Skip to content

HI-TORQUE BALANCE MIDDLEWEIGHT™ 1011882Recalls & reported problems

Abbott Vascular Inc. · HT BALANCE MIDDLEWEIGHT ELITE™ Guide Wire .014" 3 cm Straight Tip 190 cm w/o marker · Class 2

All models of HI-TORQUE BALANCE MIDDLEWEIGHT (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648144820), received between Nov 28, 2022 and Nov 28, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Difficult to Advance1
  • Difficult to Remove1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 28, 2022Peeled/Delaminated; Difficult to Remove; Difficult to Advance; Peeled/Delaminated; Difficult to Remove; Difficult to Advance

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY TORTUOUS LESION IN THE PROXIMAL RIGHT CORONARY ARTERY (PRCA). A 014 HT BMW ELITE GUIDE WIRE (GW) WAS BEING ADVANCED THROUGH THE GUIDE CATHETER WHEN RESISTANCE WAS FELT BETWEEN THE GUIDE WIRE AND THE GUIDE CATHETER. THE GUIDE WIRE WAS REMOVED AND ONCE OUTSIDE THE PATIENT ANATOMY, SEGMENTS OF THE GUIDE WIRE FELT AND LOOKED LIKE IT HAD BEEN SHAVED. IMAGING SHOWED NO GUIDE WIRE COATING IN THE ANATOMY AND NO FOREIGN MATERIAL WAS REMOVED FROM THE PATIENT. A NEW BMW GUIDE WIRE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15874928 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page