HI-TORQUE VERSACORE 1012068-01Recalls & reported problems
Abbott Vascular Inc. · Hi-Torque Versacore Modified J Guide Wire System 145 cm · Class 2
All models of HI-TORQUE VERSACORE (7)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648124150), received between May 7, 2018 and May 22, 2018 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Device Operates Differently Than Expected1
- Difficult to Remove1
- Entrapment of Device1
- Stretched1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 22, 2018Break; Device Operates Differently Than Expected
IT WAS REPORTED THE PROCEDURE WAS TO TREAT A LESION IN THE LEFT RADIAL ARTERY. IT WAS NOTED AFTER THE FIRST CATHETER EXCHANGE THAT THE WHITE JACKET OF THE VERSACORE GUIDE WIRE KEPT MOVING. A NEW VERSACORE GUIDE WIRE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT EFFECTS. RETURN DEVICE ANALYSIS NOTED THE DISTAL END OF THE WHITE JACKET WAS SEPARATED AT THE PROXIMAL END OF THE TEFLON COATING. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7534244 ↗InjuryMay 7, 2018Entrapment of Device; Difficult to Remove; Stretched
IT WAS REPORTED THAT THE DIAGNOSTIC PROCEDURE WAS TO TREAT A MODERATELY TORTUOUS AND NON-CALCIFIED LESION LOCATED IN THE RIGHT RADIAL ARTERY. A 145 CM HT VERSACORE GUIDE WIRE WAS INSERTED INTO THE RIGHT ARM OF THE PATIENT, AND AFTER GAINING RADIAL ACCESS, THE WIRE ADVANCED TO THE AORTIC ARCH TO TAKE IMAGES. AN UNSPECIFIED BALLOON CATHETER THAT WAS ADVANCED OVER THE GUIDE WIRE WAS REMOVED AND ANOTHER UNSPECIFIED CATHETER WAS INSERTED. RESISTANCE MET WITH THE GUIDE WIRE AND THE BALLOON CATHETER COULD NOT ADVANCE AND IT WAS REMOVED, BUT THE GUIDE WIRE COULD NOT MOVE FORWARD OR RETRACT. FLUOROSCOPY SHOWED THE GUIDE WIRE LOOKED STRETCHED. THE GUIDE WIRE REMAINED EMBEDDED IN THE VESSEL WHILE A FEMORAL APPROACH WAS USED TO COMPLETE THE REST OF THE PROCEDURE. THE PATIENT WAS THEN TRANSFERRED TO THE OPERATING ROOM FOR A SURGICAL CUT DOWN. AN INCISION WAS MADE IN THE PATIENT'S ARM, AND A DECISION WAS MADE TO BREAK THE WIRE TO REMOVE 8.7 CM OF THE WIRE FROM THE ARTERY. THE PATIENT WAS THEN SENT TO SURGICAL PATHOLOGY FOR ANOTHER SURGICAL CUT DOWN, AND THE REMAINDER OF THE WIRE WAS REMOVED. THE PATIENT WAS DISCHARGED THE FOLLOWING DAY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7492679 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page