HI-TORQUE WINN 1012469Recalls & reported problems
Abbott Vascular Inc. · HT WINN 80 Guide Wire .014" x 300 cm · Class 2
All models of HI-TORQUE WINN (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08717648156847), received between Dec 13, 2021 and Aug 6, 2024 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Advance2
- Material Separation2
- Peeled/Delaminated2
- Difficult to Remove1
Patient problems
- Device Embedded In Tissue or Plaque2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryAug 6, 2024Material Separation; Failure to Advance; Material Separation; Failure to Advance
IT WAS REPORTED THAT RESISTANCE WAS MET WITH THE WINN GUIDEWIRE AND THE CHALLENGING ANATOMY. APPROXIMATELY 3 MM OF THE GUIDEWIRE TIP BROKE DURING ADVANCEMENT WITH RESISTANCE TO THE HEAVILY CALCIFIED, MILDLY TORTUOUS, 100% STENOSIS LESION AND CHRONIC TOTAL OCCLUSION (CTO) IN THE RIGHT POSTERIOR TIBIAL ARTERY. THE SEPARATED TIP REMAINS LODGED IN THE CTO LESION. THERE WERE NO ADVERSE PATIENT SEQUELAE. AN HT COMMAND ES GUIDEWIRE WAS USED TO COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 19923652 ↗MalfunctionAug 3, 2023Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE PERONEAL ARTERY. THE 014 HT WINN 80 GUIDE WIRE WAS USED SEVERAL TIMES AND IT WAS FELT THAT THE HYDROPHILIC COATING WAS COMING OFF THE DEVICE. THERE WAS NO RESISTANCE REPORTED DURING THE PROCEDURE. UPON REMOVAL, THE TIP OF THE WIRE SHOWED EXPOSED MATERIAL WITHOUT HYDROPHILIC COATING. ANOTHER COMMAND ES GUIDE WIRE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17451384 ↗InjuryMar 16, 2023Material Separation; Failure to Advance; Material Separation; Failure to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A CHRONIC TOTAL OCCLUSION (CTO) IN THE ANTERIOR TIBIAL ARTERY WITH HEAVY CALCIFICATION. THE 014 WINN 80 300CM GUIDEWIRE (GW) WAS ATTEMPTED TO BE ADVANCED TO THE TARGET LESION; HOWEVER, WHEN TORQUEING THE GW 10MM OF THE TIP SNAPPED OFF IN THE OCCLUDED VESSEL AND THE TIP OF THE GW BECAME EMBEDDED IN THE VESSEL. DUE TO THE VESSEL BEING TOTALLY OCCLUDED THERE WAS NO ATTEMPT TO RETRIEVE THE SEPARATED GW TIP; THEREFORE, THE TIP OF THE GW REMAINED EMBEDDED IN THE VESSEL AND NO FURTHER TREATMENT WAS PERFORMED. THERE WAS NO ADVERSE PATIENT SEQUELA. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 16553839 ↗MalfunctionDec 13, 2021Peeled/Delaminated; Difficult to Remove; Peeled/Delaminated; Difficult to Remove
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT AN UNSPECIFIED LESION IN THE ANTERIOR TIBIAL ARTERY. WHILE TRYING TO REMOVE THE DEVICE FROM THE ANATOMY, RESISTANCE WAS FELT WITH THE MICROCATHETER AND THE COATING ON THE WINN GUIDE WIRE STRIPPED OFF INSIDE OF THE MICROCATHETER. THE GUIDE WIRE AND THE MICROCATHETER HAD TO BE REMOVED AS A SINGLE UNIT. NO COATING STRIPPED OFF IN THE ANATOMY, AS IT REMAINED INSIDE OF THE MICROCATHETER. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY DURING THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12985034 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page