StarClose SE™ 14679-01Recalls & reported problems
Abbott Vascular Inc. · StarClose SE™ Vascular Closure System · Class 3 · Implant
All models of StarClose SE (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Abbott StarClose SE vascular closure device is used to close an artery after a catheter procedure. Many of the 588 reports describe the clip firing early, not deploying or attaching, or the plunger jamming, and a few mention hematomas. The reports include 535 injuries and no deaths, and reports fell from 29 in the previous 12 months to 16 in the latest 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 586
- Deaths reported
- 0
- Injuries reported
- 535
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 586 adverse event reports that list this exact device (DI 08717648079467), received between Aug 5, 2016 and Sep 29, 2026 — classified as 535 injuries, 51 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (128 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Operates Differently Than Expected128
- Patient Device Interaction Problem83
- Improper or Incorrect Procedure or Method61
- Difficult to Remove59
- Firing Problem54
- Mechanical Jam52
- Positioning Failure51
- Off-Label Use49
Patient problems
- No Consequences Or Impact To Patient332
- No Clinical Signs, Symptoms or Conditions158
- Hematoma29
- Pain18
- Hypersensitivity/Allergic reaction12
- Hemorrhage/Blood Loss/Bleeding10
- Foreign Body In Patient8
- Insufficient Information6
Recent adverse event reports
The 25 most recent of 586 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Difficult to Remove; Failure to Fire; Mechanical Jam
It was reported that this was an arteriotomy closure of a left common femoral artery using a StarClose SE device after a diagnostic angiogram procedure with a 5F sheath. Reportedly, the clip did not deploy and the emergency release did not work; however, the device was ultimately able to be removed. Manual compression was applied to achieve hemostasis. There were no reported adverse patient effects. No additional information was provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936933 ↗MalfunctionSep 28, 2026Noise, Audible; Firing Problem
It was reported that this was an arteriotomy closure using a StarClose device after a procedure. Reportedly, after deploying the clip, the flex guide or delivery tube snapped "[audible noise]", forcing the physician to pull the clip out of artery "[clip stuck on distal end of sheath or device when device was withdrawn]". Manual compression was used to achieve hemostasis. There were no reported adverse patient effects. No additional information was provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917956 ↗MalfunctionMar 25, 2026Malposition of Device; Firing Problem
IT WAS REPORTED THAT THIS WAS A CLOSURE USING A STARCLOSE SE DEVICE AFTER A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE. REPORTEDLY, THE STARCLOSE DID NOT ATTACH TO VESSEL. DIRECT PRESSURE WAS HELD TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24694429 ↗MalfunctionMar 9, 2026Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY USING A STARCLOSE SE DEVICE AFTER A LEFT HEART CATHETERIZATION PROCEDURE. REPORTEDLY, THE STARCLOSE SE DEVICE WAS USED 10 DAYS AFTER EXPIRATION. THE STARCLOSE SE DEVICE WAS SUCCESSFULLY USED TO ACHIEVE HEMOSTASIS. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24547340 ↗MalfunctionFeb 12, 2026Break; Failure to Fire
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY USING A STARCLOSE SE DEVICE AFTER A HEMORRHOIDAL ARTERY EMBOLIZATION PROCEDURE. REPORTEDLY, THE TRIGGER BUTTON WAS PRESSED; HOWEVER, THE CLIP DID NOT DEPLOY. THE CLIP WAS NOT IN THE PATIENT ANATOMY OR THE DEVICE. THE CLIP WAS LATER SEEN LOOSE INSIDE THE PACKAGE WITH THE STARCLOSE SE DEVICE WHEN PACKAGING THE DEVICE AFTER THE PROCEDURE. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED. ANALYSIS OF THE RETURNED STARCLOSE SE DEVICE REVEALED DAMAGE NOTED TO THE ONE VESSEL LOCATOR WING AT THE DISTAL END WAS SEPARATED BUT STILL INTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24329058 ↗MalfunctionDec 17, 2025Product Quality Problem; Failure to Cut; Product Quality Problem; Failure to Cut
IT WAS REPORTED THAT THIS WAS A CLOSURE OF AN UNSPECIFIED VESSEL USING A STARCLOSE DEVICE AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, WHEN COMPLETING STEP 2, (PUSHING THE PLUNGER TO OPEN THE VESSEL LOCATOR WINGS AND INITIATE THE SHEATH SPLIT), THE ENTIRE SHEATH SPLIT AND SEEMED LIKE IT WAS WARPED. THEREFORE, THE CLIP WAS NOT DEPLOYED. THE METHOD OF ACHIEVING HEMOSTASIS WAS NOT REPORTED. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23829934 ↗MalfunctionDec 17, 2025Product Quality Problem; Failure to Cut; Product Quality Problem; Failure to Cut
IT WAS REPORTED THAT THIS WAS A CLOSURE OF AN UNSPECIFIED VESSEL USING A STARCLOSE DEVICE AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, WHEN COMPLETING STEP 2, (PUSHING THE PLUNGER TO OPEN THE VESSEL LOCATOR WINGS AND INITIATE THE SHEATH SPLIT), THE ENTIRE SHEATH SPLIT AND SEEMED LIKE IT WAS WARPED. THEREFORE, THE CLIP WAS NOT DEPLOYED. THE METHOD OF ACHIEVING HEMOSTASIS WAS NOT REPORTED. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23829726 ↗MalfunctionDec 12, 2025Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF A MILDLY CALCIFIED LEFT COMMON FEMORAL ARTERY USING A STARCLOSE DEVICE AFTER AN INTERVENTIONAL PERIPHERAL PROCEDURE WITH A 6F SHEATH. REPORTEDLY WHEN ATTEMPTING TO DEPLOY THE STARCLOSE DEVICE, THE SHEATH WAS UNABLE TO SPLIT WHEN PERFORMING STEP 2. A NEW STARCLOSE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23791311 ↗MalfunctionDec 11, 2025Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF A COMMON FEMORAL ARTERY USING A STARCLOSE SE DEVICE AFTER A NEURO INTERVENTION PROCEDURE, USING A 6F SHEATH. REPORTEDLY, THE SHEATH SPLIT CROOKED DURING STEP 2 (SHEATH SPLIT INITIATION). THEREFORE, THE CLIP WAS NOT DEPLOYED. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23783552 ↗MalfunctionDec 9, 2025Malposition of Device; Malposition of Device
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF A CALCIFIED RIGHT COMMON FEMORAL ARTERY USING A STARCLOSE DEVICE AFTER A DIAGNOSTIC PROCEDURE WITH A 6F SHEATH. REPORTEDLY, THE STARCLOSE SE DID NOT DEPLOY CORRECTLY. COMPUTED TOMOGRAPHY WAS PERFORMED AND IT WAS FOUND THAT THE CLIP HAD DEPLOYED INSIDE THE POSTERIOR WALL. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23761079 ↗MalfunctionNov 24, 2025Malposition of Device; Malposition of Device
IT WAS REPORTED THIS WAS AN ARTERIOTOMY CLOSURE OF A RIGHT COMMON FEMORAL ARTERY USING A PROSTYLE DEVICE AFTER AN INTERVENTIONAL PROCEDURE WITH A 6F SHEATH. REPORTEDLY, DUE TO SCAR TISSUE, THE STARCLOSE DID NOT TAKE TO THE VESSEL WALL. MANUAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23638615 ↗MalfunctionNov 20, 2025Off-Label Use; Mechanical Jam; Off-Label Use; Mechanical Jam
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE MILDLY CALCIFIED RIGHT COMMON FEMORAL ARTERY USING A STARCLOSE SE DEVICE AFTER A LEFT SUPERFICIAL FEMORAL ARTERY INTERVENTIONAL PROCEDURE USING A 6F SHEATH. SCAR TISSUE WAS PRESENT AT THE ACCESS SITE. REPORTEDLY, RESISTANCE WAS MET DURING THUMB ADVANCEMENT IN STEP 3 AND AT THE SAME TIME THE WHOLE STARCLOSE DEVICE POPPED OUT OF THE VESSEL; HOWEVER, THE VESSEL LOCATOR WINGS REMAINED OPENED WITH THE PLUNGER WAS STILL ENGAGED. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23607651 ↗MalfunctionNov 17, 2025Unintended System Motion; Mechanical Jam; Unintended System Motion; Mechanical Jam
PATIENT ID: (B)(6). IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE MILDLY CALCIFIED, RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED WITH A STARCLOSE SE DEVICE AFTER A PERCUTANEOUS TRANSLUMINAL PERIPHERAL ANGIOPLASTY PROCEDURE USING A 6F SHEATH. DURING STEP 3 WHEN THE PHYSICIAN WAS DELIVERING THE STARCLOSE SE DEVICE DOWN THE TUBE TO THE COMMON FEMORAL ARTERY, HE ENCOUNTERED SOME PRESSURE AND THE DEVICE JUMPED A BIT. THERE WAS NO ADVERSE EVENT. MANUAL PRESSURE WAS HELD TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE CONDITION RESOLVED AND THE PATIENT WAS DISCHARGED HOME THE SAME DAY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23570502 ↗MalfunctionNov 10, 2025Firing Problem; Firing Problem
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE CALCIFIED COMMON FEMORAL ARTERY USING A STARCLOSE DEVICE AFTER AN INTERVENTIONAL LOWER EXTREMITY ANGIOGRAM PROCEDURE WITH AN 6F SHEATH. REPORTEDLY, THE STARCLOSE DEVICE DID NOT DEPLOY CORRECTLY AS IT REMAINED ON THE DISTAL END OF THE DEVICE. THERE WAS NO ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23510830 ↗MalfunctionOct 30, 2025Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT THIS WAS A CLOSURE OF AN UNKNOWN VESSEL USING A STARCLOSE DEVICE RELATED TO AN INTERVENTIONAL PROCEDURE. REPORTEDLY, WHEN DEPRESSING THE BLUE PLUNGER, THE PLUNGER DID NOT FULLY DEPRESS AS THE NUMBER TWO WAS NOT VISIBLE IN THE WINDOW. AN UNKNOWN METHOD WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23429695 ↗MalfunctionOct 28, 2025Off-Label Use; Difficult to Open or Close; Off-Label Use; Difficult to Open or Close
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE CALCIFIED RIGHT COMMON FEMORAL ARTERY WITH A STARCLOSE DEVICE USING A 6F SHEATH AFTER A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE. REPORTEDLY, AFTER DEPLOYING THE CLIP, THE VESSEL LOCATOR WINGS STAYED OPEN. THE SAFETY RELEASE MECHANISM WAS USED FOR SUCCESSFUL RELEASE OF THE DEVICE. THE DEPLOYED CLIP DID NOT ACHIEVE HEMOSTASIS; THEREFORE, MANUAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23405135 ↗MalfunctionOct 28, 2025Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY USING A STARCLOSE DEVICE AFTER A RENAL PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) INTERVENTIONAL PROCEDURE USING A 6F SHEATH. REPORTEDLY, THE STARCLOSE DEVICE CAME OUT OF THE VESSEL AFTER ADVANCING THE THUMB ADVANCER. IT HAD CAME OUT WITH VESSEL LOCATOR WINGS OPEN. NO VESSEL OR TISSUE DAMAGE WAS NOTED. THE SHEATH HAD NOT SPLIT COMPLETELY AND THEREFORE THE CLIP WAS NOT DEPLOYED (REMAINED IN DEVICE). MANUAL ARTERIAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23403798 ↗InjuryOct 7, 2025Entrapment of Device; Mechanical Jam; Firing Problem; Entrapment of Device; Mechanical Jam; Firing Problem
IT WAS REPORTED THAT THIS WAS ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A STARCLOSE DEVICE AFTER A CEREBRAL ANGIOGRAM DIAGNOSTIC PROCEDURE USING A 6F SHEATH. STEP ONE THROUGH THREE WERE PERFORMED AS INTENDED. REPORTEDLY, THE CLIP DID NOT DEPLOY AT STEP FOUR, AND THE SAFETY RELEASE DID NOT WORK, AND THE DEVICE WAS STUCK IN THE ANATOMY. A CUT-DOWN WAS THEN PERFORMED TO REMOVE THE DEVICE AND CLIP. HEMOSTASIS WAS ACHIEVED VIA CUT-DOWN AND SURGICAL SUTURING. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23240389 ↗InjuryOct 3, 2025Failure to Cut; Failure to Cut
IT WAS REPORTED DURING USE OF THE STARCLOSE DEVICE, THE SHEATH FAILED TO PROPERLY DEPLOY, DUE TO THE ACCESS SHEATH NOT TEARING PROPERLY. AN UNKNOWN METHOD WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23214196 ↗InjurySep 11, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF A RIGHT COMMON FEMORAL ARTERY USING A STARCLOSE DEVICE AFTER AN INTERVENTIONAL GASTROINTESTINAL BLEED PROCEDURE WITH A 6FR SHEATH. REPORTEDLY, THE VALVE GOT STUCK CAUSING THE INTRODUCER TO BLEED BACK. A NEW STARCLOSE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23025533 ↗InjurySep 11, 2025Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED WITH A STARCLOSE DEVICE AFTER AN INTERVENTIONAL PERIPHERAL PROCEDURE USING A 6F SHEATH. REPORTEDLY, RESISTANCE WAS NOTED WHILE DEPRESSING THE PLUNGER DURING STEP 2. THE #2 WAS VISIBLE HOWEVER, THE DEVICE DID NOT WORK. THE PHYSICIAN ATTEMPTED TO USE THE SAFETY RELEASE BUTTON, BUT DID NOT WORK SINCE THE LOCATOR WINGS HAD NOT BEEN DEPLOYED. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23025049 ↗InjuryAug 18, 2025Failure to Cut; Deformation Due to Compressive Stress; Failure to Cut; Deformation Due to Compressive Stress
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE LEFT COMMON FEMORAL ARTERY USING A SMALL-BORE SHEATH WITH A STARCLOSE DEVICE AFTER A BILATERAL ILIAC STENTING PROCEDURE. REPORTEDLY, THE THUMB ADVANCER COULD NOT BE FULLY ADVANCED. THE SHEATH BECAME SHREDDED, AND IT WAS NOTED THAT THE GARAGE LEAVES WERE BENT. MANUAL COMPRESSION WAS HELD FOR 30 MINUTES TO ACHIEVE HEMOSTASIS. ALTHOUGH A DELAY WAS REPORTED DUE TO THE NEED FOR MANUAL COMPRESSION, THERE WAS NO SIGNIFICANT IMPACT TO THE PATIENT DUE TO THE DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22828563 ↗InjuryAug 13, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A COMMON FEMORAL ARTERY WAS ATTEMPTED USING A STARCLOSE SE DEVICE WITH A 6F SHEATH AFTER AN INTERVENTIONAL LOWER LEFT EXTREMITY ARTERIOGRAM PROCEDURE. REPORTEDLY, THE CLIP DIDN'T DEPLOY IN STEP 4. MANUAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22793635 ↗InjuryJul 23, 2025Mechanical Jam; Mechanical Jam
IT WAS REPORTED THIS WAS AN ARTERIOTOMY CLOSURE OF A RIGHT COMMON FEMORAL ARTERY USING A STARCLOSE DEVICE AFTER AN INTERVENTIONAL PROCEDURE WITH A 6F SHEATH. REPORTEDLY, WHEN PERFORMING STEP 3, THE THUMB ADVANCER WAS UNABLE TO FULLY ADVANCE. THEREFORE, THE STARCLOSE DEVICE WAS REMOVED AND MANUAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22582731 ↗InjuryJul 10, 2025Improper or Incorrect Procedure or Method; Firing Problem; Improper or Incorrect Procedure or Method; Firing Problem
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE LEFT COMMON FEMORAL ARTERY USING A STARCLOSE SE DEVICE AFTER A CAROTID STENT INTERVENTIONAL PROCEDURE USING A 6F SHEATH. REPORTEDLY, ALL STEPS WERE PERFORMED WITHOUT ISSUE; HOWEVER, THE CLIP WAS SEEN AT THE DISTAL END OF THE DEVICE. THE SAFETY RELEASE BUTTON WAS USED TO RETRACT THE VESSEL LOCATOR WINGS; HOWEVER, THE ACCESS PORTS WERE NOT USED TO RETRACT THE THUMB ADVANCER PRIOR TO SLIDING THE SAFETY RELEASE. THE DEVICE WAS PULLED OUT, NO RESISTANCE WAS REPORTED. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22472292 ↗
FDA clearance and classification
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the StarClose SE™ 14679-01
Has the StarClose SE™ 14679-01 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08717648079467), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the StarClose SE™ 14679-01?
The FDA's MAUDE database lists 586 adverse event reports for this device, received between Aug 5, 2016 and Sep 29, 2026. In those reports, the most common specific device problems are “Device Operates Differently Than Expected” (128), “Patient Device Interaction Problem” (83) and “Improper or Incorrect Procedure or Method” (61); the most common patient problems are “Hematoma” (29), “Pain” (18) and “Hypersensitivity/Allergic reaction” (12). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the StarClose SE™ 14679-01?
None of the reports are classified as deaths, 535 as injuries and 51 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the StarClose SE™ 14679-01 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the StarClose SE™ 14679-01, and is it still sold?
The FDA's device database lists Abbott Vascular Inc. as the company responsible for it (brand StarClose SE™, model 14679-01). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DEVICE, HEMOSTASIS, VASCULAR devices
- Perclose™ ProStyle™ 12773-03Abbott Vascular Inc. · Device, hemostasis, vascular8 deaths17,987 reports
- Perclose ProGlide™ 12673-03Abbott Vascular Inc. · DEVICE, HEMOSTASIS, VASCULAR13 deaths10,576 reports
- ANGIO-SEAL 610130Terumo Medical Corporation · Device, hemostasis, vascular3 deaths1,626 reports
- MYNXGRIP MX6721Cordis US Corp. · Device, hemostasis, vascular3 deaths1,336 reports
- MYNXGRIP MX5021Cordis US Corp. · Device, hemostasis, vascular2 recalls897 reports
- MANTA 2115Essential Medical, Inc. · Device, Hemostasis, Vascular14 deaths966 reports
- Cordis ExoSeal Vascular Closure Device EX500Cordis US Corp. · DEVICE, HEMOSTASIS, VASCULAR470 reports
- ANGIO-SEAL 610131Terumo Medical Corporation · Device, hemostasis, vascular1 death406 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page