Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648173974), received between Feb 26, 2019 and Aug 18, 2023 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Difficult or Delayed Positioning” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Difficult or Delayed Positioning1
- Material Separation1
- Material Too Rigid or Stiff1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Intimal Dissection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 18, 2023Difficult or Delayed Positioning; Difficult or Delayed Positioning
IT HAS BEEN REPORTED THE PROCEDURE WAS IN THE SUBCLAVIAN VEIN WITH NO STENOSIS OR TORTUOSITY. THE HI-TORQUE 014 BHW HYDRO 190CM GUIDE WIRE AND THE HT POWERTURN 190CM GUIDE WIRE COULD NOT BE SEEN DURING PROCEDURE. THE RADIOPACITY TIP MARK OF BOTH GUIDE WIRES COULD NOT BE VISUALIZED. ANOTHER HT BHW GUIDE WIRE WAS USED IN THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17574714 ↗MalfunctionJun 12, 2023Material Separation; Material Separation
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MODERATELY CALCIFIED, 100% STENOSED LESION IN THE DISTAL RIGHT CORONARY ARTERY (RCA). A POWERTURN GUIDE WIRE (GW) WAS ADVANCED TO THE LESION WITHOUT ISSUE, AND AN OPTICAL COHERENCE TOMOGRAPHY (OCT) CATHETER WAS ADVANCED OVER THE GW TO THE LESION WITHOUT ISSUE. THE PROCEDURE WAS PERFORMED WITHOUT ISSUE. THE OCT CATHETER AND POWERTURN GW WERE REMOVED TOGETHER, WITHOUT RESISTANCE. AFTER REMOVAL OF THE POWERTURN FROM THE OCT CATHETER, IT WAS NOTED THAT THE TIP OF THE GW HAD SEPARATED AND REMAINED IN THE OCT CATHETER. THERE WAS NO PART OF ANY DEVICE REMAINING IN THE PATIENT ANATOMY. THE PROCEDURE WAS COMPLETED AT THAT POINT SO NO REPLACEMENT DEVICES WERE REQUIRED. THERE WAS NO ADVERSE PATIENT EFFECT. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17110899 ↗InjuryFeb 26, 2019Material Too Rigid or Stiff; Material Too Rigid or Stiff
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN THE RIGHT CORONARY ARTERY. WHILE ADVANCING THE HI-TORQUE POWERTURN GUIDE WIRE TO THE VESSEL, THE WIRE JUMPED FORWARD, STRAIGHTENING THE VESSEL AND DISSECTING IT NEAR THE PROXIMAL RIGHT OSTIAL. PATIENT WAS SENT TO SURGERY FOR SUCCESSFUL REPAIR. THE PATIENT WAS REPORTED DOING FINE AFTER SURGERY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Intimal Dissection
MAUDE report 8372173 ↗
Other Wire, guide, catheter devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page