Skip to content

HI-TORQUE FLEX-T 1012067-02Recalls & reported problems

Abbott Vascular Inc. · HI-TORQUE FLEX-T Guide Wire System with MICROGLIDE Coating .018 145 cm DOC · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648124112), received between Aug 1, 2024 and Aug 1, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Difficult to Advance3
  • Stretched3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 1, 2024Break; Stretched; Difficult to Advance; Break; Stretched; Difficult to Advance

    IT WAS REPORTED THE PROCEDURE WAS TO TREAT AN UNSPECIFIED VESSEL. THERE WAS SOME RESISTANCE DURING USE OF THE HT FLEX-T GUIDE WIRE AND THE DEVICE WAS REMOVED. IT WAS NOTED UPON REMOVAL THAT THE TIP OF THE WIRE HAD UNRAVELED. THE ISSUE OCCURRED WITH ANOTHER DEVICE FROM THE SAME LOT. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19886095 ↗
  • MalfunctionAug 1, 2024Break; Stretched; Difficult to Advance; Break; Stretched; Difficult to Advance

    RETURNED DEVICE ANALYSIS IDENTIFIED THE GUIDE WIRES WERE KINKED/BENT AND NO BREAKS WERE IDENTIFIED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19885058 ↗
  • MalfunctionAug 1, 2024Break; Stretched; Difficult to Advance; Break; Stretched; Difficult to Advance

    IT WAS REPORTED THE PROCEDURE WAS TO TREAT AN UNSPECIFIED VESSEL. THERE WAS SOME RESISTANCE DURING USE OF THE HT FLEX-T GUIDE WIRE AND THE DEVICE WAS REMOVED. IT WAS NOTED UPON REMOVAL THAT THE TIP OF THE WIRE HAD UNRAVELED. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19885038 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page