HI-TORQUE WINN 1012469Recalls & reported problems

Abbott Vascular Inc. · HT WINN 80 Guide Wire .014" x 300 cm · Class 2

All models of HI-TORQUE WINN (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08717648156847), received between Dec 13, 2021 and Aug 6, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Failure to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Advance2
  • Material Separation2
  • Peeled/Delaminated2
  • Difficult to Remove1

Patient problems

  • Device Embedded In Tissue or Plaque2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 6, 2024Material Separation; Failure to Advance; Material Separation; Failure to Advance

    IT WAS REPORTED THAT RESISTANCE WAS MET WITH THE WINN GUIDEWIRE AND THE CHALLENGING ANATOMY. APPROXIMATELY 3 MM OF THE GUIDEWIRE TIP BROKE DURING ADVANCEMENT WITH RESISTANCE TO THE HEAVILY CALCIFIED, MILDLY TORTUOUS, 100% STENOSIS LESION AND CHRONIC TOTAL OCCLUSION (CTO) IN THE RIGHT POSTERIOR TIBIAL ARTERY. THE SEPARATED TIP REMAINS LODGED IN THE CTO LESION. THERE WERE NO ADVERSE PATIENT SEQUELAE. AN HT COMMAND ES GUIDEWIRE WAS USED TO COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 19923652 ↗
  • MalfunctionAug 3, 2023Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE PERONEAL ARTERY. THE 014 HT WINN 80 GUIDE WIRE WAS USED SEVERAL TIMES AND IT WAS FELT THAT THE HYDROPHILIC COATING WAS COMING OFF THE DEVICE. THERE WAS NO RESISTANCE REPORTED DURING THE PROCEDURE. UPON REMOVAL, THE TIP OF THE WIRE SHOWED EXPOSED MATERIAL WITHOUT HYDROPHILIC COATING. ANOTHER COMMAND ES GUIDE WIRE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17451384 ↗
  • InjuryMar 16, 2023Material Separation; Failure to Advance; Material Separation; Failure to Advance

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A CHRONIC TOTAL OCCLUSION (CTO) IN THE ANTERIOR TIBIAL ARTERY WITH HEAVY CALCIFICATION. THE 014 WINN 80 300CM GUIDEWIRE (GW) WAS ATTEMPTED TO BE ADVANCED TO THE TARGET LESION; HOWEVER, WHEN TORQUEING THE GW 10MM OF THE TIP SNAPPED OFF IN THE OCCLUDED VESSEL AND THE TIP OF THE GW BECAME EMBEDDED IN THE VESSEL. DUE TO THE VESSEL BEING TOTALLY OCCLUDED THERE WAS NO ATTEMPT TO RETRIEVE THE SEPARATED GW TIP; THEREFORE, THE TIP OF THE GW REMAINED EMBEDDED IN THE VESSEL AND NO FURTHER TREATMENT WAS PERFORMED. THERE WAS NO ADVERSE PATIENT SEQUELA. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 16553839 ↗
  • MalfunctionDec 13, 2021Peeled/Delaminated; Difficult to Remove; Peeled/Delaminated; Difficult to Remove

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT AN UNSPECIFIED LESION IN THE ANTERIOR TIBIAL ARTERY. WHILE TRYING TO REMOVE THE DEVICE FROM THE ANATOMY, RESISTANCE WAS FELT WITH THE MICROCATHETER AND THE COATING ON THE WINN GUIDE WIRE STRIPPED OFF INSIDE OF THE MICROCATHETER. THE GUIDE WIRE AND THE MICROCATHETER HAD TO BE REMOVED AS A SINGLE UNIT. NO COATING STRIPPED OFF IN THE ANATOMY, AS IT REMAINED INSIDE OF THE MICROCATHETER. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY DURING THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12985034 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →