Skip to content

HI-TORQUE ALL STAR 1001740JRecalls & reported problems

Abbott Vascular Inc. · HI-TORQUE ALL STAR Guide Wire .014 J Tip 3.0 cm x 190 cm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648025945), received between Sep 23, 2019 and Sep 23, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Difficult to Advance1
  • Material Separation1
  • Off-Label Use1

Patient problems

  • Vessel Or Plaque, Device Embedded In1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 23, 2019Off-Label Use; Material Separation; Difficult to Advance; Off-Label Use; Material Separation; Difficult to Advance

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A TORTUOUS AND CALCIFIED PERIPHERAL LOWER LIMB LESION. A HIGH-TORQUE ALLSTAR GUIDE WIRE WAS ADVANCED TO THE LESION WITH RESISTANCE FROM THE ANATOMY. A NON-ABBOTT RE-ENTRY CATHETER WAS ADVANCED OVER THE GUIDE WIRE TO TREAT THE LESION; HOWEVER, WAS ADVANCED TOO FAR OVER THE WIRE AND MET RESISTANCE WITH THE GUIDE WIRE TIP. THE TIP OF THE GUIDE WAS SUBINITIMAL [INTENTIONALLY] AT THE TIME AND WAS SHEERED OFF [SEPARATED]. AS THE TIP REMAINED SUBINTIMAL IN THE VESSEL WALL, NO ATTEMPTS TO RETRIEVE IT WERE PERFORMED. THE PATIENT WAS REPORTED TO BE DOING WELL POST PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Vessel Or Plaque, Device Embedded In

    MAUDE report 9107493 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page