Skip to content

HI-TORQUE ADVANCE 1044586Recalls & reported problems

Abbott Vascular Inc. · HI-TORQUE ADVANCE LITE Guide Wire .014 Straight Tip 3.0 cm x 190 cm · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648014468), received between Nov 24, 2020 and Nov 24, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Difficult to Remove1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2020Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE LEFT ANTERIOR DESCENDING ARTERY. A 4.5X20MM NC TREK RX BALLOON DILATATION CATHETER (BDC) WAS PREPARED PER INSTRUCTIONS FOR USE AND WITH A 50/50 RATIO CONTRAST FOR POST-DILATATION. THE BDC WAS INFLATED TWO TIMES AT 14 ATMOSPHERES; HOWEVER, IT WAS NOTED THAT THE BALLOON DID NOT DEFLATE AFTER THE SECOND INFLATION. A NEGATIVE WAS HELD FOR 10 SECONDS, BUT THE BALLOON ONLY PARTIALLY DEFLATED. THE BDC BECAME CAUGHT ON THE NON-ABBOTT GUIDE WIRE. ALL DEVICES WERE REMOVED AS A UNIT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN UNSPECIFIED BALLOON CATHETER. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. ADDITIONAL INFORMATION RECEIVED INDICATING THAT THE GUIDE WIRE THAT BECAME CAUGHT IN THE BDC WAS AN ABBOTT VASCULAR HT ADVANCE LITE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10892314 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →