HI-TORQUE EXTRA S'PORT 22235MRecalls & reported problems
Abbott Vascular Inc. · HI-TORQUE EXTRA S'PORT Guide Wire .014 Straight Tip 3.0 cm x 300 cm · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648079603), received between Nov 3, 2016 and Jun 29, 2021 — classified as 3 injuries.
The most frequently reported specific device problem is “Detachment Of Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment Of Device Component2
- Difficult to Remove2
- Failure to Advance1
- Material Separation1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 29, 2021Difficult to Remove; Material Separation; Failure to Advance; Difficult to Remove; Material Separation; Failure to Advance
InjuryJan 3, 2017Detachment Of Device Component; Difficult to Remove
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE RIGHT CORONARY ARTERY. A 014 S'PORT GUIDE WIRE WAS INSERTED INTO THE LESION AND A 2.25X18MM XIENCE ALPINE STENT WAS IMPLANTED PROXIMALLY. THE GUIDE WIRE WAS PULLED BACK AND A 2.25X23MM XIENCE ALPINE STENT DELIVERY SYSTEM (SDS) WAS ADVANCED. AN ATTEMPT TO ADVANCE THE 2.25X23MM XIENCE ALPINE STENT DISTALLY WAS MADE; HOWEVER, IT COULD NOT CROSS PAST THE 2.25X18MM XIENCE ALPINE STENT. THE 2.25X23MM XIENCE ALPINE SDS WAS REMOVED; HOWEVER, THE GUIDE WIRE BECAME STUCK IN THE STENT. THE GUIDE WIRE SEPARATED AND A MICRO SNARE GOOSENECK RETRIEVAL WIRE WAS USED TO REMOVE THE DISTAL TIP OF THE GUIDE WIRE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6220412 ↗InjuryNov 3, 2016Detachment Of Device Component
IT WAS REPORTED THAT AT THE END OF THE PROCEDURE IN THE ANTERIOR TIBIAL ARTERY, THE TIP OF THE SPORT GUIDE WIRE WAS NOTED TO HAVE SEPARATED. THE TIP WAS EMBEDDED IN A COLLATERAL VESSEL, BUT IT DID NOT IMPACT BLOOD FLOW. TIP EMBOLIZATION WAS NOT A CONCERN AS THE TIP WAS NOTED TO BE STABLE IN THE VESSEL. NO ADDITIONAL INTERVENTION WAS PERFORMED FOR THE TIP DETACHMENT. THE PATIENT DID NOT REPORT SYMPTOMS RELATED TO THIS TIP SEPARATION. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 6077842 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page