE DEVICES NOTED 3 OF THE STOPCOCKS LEAKED DURING FUNCTIONAL TESTING.
Patient problems: No Patient Involvement
MAUDE report 6939070 ↗MalfunctionOct 10, 2017Leak/Splash
UNUSED/STERILE DEVICES WERE RETURNED FOR LEAK TESTING. ANALYSIS OF THE RETURNED STERILE DEVICES NOTED 3 OF THE STOPCOCKS LEAKED DURING FUNCTIONAL TESTING.
Patient problems: No Patient Involvement
MAUDE report 6939004 ↗MalfunctionAug 1, 2017Leak/Splash
SUBSEQUENT TO THE INITIAL FILED MEDWATCH REPORT, THE FOLLOWING ADDITIONAL INFORMATION WAS RECEIVED: THE INDEFLATOR WAS PREPPED PER STANDARD PROCEDURE; THE INDEFLATOR WAS CONNECTED TO THE STRAIGHT PORT, A SYRINGE WAS CONNECTED TO THE SIDE PORT, AND NEGATIVE PRESSURE WAS PULLED. THE LEAK AS NOTED WHEN THE INDEFLATOR WAS PRESSURIZED WITH A STARTING PRESSURE OF APPROXIMATELY 6 ATMOSPHERES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6757760 ↗
Other Syringe, balloon inflation devices
- Indeflator 1000184Abbott Vascular Inc. · Syringe, balloon inflation2 recalls155 reports
- Indeflator 1000186-115Abbott Vascular Inc. · Syringe, balloon inflation2 recalls124 reports
- Alliance II M00550600Boston Scientific Corporation · Syringe, balloon inflation191 reports
- Indeflator 1000185Abbott Vascular Inc. · Syringe, balloon inflation1 recall131 reports
- Indeflator 1000186Abbott Vascular Inc. · Syringe, balloon inflation2 recalls56 reports
- Encore™ 26 M001151050Boston Scientific Corporation · Syringe, balloon inflation1 recall54 reports
- PRESTO ID4030Bard Peripheral Vascular, Inc. · Syringe, balloon inflation1 death98 reports
- Encore™ 26 Advantage Kit H74904527011Boston Scientific Corporation · Syringe, balloon inflation1 recall28 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.